FREESTYLE® FR995-25Recalls & reported problems
Medtronic, Inc. · VLV FR995-25 FREESTYLE AOR US TEMP IND · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic FREESTYLE FR995-25 is a bioprosthetic aortic heart valve. Its 99 adverse event reports include 9 deaths, 83 injuries and 7 malfunctions. The reports mostly describe the valve leaking backward or narrowing over time, often years after implant, with calcium buildup, and many patients needing a second valve placed inside it. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 100
- Deaths reported
- 9
- Injuries reported
- 84
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 100 adverse event reports that list this exact device (DI 00643169002159), received between Nov 30, 2016 and Sep 16, 2026 — classified as 9 deaths, 84 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (47 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information47
- Intravalvular regurgitation20
- Adverse Event Without Identified Device or Use Problem17
- Patient Device Interaction Problem14
- Device Stenosis8
- Calcified5
- Gradient Increase3
- Material Split, Cut or Torn3
Patient problems
- Insufficient Information32
- Aortic Valve Insufficiency/ Regurgitation26
- Aortic Valve Stenosis13
- No Information8
- Non specific EKG/ECG Changes8
- Endocarditis7
- Calcium Deposits/Calcification6
- Heart Block4
Recent adverse event reports
The 25 most recent of 100 reports, in the reporter's own words as published by the FDA.
InjurySep 16, 2026Insufficient Device Problem Information
Medtronic received information that 13 years post implant of this 25mm aortic bioprosthetic valve, a 26mm aortic transcatheter valve was implanted valve-in-valve. The reason for intervention was not reported. No additional adverse patient effects were reported.
Patient problems: Insufficient Information
MAUDE report 50784941 ↗InjuryJun 12, 2026Patient Device Interaction Problem
IT WAS REPORTED THAT 4 YEARS AND 5 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A BI-VENTRICULAR DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE BI-VENTRICULAR DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY, COMPLETE HEART BLOCK (THIRD-DEGREE ATRIOVENTRICULAR BLOCK) AND SECONDARY PREVENTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiomyopathy; Heart Block
MAUDE report 25483898 ↗InjuryJun 9, 2026Calcified; Device Stenosis; Intravalvular regurgitation
IT WAS REPORTED THAT 13 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SEVERE AORTIC CENTRAL REGURGITATION AND VALVE STENOSIS. EVIDENCE OF CALCIFICATION WAS SEEN IN THE PATIENT ANATOMY. SUBSEQUENTLY ON (B)(6) 2026 THE PATIENT SUCCESSFULLY HAD A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Calcium Deposits/Calcification
MAUDE report 25440500 ↗InjuryApr 16, 2026Intravalvular regurgitation
IT WAS REPORTED THAT 11 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE CENTRAL AORTIC REGURGITATION. IT WAS REPORTED THAT AN INTERVENTION WAS PLANNED FOR (B)(6) 2026. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 24894765 ↗InjuryMar 31, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT 11 YEARS AND 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 25MM AORTIC BIOPROSTHETIC VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24739838 ↗InjuryMar 18, 2026Intravalvular regurgitation
IT WAS REPORTED THAT 11 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC CENTRAL REGURGITATION. IT WAS REPORTED THAT AN INTERVENTION IS REQUIRED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 24625905 ↗InjuryMar 17, 2026Calcified; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR) AND MILD AORTIC STENOSIS (AS). IT WAS ALSO REPORTED THAT THE VALVE HAD CALCIFICATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Calcium Deposits/Calcification
MAUDE report 24619059 ↗InjuryMar 10, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT 4 YEARS AND 1 MONTH POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS REPORTED THAT EVIDENCE OF ANEURYSMS WAS SEEN BELOW ANNULUS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aneurysm
MAUDE report 24562302 ↗MalfunctionJan 26, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT 12 YEARS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 24177677 ↗InjuryJan 22, 2026Calcified; Insufficient Device Problem Information
