Affinity Fusion® BB811Recalls & reported problems
Medtronic, Inc. · OXYGENATOR BB811 FUSION W/BALANCE · Class 2
All models of Affinity Fusion (10)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Medtronic Affinity Fusion oxygenator has 90 adverse event reports: no deaths, 7 injuries, and 83 malfunctions. Reports mostly describe material separation, fluid/blood leaks, cracks, and related bypass-circuit issues. This device has 2 recalls: one open about the temperature monitoring adapter coming loose from the oxygenator, and one closed for specific lot numbers of Affinity Fusion oxygenators with Balance1 Biosurface.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 90
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 90 adverse event reports that list this exact device (DI 00643169178168), received between Mar 25, 2019 and Nov 10, 2023 — classified as 7 injuries, 83 malfunctions.
The most frequently reported specific device problem is “Material Separation” (60 reports).
2 FDA recall records name this device's identifier, 1 still open.
5 other recalls cover devices cleared under the same FDA submission (K122827, K203111); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Recalls naming this device
- Dec 12, 2022Class IIOpen, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1023-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Nov 16, 2016Class IITerminated
Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity Fusion Oxygenators with Balance1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous Reservoir, or as part of tubing packs. Medtronic has identified an out-of-specification condition exhibiting additional plastic (flash) in the arterial sampling port adjacent to the arterial outlet port of the oxygenator.
Z-0826-2017 · Root cause: Process control · Recall details · FDA record ↗
Most reported problems
Device problems
- Material Separation60
- Fluid/Blood Leak28
- Crack7
- Increase in Pressure4
- Air/Gas in Device2
- Material Puncture/Hole2
- Restricted Flow rate2
- Gas Output Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions85
- Hemorrhage/Blood Loss/Bleeding9
- Infarction, Cerebral2
- Hypoxia1
- Ischemia1
Recent adverse event reports
The 25 most recent of 90 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 10, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, THE CUSTOMER REPORTED THAT THE CARDIOPULMONARY BYPASS CIRCUIT HAD BEEN PRIMED READY FOR THE CASE. THE CUSTOMER OBSERVED A FEW DROPS OF CLEAR FLUID ON THE FLOOR BENEATH THE OXYGENATOR. THE CUSTOMER THOUGHT IT WAS CONDENSATION OR DRIPS FROM THE HEATER COOLER LINE. ONCE THE PUMP SUCKERS WERE SWITCHED ON AND BLOOD WAS INTRODUCED INTO THE CIRCUIT THE DRIPS WERE INDICATIVE OF A BLOOD LEAK. THE CUSTOMER STATED THAT THEY COULD NOT PINPOINT THE EXACT AREA OF THE LEAK BUT BELIEVED IT WAS FROM THE CENTRAL DOME AT THE TOP OF THE OXYGENATOR. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER STATED THAT THE LEAK WAS FROM THE DEVICE ITSELF, NOT TUBING. THE CUSTOMER REPORTED THAT THEY WERE UNSURE OF EXACT LOCATION AS THE DRIPS WERE SEEN ON THE FLOOR. THE CUSTOMER DESCRIBED IT AS THE DOME AT THE TOP OF THE OXYGENATOR. THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE (EX. DEVICES IN A TRAY OR CUSTOM PACK). THE PATIENT LOST A FEW DRIPS, LESS THAN 10MLS, AS A RESULT OF THE
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 18109545 ↗MalfunctionNov 3, 2023Fluid/Blood Leak; Air/Gas in Device; Fluid/Blood Leak; Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT UNDER PRESSURE WHEN THE CIRCUIT WAS PRIMED THERE WAS A SPRAY OF PRIME NOTED WHEN THE OXYGENATOR MODULE FAILED TO DE-AIR APPROPRIATELY. THE SPRAY WAS COMING FROM THE RECIRCULATION PORT (NORMALLY THE PINK CAP PORT ON THE FUSION MODULE) WHICH HAD 1/4" TUBING AND A CABLE TIE APPLIED BY THE CLEAN ROOM TEAM IN KERKRADE. THE CABLE TIE HAD DISPLACED DUE TO THE FLUID AND THEN THE PERFUSIONIST REAPPLIED A NEW CABLE TIE THE SPRAY OF PRIME CONTINUED. AT THIS TIME THE DECISION WAS MADE TO CHANGE OUT THE CIRCUIT AND THE OXYGENATOR. