NavLock 9733734 (UDI 00643169089600)Recalls & reported problems
Medtronic Navigation, Inc. · HANDLE 9733734 RATCHETING QC SM STRGHT · Class 1
All models of NavLock (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic NavLock ratcheting handle, a straight manual surgical instrument used with navigation systems during spine and sacroiliac procedures. The 46 reports mostly describe the metal parts of the handle breaking, loosening or detaching — including the end cap that staff strike with a mallet — sometimes leaving the instrument hard to grip, with delays of under an hour in a few cases. No deaths or injuries are described. There are no recalls naming this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 46
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 46 adverse event reports that list this exact device (DI 00643169089600), received between Sep 28, 2016 and Jul 17, 2025 — classified as 46 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (20 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem20
- Break13
- Mechanical Problem13
- Material Deformation12
- Material Fragmentation4
- Device Damaged by Another Device3
- Use of Device Problem3
- Loss of or Failure to Bond2
Patient problems
- No Consequences Or Impact To Patient18
- No Clinical Signs, Symptoms or Conditions12
- No Patient Involvement10
- No Known Impact Or Consequence To Patient6
Recent adverse event reports
The 25 most recent of 46 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 17, 2025Material Fragmentation; Material Fragmentation
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED DURING A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT THE RATCHET HANDLE HAD DETACHMENT ISSUES. THERE WAS NO IMPACT TO PATIENT'S OUTCOME AND SURGICAL DELAY WAS LESS THAN ONE-HOUR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22530374 ↗MalfunctionJun 3, 2025Material Fragmentation; Material Fragmentation
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE PARTS IN THE REAR PART OF THE RATCHET HANDLE HAD BECOME DETACHED DURING PRE-USAGE PREPARATION. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22137942 ↗MalfunctionJan 23, 2025Use of Device Problem; Use of Device Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE BACK END OF THE RATCHETING HANDLE WHERE THEY CAN USE A HAMMER FOR THE AWL APPEARED TO HAVE SUNK INTO THE HAND OR THE PLATE FELL OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21222913 ↗MalfunctionJan 23, 2025Use of Device Problem; Use of Device Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE BACK END OF THE RATCHETING HANDLE WHERE THEY CAN USE A HAMMER FOR THE AWL APPEARED TO HAVE SUNK INTO THE HAND OR THE PLATE FELL OFF. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21222818 ↗MalfunctionJan 15, 2025Material Fragmentation; Mechanical Problem; Material Fragmentation; Mechanical Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THERE WAS BREAKAGE TO THE INSTRUMENT FIXATION PART. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED. THE INSTRUMENT FIXING PART OF THE HANDLE CAME OFF, SO THE INSTRUMENT COULD NOT BE GRASPED BY THE HANDLE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21162389 ↗MalfunctionAug 5, 2024Material Fragmentation; Material Fragmentation
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT ONE OF THE STRAIGHT RATCHETING HANDLES IN THE INSTRUMENT TRACKER SET WAS BROKEN. THE PROBABLE CAUSE OF THE REPORTED ISSUE WAS LIKELY DUE TO DAILY USE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19912407 ↗MalfunctionJun 28, 2022Material Integrity Problem; Material Integrity Problem
