FREESTYLE® FR995-27 (UDI 00643169008137)Recalls & reported problems
Medtronic, Inc. · VLV FR995-27 FREESTYLE AOR US TEMP IND · Class 3 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Freestyle aortic heart valve is a tissue valve implanted to replace the aortic valve in the heart. Of its 128 reports, 115 are injuries, 3 are malfunctions and 10 describe deaths; many describe the valve leaking backward or narrowing, with some patients needing a new valve placed inside it. Reports fell to 12 in the last 12 months from 29 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 129
- Deaths reported
- 10
- Injuries reported
- 116
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 129 adverse event reports that list this exact device (DI 00643169008137), received between Jan 3, 2017 and Sep 30, 2026 — classified as 10 deaths, 116 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (69 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information69
- Intravalvular regurgitation31
- Patient Device Interaction Problem23
- Adverse Event Without Identified Device or Use Problem19
- Device Stenosis8
- Calcified5
- Material Split, Cut or Torn5
- Gradient Increase3
Patient problems
- Aortic Valve Insufficiency/ Regurgitation39
- Insufficient Information38
- Aortic Valve Stenosis16
- No Information16
- Endocarditis8
- Non specific EKG/ECG Changes8
- Aortic Regurgitation6
- Heart Block6
Recent adverse event reports
The 25 most recent of 129 reports, in the reporter's own words as published by the FDA.
InjurySep 30, 2026Patient Device Interaction Problem
Medtronic received information that 6 years and 8 months post implant of this 27mm aortic bioprosthetic valve, a Bi-Ventricular Defibrillator was implanted. The reason for the Bi-Ventricular Defibrillator was reported as low ejection fraction (EF). No additional adverse patient effects were reported.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 50942994 ↗InjurySep 29, 2026Insufficient Device Problem Information
Medtronic received information that 8 days post implant of this 27mm aortic bioprosthetic valve, it was explanted and replaced with a new 27mm aortic bioprosthetic valve of the same model. The reason for the replacement was not reported. No additional adverse patient effects were reported.
Patient problems: Insufficient Information
MAUDE report 50928922 ↗InjuryMay 5, 2026Gradient Increase; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 8 MONTHS POST IMPLANT OF THIS 27MM FREESTYLE AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC REGURGITATION (AR). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 25083335 ↗InjuryMar 26, 2026Intravalvular regurgitation
IT WAS REPORTED THAT 12 YEARS AND 4 MONTHS POST IMPLANT OF THIS 27 MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC CENTRAL REGURGITATION AND SHORTNESS OF BREATH (DYSPNEA) WITH FATIGUE. EVIDENCE OF CALCIFICATION WAS SEEN IN THE SINO-TUBULAR JUNCTION (STJ) AND THE NEAR CONCENTRIC CALCIFICATION THE ABDOMINAL AORTA. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification; Dyspnea; Fatigue
MAUDE report 24697773 ↗InjuryMar 25, 2026Insufficient Device Problem Information; Intravalvular regurgitation
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THE REASON FOR EVALUATION WAS SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information
MAUDE report 24687131 ↗InjuryFeb 9, 2026Calcified; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT AN UNKNOWN DURATION POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED FOR SEVERE AORTIC INSUFFICIENCY AND MODERATE AORTIC STENOSIS. SUBSEQUENTLY, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. IT WAS ALSO REPORTED THAT THE VALVE WAS CALCIFIED, INCLUDING THE SINOTUBULAR JUNCTION AND LEAFLETS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Calcium Deposits/Calcification
MAUDE report 24293610 ↗InjuryJan 30, 2026Insufficient Device Problem Information; Device Stenosis
ADDITIONAL INFORMATION RECEIVED THE REASON FOR EVALUATION WAS REPORTED AS STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 24220134 ↗InjuryJan 19, 2026Insufficient Device Problem Information; Device Stenosis; Intravalvular regurgitation
IT WAS REPORTED THAT AN UNKNOWN DURATION POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Insufficient Information