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18063792 ↗MalfunctionOct 12, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT DURING PRIMING THE DEVICE WAS LEAKING FLUID OUT OF THE TOP. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE NOTED ON THE DEVICE MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER REPORTED THAT THERE WAS NO ANTI-COAGULANT/CHEMICALS/DRUGS USED IN PRIME OR DURING THE CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17918243 ↗MalfunctionAug 17, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, THE CUSTOMER REPORTED THAT WHILST SETTING UP THE PUMP THEY OBSERVED THAT THE TEMPERATURE PORT WAS LOOSE AND CAME APART WHEN THEY INSERTED THE TEMP PROBE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17563296 ↗MalfunctionJul 26, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT AFTER THE USE OF A FUSION OXYGENATOR, THE CUSTOMER REPORTED THAT AT THE END OF THE PROCEDURE, TH ERE WAS A PROBLEM WITH THE TEMPERATURE CONE THAT CAME WITH THE PROBE. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 17395157 ↗MalfunctionJul 24, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE DURING SET-UP OF A FUSION OXYGENATOR, THE CUSTOMER NOTICED THAT THE TEMPERATURE PROBE WAS NOT WELL SECURED. THE DEVICE WAS REPLACED FOR THE PROCEDURE. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TEMPERATURE MONITORING ADAPTER (TMA) DETACHED FROM THE PORT WHEN THE PROBE WAS INSERTED, THE CUSTOMER NOTICED THAT IT WIGGLED AT THAT POINT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17376775 ↗MalfunctionJul 12, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR IN A CARDIOPULMONARY BYPASS SURGERY, IT WAS REPORTED THAT THERE WAS A BLOOD LEAK FROM THE ARTERIAL SAMPLE PORT. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT 10 ML OF PATIENT BLOOD WAS LOST BECAUSE OF THIS LEAK. A TRANSFUSION WAS NOT REQUIRED BECAUSE OF THIS LEAK. THERE WAS NO DAMAGE TO THE PACKAGING. THERE OTHER DEVICES IN THE SAME BOX/SHIPPING CONTAINER WERE NOT DAMAGED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 17305597 ↗MalfunctionJun 22, 2023Fluid/Blood Leak; Air/Gas in Device; Fluid/Blood Leak; Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, THE CUSTOMER REPORTED THAT THE DEVICE WAS LEAKING AND FOAMING. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 17180287 ↗MalfunctionJun 16, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT AFTER THE USE OF A FUSION OXYGENATOR, THE CUSTOMER OBSERVED A LEAK FROM THE TEMPERATURE MONITORING ADAPTER (TMA). THIS WAS OBSERVED AFTER THE CASE DURING TEAR DOWN. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. THERE WAS NO PATIENT BLOOD LOSS DUE TO THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17141825 ↗MalfunctionJun 14, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT WHEN SETTING UP THE DEVICE AFTER PRIMING, THE CUSTOMER NOTED A PRIME LEAK FROM THE TEMPERATURE PORT ON THE ARTERIAL OUTLET OF THE OXYGENATOR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER WAS AWARE OF THE FIELD SAFETY NOTICE ON THE PRODUCT AND WAS VERY CAREFUL WHEN ATTACHING THE TEMPERATURE PROBE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17123916 ↗MalfunctionJun 13, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT THE POINT WHERE THE TEMPERATURE PORT WAS INSERTED CAME OFF. THE DEVICE WAS REPLACED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TEMPERATURE MONITORING ADAPTER (TMA) WAS DETACHED FROM THE PORT BY ROTATING IT NORMALLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17116104 ↗MalfunctionJun 8, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT AT AN UNSPECIFIED TIME, THIS FUSION OXYGENATOR DURING ASSEMBLY OF THE CIRCULATION EXTRA-CORPORELLE (CEC) CIRCUIT, AFTER DE-BUBBLING THE CUSTOMER OBSERVED A PRIMING LEAK AT THE LEVEL OF THE OXYGENATOR BEFORE THE DEPARTURE IN CEC. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO REPORTED ADVERSE PATIENT EFFECT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17085013 ↗MalfunctionJun 6, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT WATER WAS FOUND TO BE LEAKING AT THE WATER OUTLET CONNECTION AT THE LOWER END OF THE OXYGENATOR WHEN IT WAS PRE-FILLED WITH SALINE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17069179 ↗MalfunctionMay 24, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT BLOOD WAS DRIPPING INTERMITTENTLY FROM THE TEMPERATURE PORT. THE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO VISIBLE AIR NOTED IN THE SYSTEM/TUBING. THERE WAS A FEW SMALL DROPS OF PATIENT BLOOD LOSS. NO TRANSFUSION WAS REQUIRED AS A RESULT OF THIS LEAK.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 16990686 ↗MalfunctionMay 24, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT AFTER THE USE OF A FUSION OXYGENATOR, IT WAS REPORTED POST CARDIOPULMONARY BYPASS (CPB), DURING BREAKDOWN OF THE PUMP, THAT THE TEMPERATURE PROBE ADAPTER PULLED OUT DURING REMOVAL OF THE TEMPERATURE PROBE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE TEMPERATURE MONITORING ADAPTER (TMA) PULLED OUT OF THE DEVICE WHILE REMOVING THE TEMPERATURE PROBE. THE TMA WAS NOT LOOSE, THE USER DID NOT APPLY FORCE OR PRESSURE TO TEST THE TMA PRIOR TO USE. NO DAMAGE TO THE DEVICE WAS NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16990665 ↗MalfunctionMay 18, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT AFTER THE USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT THE PERFUSIONIST INFORMED THAT THE TE MPERATURE MONITORING ADAPTER (TMA) HAD COME LOOSE FROM THE DEVICE POST-PROCEDURE WHEN DISASSEMBLING THE PERFUSION CIRCUIT. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE PROBLEM OCCURRED WHEN THE PERFUSIONIST WAS DISASSEMBLING THE DISPOSABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16951817 ↗MalfunctionApr 14, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT THE CUSTOMER OBSERVED A LEAK IN THE ARTERIAL LINE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE DEVICE ITSELF LEAKED. AT FIRST THE CUSTOMER REPLACED THE TUBING BUT THE LEAK STILL OCCURRED, THEN THEY DECIDED TO REPLACE THE DEVICE. THERE WAS NO DAMAGE TO THE DEVICE IDENTIFIED. THERE WAS NO DAMAGE TO THE PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16741651 ↗MalfunctionApr 14, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR , IT WAS REPORTED THAT DURING PRIMING THE PERFUSIONIST OBSERVED PRIME SOLUTION TO BE LEAKING OUT FROM THE TOP OF THE OXYGENATOR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. THE CUSTOMER STATED THE SPECIFIC LOCATION OF THE LEAK COULD NOT BE IDENTIFIED. NO HEPARIN OR OTHER DRUGS WERE ADDED TO THE PRIME SOLUTION. THERE WERE NO NOTICEABLE CRACKS OR DAMAGE TO THE OXYGENATOR OR THE PACKAGING. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO OTHER CONTENTS WITHIN THE PACKAGE. THERE WAS NO AIR IN SYSTEM AND NO ADVERSE PATIENT EFFECTS, AS THE ISSUE OCCURRED DURING PRIMING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16741468 ↗MalfunctionApr 4, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT DURING PRIMING THE ARTERIAL OUTPUT AT THE PURGE LINE OF THE AV LINE, DROPS OF PRIMING SOLUTION DRIPPED OUT. USE OF THE DEVICE WAS REQUESTED BUT IT WAS REPORTED AS ASKED BUT UNKNOWN. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO VISUAL DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16672375 ↗MalfunctionMar 16, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT A TEMPERATURE PROBE WITH A KNOWN KINK WAS INSERTED INTO THE ARTERIAL TEMPERATURE PORT. THE CUSTOMER DID NOT LIKE THE POSITION OF THE OXYGENATOR AND REMOVED THE TEMPERATURE PROBE (WHICH HAD A KNOWN KINK IN IT) AND THE PROBE GOT STUCK INSIDE THE OXYGENATOR TEMPERATURE PORT. THE CUSTOMER USED SOME FORCE AND THE TEMPERATURE MONITORING PORT CAME OFF THE OXYGENATOR. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER USED FORCE TO REMOVE THE TEMPERATURE PROBE OUT OF THE TEMPERATURE MONITORING ADAPTER (TMA) AND THEN THE TMA DETACHED. THE EVENT HAPPENED DURING SETUP. THE CUSTOMER KNOWINGLY USED A TEMPERATURE PROBE THAT WAS BENT. IT GOT STUCK IN THE TMA PORT AND WHEN THEY TRIED TO PULL OUT THE TEMPERATURE PROBE, THE TMA CAME OUT WITH IT. THE CUSTOMER IS AWARE THAT THE PROBLEM WOULD NOT HAVE HAPPENED IF THE BENT TEMPERATURE PROBE WAS NOT USED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16552432 ↗MalfunctionMar 16, 2023Fluid/Blood Leak; Material Separation; Fluid/Blood Leak; Material Separation