2022-JUN-24 00740686 (REP): MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE RATCHET HANDLE WAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14844166 ↗MalfunctionNov 18, 2021Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. THE HANDLES WERE USED THROUGHOUT VARIOUS ASPECTS OF SPINAL PROCEDURES. AT TIMES, IT BECAME NECESSARY TO MALLET THE HANDLES IN AN EFFORT TO BREAK THROUGH SCLEROTIC BONE. IT WAS OBSERVED OVER TIME THAT THE METAL IMPACTION CAP ON THE TOP OF THE HANDLE BECAME LOOSE AND SOMETIMES COMPLETELY DETACHED. ONCE DETACHED, THE ABILITY TO EFFICACIOUSLY MALLET THE HANDLE WAS LOST. SINCE THE INTEGRITY OF THE HANDLE WAS COMPROMISED, PROPER SANITATION CANNOT BE COMPLETED FOR FURTHER USE IN NON-MALLETING SITUATIONS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12836641 ↗MalfunctionNov 18, 2021Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. THE HANDLES WERE USED THROUGHOUT VARIOUS ASPECTS OF SPINAL PROCEDURES. AT TIMES, IT BECAME NECESSARY TO MALLET THE HANDLES IN AN EFFORT TO BREAK THROUGH SCLEROTIC BONE. IT WAS OBSERVED OVER TIME THAT THE METAL IMPACTION CAP ON THE TOP OF THE HANDLE BECAME LOOSE AND SOMETIMES COMPLETELY DETACHED. ONCE DETACHED, THE ABILITY TO EFFICACIOUSLY MALLET THE HANDLE WAS LOST. SINCE THE INTEGRITY OF THE HANDLE WAS COMPROMISED, PROPER SANITATION CANNOT BE COMPLETED FOR FURTHER USE IN NON-MALLETING SITUATIONS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12836640 ↗MalfunctionNov 18, 2021Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. THE HANDLES WERE USED THROUGHOUT VARIOUS ASPECTS OF SPINAL PROCEDURES. AT TIMES, IT BECAME NECESSARY TO MALLET THE HANDLES IN AN EFFORT TO BREAK THROUGH SCLEROTIC BONE. IT WAS OBSERVED OVER TIME THAT THE METAL IMPACTION CAP ON THE TOP OF THE HANDLE BECAME LOOSE AND SOMETIMES COMPLETELY DETACHED. ONCE DETACHED, THE ABILITY TO EFFICACIOUSLY MALLET THE HANDLE WAS LOST. SINCE THE INTEGRITY OF THE HANDLE WAS COMPROMISED, PROPER SANITATION CANNOT BE COMPLETED FOR FURTHER USE IN NON-MALLETING SITUATIONS. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12836624 ↗MalfunctionDec 30, 2020Material Integrity Problem; Material Integrity Problem
ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THE MANUFACTURER REPRESENTATIVE CONFIRMED THAT THERE WAS NO SURGERY DELAY AND STATED THAT THEY NOTICED THAT THE HANDLE CAP WAS SHIFTING, AND INSTRUCTED THE HEALTH CARE PROFESSIONAL TO CHANGE THE HANDLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11095022 ↗MalfunctionDec 5, 2020Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING NAVIGATION AND SPINAL INSTRUMENTS BEING USED INTRA-OPERATIVELY OF A SPINAL L4/5 STENOSIS PROCEDURE. IT WAS REPORTED THAT THE LEGACY RATCHET AND PROVISIONAL BREAK OFF. THE RATCHET HANDLE BROKE. THEY DID NOT BREAK INSIDE THE PATIENT ANATOMY. THERE WAS NO IMPACT ON THE PATIENT INCOME. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT ALL THREE INSTRUMENTS WERE ALL BROKEN. THE PROVISIONAL WOULD NO LONGER RETAIN THE BREAK OFF SCREWS. THE RATCHET HANDLES WERE NO LONGER WORKING. THE CAPS ON TOP FELL OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10950980 ↗MalfunctionFeb 13, 2020Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM THAT WAS USED OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT THE RATCHETING HANDLE IS DAMAGED OUT OF THEIR LOANER SET. NO PATIENT PRESENT.
Patient problems: No Patient Involvement
MAUDE report 9706052 ↗MalfunctionJun 20, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM WHILE OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT THE SILVER TAP CAP WAS COMING OFF THE RATCHETING HANDLE. THERE WAS NO PATIENT PRESENT WHEN THE ISSUE OCCURRED.
Patient problems: No Patient Involvement
MAUDE report 8719903 ↗MalfunctionJun 19, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM WHILE OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT THE RATCHETING HANDLE WAS DAMAGED. THE UPPER METAL CAP WAS LOOSE AND LOST. THERE WAS NO PATIENT PRESENT WHEN THE ISSUE OCCURRED.
Patient problems: No Patient Involvement
MAUDE report 8714761 ↗MalfunctionMay 29, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE RATCHETING HANDLE CAP CAME OFF THE HANDLE DURING STERILE PROCESSING. THERE WAS NO PATIENT PRESENT AT THE TIME OF THE EVENT.
