Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
6
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00643169027107), received between Apr 13, 2023 and Sep 15, 2025 — classified as 6 injuries, 1 malfunction.

The most frequently reported specific device problem is “Insufficient Device Problem Information” (5 reports).

4 other recalls cover devices cleared under the same FDA submission (P980043); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Insufficient Device Problem Information5
  • Break1
  • Intravalvular regurgitation1

Patient problems

  • Insufficient Information6
  • Mitral Valve Insufficiency/ Regurgitation1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • InjurySep 15, 2025Intravalvular regurgitation; Intravalvular regurgitation

    MEDTRONIC RECEIVED INFORMATION THAT 12 YEARS AND 7 MONTHS POST IMPLANT OF THIS HANCOCK II MITRAL VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE MITRAL REGURGITATION (MR) AND DEGENERATIVE VALVE. IT WAS REPORTED THAT AN INTERVENTION IS SCHEDULED, NO ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Mitral Valve Insufficiency/ Regurgitation

    MAUDE report 23051095 ↗
  • InjuryJun 24, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PREVIOUSLY IMPLANTED MITRAL VALVE WAS INACTIVATED AND REPLACED WITH A NEW MITRAL VALVE IMPLANT. THE PATIENT REPORTED UNCERTAINTY REGARDING THE MANUFACTURER OF THE NEW MITRAL VALVE IMPLANT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Insufficient Information

    MAUDE report 22317920 ↗
  • MalfunctionNov 22, 2024Break; Break

    MEDTRONIC RECEIVED INFORMATION THAT DURING THE OPERATION OF THIS 29MM MITRAL BIOPROSTHETIC VALVE, IT WAS REPORTED THAT THE HOLDING SUTURES BROKE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS THE VALVE WAS CINCHED HALF A DOZEN TIMES AND THE SUTURES BROKE, AS THE CINCH WAS OVERTIGHTENED. NO PATIENT INVOLVEMENT WAS REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 20764639 ↗
  • InjuryJul 1, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 10 YEARS POST IMPLANT OF THIS 29MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 27MM VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 19645154 ↗
  • InjuryOct 5, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 11 YEARS, 8 MONTHS POST IMPLANT OF THIS 29MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A UNKNOWN VALVE OF A DIFFERENT MANUFACTURER. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 17876657 ↗
  • InjuryAug 15, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 2 MONTHS POST IMPLANT OF THIS 29MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 17548348 ↗
  • InjuryApr 13, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    MEDTRONIC RECEIVED INFORMATION THAT 9 YEARS AND 10 MONTHS POST IMPLANT OF THIS 29MM MITRAL BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 31MM VALVE OF THE SAME EXACT MAKE AND MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

    Patient problems: Insufficient Information

    MAUDE report 16733031 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980043. One submission often covers many models, so check the product and lot details in each record.

  • Z-1882-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T505U227 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1883-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1884-2019 · May 21, 2019 · Terminated
    Hancock II T510 Bioprosthesis, Mitral, REF T510C27 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
  • Z-1226-2009 · Jan 26, 2009 · Terminated
    Hancock II Bioprostheses with Cinch, Heart Valve, Model # T505 & T510; Catalog #s T505C2501, T505C2101, T505U25ID , T505C2301, T505C2301, T505C2501, T505C2101, T510C29ID, T510C25ID

Other heart-valve, non-allograft tissue devices