ADDITIONAL INFORMATION RECEIVED BY MEDTRONIC THAT DURING UNPACKING AND PRIMING THE CIRCUIT THE CUSTOMER SAW WHEN CONNECTING THE TEMPERATURE PROBE THAT THERE WAS A LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16551621 ↗MalfunctionMar 9, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT THE TEMPERATURE MONITOR ADAPTER (TMA) PORT WAS VISIBLY INSPECTED AND WHEN THE TEMPERATURE PROBE ATP210 WAS GENTLY INSERTED THE PORT CAME AWAY FROM THE OXYGENATOR BODY WITH EASE. THE CUSTOMER STATED THE ISSUE WAS OBSERVED DURING SET UP AND BEFORE PRIMING, THERE WAS NO FORCE APPLIED AND THERE WAS NO DAMAGE TO ANY PACKAGING. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. THERE WERE TWO LOTS OF FUSION OXYGENATOR USED IN THE MANUFACTURE OF THIS CUSTOM TUBING PACK: 225199429 AND 225184613.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16511569 ↗MalfunctionMar 7, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, THE CUSTOMER WAS INSPECTING THE DEVICE DURING SETUP AND BEFORE PRIMING, THE TEMPERATURE MONITORING ADAPTER (TMA) PORT WAS VISIBLY INSPECTED AND WHEN THE TEMPERATURE PROBE ATP210 WAS GENTLY INSERTED THE PORT CAME AWAY FROM THE BODY OF THE DEVICE WITH EASE. THE OXYGENATOR WAS CUT OUT OF THE CUSTOM PACK AND REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. IT WAS ALSO REPORTED THAT NO FORCE WAS APPLIED AND THERE WAS NO DAMAGE TO ANY PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16494059 ↗MalfunctionMar 7, 2023Material Separation; Material Separation
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, THE CUSTOMER WAS INSPECTING THE DEVICE PRIOR TO USE. DURING SETUP AND BEFORE PRIMING, THE TMA PORT WAS VISIBLY INSPECTED AND WHEN THE TEMPERATURE PROBE ATP210 WAS GENTLY INSERTED THE PORT CAME AWAY FROM THE BODY OF THE DEVICE WITH EASE. THE OXYGENATOR WAS CUT OUT OF THE CUSTOM PACK AND REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. IT WAS ALSO REPORTED THAT NO FORCE WAS APPLIED AND THERE WAS NO DAMAGE TO ANY PACKAGING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16494009 ↗MalfunctionFeb 27, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, A LEAK WAS NOTED AT THE INLET OF THE OXYGENATOR MEMBRANE DURING HIGH-FLOW EXTRACORPOREAL CIRCULATION EVEN THOUGH THERE WERE NO ABNORMALITIES DEPRESSION. EXTRACORPOREAL CIRCULATION WAS CONTINUED AT A LOWER FLOW RATE, WHICH RESULTED IN PATIENT COOLING. THERE WAS NO ADVERSE PATIENT EFFECT REPORTED WITH THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 16443038 ↗
FDA clearance and classification
Cleared through 510(k) K203111 on Sep 16, 2021 (decision: Substantially Equivalent). Submitted by Medtronic as “Affinity Fusion Oxygenator System, Affinity Pixie Oxygenation System”. The FDA review took 336 days. FDA record ↗
Cleared through 510(k) K122827 on Dec 14, 2012 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Affinity Fusion Oxygenator With Integrated Arterial Filter And Balance Biosurface Model Bb811”. The FDA review took 91 days. FDA record ↗
The FDA regulates product code DTZ (oxygenator, cardiopulmonary bypass) under 21 CFR 870.4350 as a Class II device. 21 CFR 870.4350 ↗
The FDA lists this type of device as life-sustaining or life-supporting.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K122827, K203111. One submission often covers many models, so check the product and lot details in each record.
