FREESTYLE® FR995-19 (UDI 00643169002098)Recalls & reported problems

Medtronic, Inc. · VLV FR995-19 FREESTYLE AOR US TEMP IND · Class 3 · Implant

All models of FREESTYLE (33)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
1
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00643169002098), received between Dec 8, 2016 and Sep 30, 2026 — classified as 1 death, 10 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information9
  • Adverse Event Without Identified Device or Use Problem2
  • Device Stenosis2
  • Gradient Increase2
  • Appropriate Device Problem Term/Code Not Available1
  • Calcified1
  • Increase in Pressure1
  • Intravalvular regurgitation1

Patient problems

  • No Information6
  • Aortic Valve Stenosis3
  • Insufficient Information3
  • Calcium Deposits/Calcification2
  • Corneal Pannus2
  • Aortic Regurgitation1
  • Aortic Valve Insufficiency/ Regurgitation1
  • Blood Loss1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Calcified; Device Stenosis

    It was reported that 11 years and 1 month post implant of this 19mm aortic bioprosthetic valve, the patient is being evaluated for a transcatheter valve-in-valve replacement. The reason for evaluation was reported as aortic stenosis. Evidence of calcification was seen in the sino-tubular junction (STJ). No intervention or adverse patient effects were reported.

    Patient problems: Aortic Valve Stenosis; Calcium Deposits/Calcification

    MAUDE report 50943946 ↗
  • InjuryAug 23, 2026Gradient Increase; Device Stenosis

    It was reported that 4 years and 6 months post implant of this 19mm aortic bioprosthetic valve implanted in the pulmonary position, in a 4 year old pediatric patient, a 16mm pulmonary transcatheter valve was implanted valve-in-valve. The reason for intervention was reported as severe conduit stenosis with peak gradient of 50-55mmHg. It was noted that the reason for replacement was not the result of somatic outgrowth of the original conduit. No additional adverse patient effects were reported.

    Patient problems: Pulmonary Valve Stenosis

    MAUDE report 50397181 ↗
  • InjuryMar 24, 2026Insufficient Device Problem Information; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 2 MONTHS POST IMPLANT OF THIS 19MM FREESTYLE AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. A COMPUTED TOMOGRAPHY WAS CONDUCTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information

    MAUDE report 24673122 ↗
  • InjuryJul 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 7 MONTHS POST IMPLANT OF THIS 19MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM BIOPROSTHETIC VALVE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19745604 ↗
  • InjuryMay 17, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 5 MONTHS POST IMPLANT OF THIS 19MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19347510 ↗
  • InjuryFeb 28, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 2 MONTHS POST IMPLANT OF THIS 19MM BIOPROSTHETIC AORTIC VALVE IMPLANTED IN A (B)(6)-YEAR-OLD PEDIATRIC PATIENT, IT WAS EXPLANTED AND REPLACED WITH A 25MM BIOPROSTHETIC VALVE OF THE SAME MODEL. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: No Information

    MAUDE report 9769460 ↗
  • InjurySep 18, 2019Gradient Increase; Increase in Pressure; Insufficient Device Problem Information; Gradient Increase; Increase in Pressure; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 10 MONTHS POST IMPLANT OF THIS BIOPROSTHETIC VALVE, IMPLANTED IN THE PULMONARY POSITION IN A (B)(6) YEAR OLD PEDIATRIC PATIENT, A TRANSCATHETER PULMONARY BIOPROSTHETIC VALVE (TPBV) WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: High Pulmonary Arterial Wedge Pressure; No Information

    MAUDE report 9083097 ↗
  • InjuryMar 19, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 7 MONTHS POST IMPLANT OF THIS 19MM AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A 21MM VALVE OF THE SAME MODEL. THE REASON FOR THE EXPLANT WAS NOT REPORTED.  NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Corneal Pannus; No Information

    MAUDE report 8431529 ↗
  • InjuryFeb 20, 2019Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 7 MONTHS POST IMPLANT OF THIS 19MM AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A 21MM VALVE OF THE SAME MODEL.  THE REASON FOR THE EXPLANT WAS NOT REPORTED.  NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Valve Stenosis; Corneal Pannus; No Information

    MAUDE report 8356047 ↗
  • DeathApr 19, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    MEDTRONIC RECEIVED INFORMATION THAT 14 YEARS AND 7 MONTHS POST IMPLANT OF THIS 23MM AORTIC BIOPROSTHETIC VALVE, THE VALVE WAS EXPLANTED AND REPLACED WITH A 19MM VALVE OF THE SAME MODEL.  THE REASON FOR THE EXPLANT WAS NOT REPORTED.  NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED THAT THE REASON FOR THE VALVE REPLACEMENT WAS THAT A COMPUTED TOMOGRAPHY (CT) SCAN REVEALED AN AORTIC RUPTURE. EMERGENT SURGERY WAS PERFORMED AND THE 23MM VALVE WAS EXPLANTED AND REPLACED WITH A 19MM VALVE OF THE SAME MODEL. THE PHYSICIAN REPORTED THAT THE PATIENT WAS UNABLE TO BE WEANED OFF OF CARDIOPULMONARY BYPASS (CPB) AND THE STATED CAUSE OF DEATH WAS PERSISTENT BLEEDING.

    Patient problems: Blood Loss; Death; Rupture

    MAUDE report 7442576 ↗
  • InjuryOct 23, 2017Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY FOUR YEARS AND NINE MONTHS POST IMPLANT OF THIS BIOPROSTHETIC VALVE, THE VALVE WAS REPLACED VALVE-IN-VALVE WITH A MEDTRONIC TRANSCATHETER VALVE FOR UNKNOWN REASONS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Calcium Deposits/Calcification; No Information

    MAUDE report 6970418 ↗
  • InjuryDec 8, 2016Structural Problem; Insufficient Device Problem Information; Structural Problem; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT APPROXIMATELY THREE YEARS, THREE MONTHS POST IMPLANT OF THIS BIOPROSTHETIC AORTIC VALVE, THIS VALVE WAS REPLACED, VALVE-IN-VALVE WITH A TRANSCATHETER BIOPROSTHETIC VALVE. NO FAILURE MECHANISM AND NO OTHER ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Aortic Regurgitation; Dyspnea; No Information

    MAUDE report 6156550 ↗

Other heart-valve, non-allograft tissue devices

All heart-valve, non-allograft tissue devices →