MAUDE report 24108851 ↗InjuryDec 19, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT 10 DAYS POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS LOW EJECTION FRACTION (EF). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23848491 ↗InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT 4 YEARS AND 2 MONTHS POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM AORTIC BIOPROSTHETIC VALVE OF A DIFFERENT MODEL. THE REASON FOR REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23642851 ↗InjuryNov 24, 2025Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS REPORTED THAT 10 YEARS AND 9 MONTHS POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH 25MM AORTIC BIOPROSTHETIC VALVE OF A DIFFERENT MODEL. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23639810 ↗InjuryNov 18, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT 2 YEARS AND 6 MONTHS POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR PERMANENT PACEMAKER WAS REPORTED AS ATRIOVENTRICULAR BLOCK (MOBITZ II). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 23581105 ↗InjuryNov 4, 2025Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation
IT WAS REPORTED THAT 11 YEARS AND 1 MONTH POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information
MAUDE report 23462318 ↗InjuryOct 17, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 6 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC REGURGITATION. AN INTERVENTION IS REQUIRED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 23327846 ↗InjurySep 18, 2025Intravalvular regurgitation; Intravalvular regurgitation
IT WAS REPORTED THAT 13 YEARS AND 1 MONTH POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC CENTRAL REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 23084162 ↗InjurySep 10, 2025Insufficient Device Problem Information; Patient Device Interaction Problem; Intravalvular regurgitation; Insufficient Device Problem Information; Patient Device Interaction Problem; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 13 YEARS AND 4 MONTHS POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information
MAUDE report 23022154 ↗InjuryAug 26, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 6 DAYS POST IMPLANT OF THIS 27MM AORTIC BIOPROSTHETIC VALVE, IT WAS REPORTED THAT THE PATIENT HAD A DEFIBRILLATOR IMPLANTED. THE REASON FOR DEFIBRILLATOR IMPLANT WAS REPORTED AS VENTRICULAR TACHYCARDIA. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Tachycardia
MAUDE report 22898599 ↗InjuryAug 19, 2025Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED THE REASON FOR REPLACEMENT AS PATIENT OUTGROWTH. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 22842560 ↗InjuryAug 11, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
ADDITIONAL INFORMATION WAS RECEIVED THAT PRIOR TO THE DEFIBRILLATOR IMPLANT, THE PATIENT UNDERWENT COMPLETE ELECTROPHYSIOLOGY STUDY. IT WAS REPORTED THAT THE STUDY WAS ABNORMAL AS THE PATIENT HAD A "VERY LONG ANTEROGRADE AV NODE BLOCK CYCLE LENGTH WITH MARKEDLY PROLONGED BASELINE H -V INTERVAL". NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Cardiac Arrest; Non specific EKG/ECG Changes; Ventricular Fibrillation
MAUDE report 22771011 ↗InjuryJul 29, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 4 MONTHS POST IMPLANT OF THIS FREESTYLE VALVE, A DEFIBRILLATOR WAS IMPLANTED. THE REASON FOR THE DEFIBRILLATOR WAS REPORTED AS CARDIOMYOPATHY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 22641450 ↗InjuryJul 11, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 4 MONTHS POST IMPLANT OF THIS FREESTYLE AORTIC VALVE, THE PATIENT WAS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 1 DAY LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22479683 ↗InjuryJul 8, 2025Intravalvular regurgitation; Intravalvular regurgitation
SUBSEQUENTLY, ONE MONTH LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22434438 ↗InjuryJul 2, 2025Calcified; Insufficient Device Problem Information; Intravalvular regurgitation; Calcified; Insufficient Device Problem Information; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. IT WAS REPORTED THAT ON EVALUATION, CALCIUM COULD BE SEEN IN THE SINOTUBULAR JUNCTION, THE PROXIMAL LEFT CORONARY ARTERY (LCA) AND RIGHT CORONARY ARTERY (RCA). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification
MAUDE report 22384278 ↗InjuryJul 2, 2025Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS AND 4 MONTHS POST IMPLANT OF THIS 27MM FREESTYLE AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE REGURGITATION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 22380312 ↗InjuryJun 17, 2025Calcified; Gradient Increase; Insufficient Device Problem Information; Device Stenosis; Calcified; Gradient Increase; Insufficient Device Problem Information; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 7 MONTHS POST IMPLANT OF THIS FREESTYLE AORTIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 22228682 ↗
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page