NA 97754 (UDI 00643169109469)Recalls & reported problems
Medtronic, Inc. · INSR 97754 MRICS · Class 3
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169109469), received between Nov 22, 2016 and Sep 20, 2017 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Charging Problem” (2 reports).
9 other recalls cover devices cleared under the same FDA submission (P840001); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Charging Problem2
- Communication or Transmission Problem1
- Device Displays Incorrect Message1
- Failure to Interrogate1
- Overheating of Device1
Patient problems
- Burn, Thermal1
- Erythema1
- Pain1
- Therapeutic Response, Decreased1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 20, 2017Failure to Interrogate; Device Displays Incorrect Message; Charging Problem; Communication or Transmission Problem; Failure to Interrogate; Device Displays Incorrect Message; Charging Problem; Communication or Transmission Problem
INFORMATION WAS RECEIVED FROM A MANUFACTURING REPRESENTATIVE REGARDING A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR FOR NON-MALIGNANT PAIN, POST LUMBAR LAMINECTOMY SYNDROME. IT WAS REPORTED THAT THE DESKTOP CHARGER HAD A BROKEN CONNECTOR PIN AND THE MANUFACTURING REPRESENTATIVE WOULD NOT INSPECT THE RECHARGER FOR DAMAGE. THE RECHARGER WAS UNABLE TO CHARGE NORMALLY. NO PATIENT INJURY WAS REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED OR ANTICIPATED. ADDITIONAL INFORMATION: IT WAS REPORTED THAT A PATIENT CAN'T CHARGE HER INS BECAUSE ALL SHE GETS IS THE REPOSITION ANTENNA SCREEN ON THE RECHARGER. PATIENT STATED THAT SHE HAD CALLED HER MANUFACTURING REPRESENTATIVE (REP), ABOUT THE REPOSITION ANTENNA ISSUE. PATIENT STATED THAT HER CONNECTOR PIN HAD BROKEN PREVIOUSLY AND THE REP HAD CALLED PATIENT SERVICES FOR HER AND SHE WAS SENT A REPLACEMENT DESK TOP CHARGER (DTC) THAT SHE RECEIVED FRIDAY ((B)(6) 2017) AND PATIENT SAID SHE TOLD THE REP THAT SHE RECEIVED THE REPLACEMENT DTC, BUT THE REPLACEMENT DIDN'T RESOLVE THE ISSUE OF THE 'REPOSITION ANTENNA SCREEN' SHE WAS SEEING. PATIENT SAID THAT THE REP TOLD HER TO CALL TO GET A NEW RECHARGER AND TOLD HER THAT HE WOULD MEET UP WITH HER THIS
Patient problems: Pain; Therapeutic Response, Decreased
MAUDE report 6880240 ↗MalfunctionNov 22, 2016Overheating of Device; Charging Problem; Overheating of Device; Charging Problem
A PATIENT WITH AN IMPLANTABLE NEUROSTIMULATOR (INS) REPORTED THAT HER CHARGING ANTENNA BECAME HOT WHEN IN USE. SHE VISITED HER HEALTHCARE PROVIDER (HCP) BECAUSE SHE WAS HAVING TROUBLE GETTING THE INS CHARGED, AS THE ANTENNA BURNS HER BACK AFTER CHARGING FOR AN HOUR, AND HER BACK BECOMES RED. THE PATIENT MET WITH A MANUFACTURER'S REPRESENTATIVE (REP) AT THE CLINIC AND CHARGED FOR AN HOUR WITH THE REP'S CHARGER, AND THE ANTENNA BECAME WARM BUT DID NOT GET TOO HOT. PATIENT DID NOT RECALL SEEING THE THERMOMETER ICON ON THE RECHARGER. PATIENT WAS IMPLANTED FOR NON-MALIGNANT PAIN.
Patient problems: Burn, Thermal; Erythema
MAUDE report 6121428 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P840001. One submission often covers many models, so check the product and lot details in each record.
- Z-0372-2025 · Sep 18, 2024 · Open, ClassifiedMedtronic Recharger Kits, RS6230, RS7230, and RS5220A, containing components WR9230 and WR9220A Recharger for rechargeable implantable neurostimulators. The Rechargers are used wi
- Z-1691-2024 · Mar 28, 2024 · Open, ClassifiedIntellis AdaptiveStim Product Number 97715, Spinal Cord Stimulator
- Z-0780-2024 · Dec 19, 2023 · Open, ClassifiedModel A710 Intellis Clinician Programmer Application, version 2.0.97
- Z-2159-2021 · Jul 9, 2021 · Open, ClassifiedMedtronic Vectris SureScan MRI, Catalog Numbers: a) 977A260 b) 977A275 c) 977A290
- Z-1926-2016 · May 3, 2016 · TerminatedMedtronic RestoreSensor, Model 37714; and, RestoreSensor SureScan, (MRI), Model 97714, Implantable Neurostimulators. Used to treat chronic, intractable pain of the trunk and/or li
- Z-1147-2015 · Jan 9, 2015 · TerminatedThe Medtronic, Quadripolar ITREL 3 IPG, Model 7425, is a multiprogrammable device designed for use in the Stimulation of the Central Nervous System, Peripheral Nervous System, or m
- Z-1524-2014 · Feb 27, 2014 · TerminatedMedtronic neurostimulators. Restore (37711), RestoreADVANCED (37713), RestoreADVANCED SureScan MRI (97713), RestoreSENSOR (37714), RestoreSENSOR SureScan MRI (97714), PrimeAdvanced
- Z-0278-2014 · Sep 30, 2013 · TerminatedMedtronic RestoreUltra, model 37712, and RestoreSensor, model 37714, Multi-program Rechargeable Neurostimulator for Spinal Cord Stimulation.
- Z-1536-06 · Aug 2, 2006 · TerminatedMedtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at: Medtronic, Inc., Villalba, Puerto Rico USA. Medtronic, Inc., Minneapolis
Other STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF devices
- Octrode™ 3186St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF30 deaths32,096 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 deaths23,976 reports
- RestoreSensor™SureScan® 97714Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF6 deaths17,903 reports
- Proclaim™ 3660St. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF2 recalls25 deaths13,336 reports
- Intellis™ AdaptiveStim® 97715Medtronic, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF11,487 reports
- Precision Spectra ™ SC-1132Boston Scientific Neuromodulation Corporation · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF53 deaths11,269 reports
- Proclaim™ 5 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall10 deaths9,661 reports
- Proclaim™ 7 EliteSt. Jude Medical, Inc. · STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF1 recall12 deaths7,299 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page