- Z-1028-2023 · Dec 12, 2022 · Open, ClassifiedCustom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF CB811. Used in Cardiopulmonary bypass.
- Z-1024-2023 · Dec 12, 2022 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, OXY/CVR COMBO BB841 FUSION W/BALANCE 26L, REF BB841. Used in Cardiopulmonary bypass.
- Z-1026-2023 · Dec 12, 2022 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB841. Used in Cardiopulmonary bypass.
- Z-1025-2023 · Dec 12, 2022 · Open, ClassifiedAffinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, Custom Packs, REF CB811. Used in Cardiopulmonary bypass.
- Z-1027-2023 · Dec 12, 2022 · Open, ClassifiedCustom Pack Perfusion Tubing Packs containing Affinity Fusion Oxygenator with Integrated Arterial Filter and Balance Biosurface, REF BB811. Used in Cardiopulmonary bypass.
Common questions about the Affinity Fusion® BB811
Has the Affinity Fusion® BB811 been recalled?
Yes. 2 FDA recall records name this device's identifier: 1 still open and 1 closed (terminated) by the FDA. The most recent, Z-1023-2023 (Class II), began on Dec 12, 2022: An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or… If you use this device, follow the manufacturer's notice and your healthcare provider's advice.
What problems have been reported with the Affinity Fusion® BB811?
The FDA's MAUDE database lists 90 adverse event reports for this device, received between Mar 25, 2019 and Nov 10, 2023. In those reports, the most common specific device problems are “Material Separation” (60), “Fluid/Blood Leak” (28) and “Crack” (7); the most common patient problems are “Hemorrhage/Blood Loss/Bleeding” (9), “Infarction, Cerebral” (2) and “Hypoxia” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the Affinity Fusion® BB811?
None of the reports are classified as deaths, 7 as injuries and 83 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the Affinity Fusion® BB811 FDA approved?
It was cleared through 510(k) K122827 on Dec 14, 2012; its record lists 2 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the Affinity Fusion® BB811, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand Affinity Fusion®, model BB811). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other OXYGENATOR, CARDIOPULMONARY BYPASS devices
- HLS Set BEQ-HLS 7050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls13 deaths87 reports
- HLS Set BEQ-HLS 5050 USAMaquet Cardiopulmonary AG · Oxygenator, Cardiopulmonary Bypass3 recalls4 reports
- NA HLS Set Advanced 5.0, Small Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS3 recalls0 reports
- A.M.G. PMP Sterile A.M.G. Module PMP NO T.P. SterileEurosets Srl · Oxygenator, Cardiopulmonary Bypass2 deaths106 reports
- CAPIOX®FX CX*FX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass2 deaths102 reports
- CAPIOX®RX CX*RX25RWTerumo Corporation · Oxygenator, Cardiopulmonary Bypass1 death102 reports
- QUADROX-i HMO 50000-USA #SQUADROX-i Small Adult, no FilterMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
- NA HLS Set Advanced 7.0, Adult, BIOLINE CoatingMaquet Cardiopulmonary AG · OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026 · each recall links to the FDA recall notice