Patient problems: No Patient Involvement
MAUDE report 8649344 ↗MalfunctionMay 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT THE BACK OF THE RATCHETING HANDLE SILVER CAP CAME OFF. THIS WAS NOTICED WHEN INSPECTING THE INSTRUMENT TRAY. THERE WAS NO PATIENT PRESENT AT THE TIME OF THE EVENT.
Patient problems: No Patient Involvement
MAUDE report 8643246 ↗MalfunctionMay 13, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED FOR A SACROILIAC AND THORACOLUMBAR PROCEDURE. IT WAS REPORTED THAT INTRA-OPERATIVELY, THE RATCHETING STRAIGHT HANDLE BROKE APART. THE PROCEDURE WAS COMPLETED USING THE NAVIGATION SYSTEM AND THERE WAS NO REPORTED IMPACT TO PATIENT OUTCOME. THERE WAS NO REPORTED SURGICAL DELAY. A SYSTEM CHECKOUT WAS PERFORMED AND FOUND THAT THE RATCHETING STRAIGHT HANDLE WAS REPLACED. A SYSTEM CHECKOUT WAS SUCCESSFULLY PERFORMED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8603879 ↗MalfunctionMar 18, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED THAT INTRA/PERI-OPERATIVELY DURING THE NAVIGATE TASK OF A SPINAL FUSION PROCEDURE THAT THE SHOCK PLATE OF THE RATCHETING HANDLE BROKE OFF. THERE WAS NO DELAY TO THE PROCEDURE. THERE WAS NO REPORTED IMPACT TO PATIENT OUTCOME.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8428313 ↗MalfunctionMar 11, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM. IT WAS REPORTED OUTSIDE OF A PROCEDURE THAT THE RATCHETING HANDLE WAS BROKEN. NO PATIENT WAS PRESENT AT THE TIME OF THE EVENT.
Patient problems: No Patient Involvement
MAUDE report 8407987 ↗MalfunctionFeb 27, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION FROM A MANUFACTURER REPRESENTATIVE (REP) WHO HAD RECEIVED INFORMATION FROM A HEALTHCARE PROVIDER (HCP) INDICATING THAT THE SCREWDRIVER HANDLE NEEDED TO BE REPLACED BECAUSE THE METAL CAP ON THE END OF THE SCREWDRIVER CAME OFF. THE ISSUE OCCURRED AFTER ANESTHESIA, BUT BEFORE INCISION IN A LUMBAR FUSION PROCEDURE. THERE WAS NO CHANGE IN PATIENT OUTCOME AND THERE WAS NO REPORTED DELAY IN SURGERY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8375557 ↗MalfunctionFeb 1, 2019Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE OF PROCEDURE. IT WAS REPORTED THAT THREE OF THE SITES RATCHET HANDLES HAVE THE TOP SILVER PIECE MISSING. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Patient problems: No Patient Involvement
MAUDE report 8297537 ↗MalfunctionDec 19, 2018Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE A PROCEDURE. IT WAS REPORTED THAT THE SILVER END OF THE RATCHETING HANDLE WHERE THE SURGEON HAMMERS THE HANDLE HAD COME OFF. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Patient problems: No Patient Involvement
MAUDE report 8182661 ↗MalfunctionDec 19, 2018Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE A PROCEDURE. IT WAS REPORTED THAT THE SILVER END OF THE RATCHETING HANDLE WHERE THE SURGEON HAMMERS THE HANDLE HAD COME OFF. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Patient problems: No Patient Involvement
MAUDE report 8182650 ↗MalfunctionNov 5, 2018Material Integrity Problem; Material Integrity Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION DEVICE BEING USED OUTSIDE A PROCEDURE. IT WAS REPORTED THAT WHEN THE MANUFACTURER REPRESENTATIVE RECEIVED THE NAC NAVLOCK TRAY BACK, THE RATCHETING HANDLE WAS DAMAGED. THERE WAS NO PATIENT PRESENT WHEN THIS ISSUE WAS IDENTIFIED.
Patient problems: No Patient Involvement
MAUDE report 8039910 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page