IT WAS REPORTED THAT 7 YEARS AND 1 MONTH POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Calcium Deposits/Calcification; Insufficient Information
MAUDE report 24144465 ↗InjuryJan 5, 2026Intravalvular regurgitation
IT WAS REPORTED THAT 6 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC INSUFFICIENCY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 23958345 ↗InjuryDec 4, 2025Perivalvular Leak; Perivalvular Leak
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED SUBSEQUENTLY, 2 DAYS LATER, A DIFFERENT MANUFACTURERS TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Valvular Insufficiency/ Regurgitation
MAUDE report 23713985 ↗MalfunctionNov 7, 2025Device Stenosis; Device Stenosis
IT WAS REPORTED THAT 10 YEARS AND 11 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED WAS AORTIC STENOSIS. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 23497734 ↗InjuryOct 23, 2025Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation
ADDITIONAL INFORMATION RECEIVED REPORTED THE REASON FOR EVALUATION AS SEVERE AORTIC INSUFFICIENCY. INTERVENTION IS SCHEDULED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification
MAUDE report 23367687 ↗InjurySep 10, 2025Material Split, Cut or Torn; Intravalvular regurgitation; Material Split, Cut or Torn; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY SECONDARY TO A PERFORATION OF THE RIGHT CORONARY LEAFLET. IT WAS REPORTED THAT THE STENTLESS AORTIC ROOT HAD BECOME DILATATED AND ANEURYSMAL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aneurysm; Aortic Valve Insufficiency/ Regurgitation; Cusp Tear
MAUDE report 23016832 ↗MalfunctionAug 25, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 1 MONTHS POST IMPLANT OF THIS 25MM FREESTYLE AORTIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 22884112 ↗MalfunctionAug 19, 2025Packaging Problem; Packaging Problem
IT WAS REPORTED THAT THE PACKAGING FOR THIS 25MM FREESTYLE VALVE WAS COMPROMISED WITH A TEAR, RIP, OR HOLE RESULTING IN A BROKEN VALVE JAR. THE DEVICE WAS INSPECTED PRIOR TO USE AND WAS NEVER IMPLANTED. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22837120 ↗InjuryJun 24, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SECOND-DEGREE ATRIOVENTRICULAR (AV) BLOCK. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 22317657 ↗InjuryMay 27, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS POST IMPLANT OF THIS FREESTYLE VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SICK SINUS SYNDROME (SSS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 22086959 ↗InjuryMay 14, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION 8 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22020404 ↗InjuryMay 14, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 5 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS "PRIMARY PREVENTION". IT WAS STATED THAT THE PATIENT WAS NEVER SHOCKED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Atrial Fibrillation; Cardiomyopathy; Dyspnea; Fatigue; Insufficient Information; Tachycardia
MAUDE report 22019092 ↗InjuryApr 22, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT THE MEDTRONIC FREESTYLE DEVICE WAS EXPLANTED FROM A PATIENT. THE EXPLANTATION PROCEDURE WAS PERFORMED ON (B)(6) 2025. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Insufficient Information
MAUDE report 21884187 ↗InjuryApr 17, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 7 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC INSUFFICIENCY. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 2 DAYS LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 21858652 ↗DeathMar 19, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 1 WEEK POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT DIED. THE CAUSE OF DEATH WAS REPORTED AS A FALSE LUMEN RUPTURE FROM AORTIC DISSECTION. IT WAS FURTHER REPORTED THAT THE IMPLANT WAS A COMPLICATED AORTIC CASE. HOWEVER, RELATEDNESS OF THE DEATH TO THE VALVE COULD NOT BE EXCLUDED. IT WAS STATED THAT THE HEALTH CARE PROFESSIONAL (HCP) DOES NOT BELIEVE THAT THE DEATH WAS RELATED TO THE VALVE. IT IS UNKNOWN WHETHER AN AUTOPSY WAS PERFORMED.
Patient problems: Aortic dissection; Rupture
MAUDE report 21649157 ↗InjuryMar 11, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 9 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE TO SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 21575258 ↗
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page