4 open recalls name at least one Affinity Fusion model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
206
Deaths reported
2
Injuries reported
10
Recalls
5

We found 10 models of Affinity Fusion by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 206 adverse event reports in the FDA's MAUDE database, received between Mar 25, 2019 and Sep 30, 2026 — classified as 2 deaths, 10 injuries, 194 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (81 reports).

5 FDA recall records name at least one of these models by its device identifier, 4 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 4 reportsDec 2023: 0 reportsJan 2024: 2 reportsFeb 2024: 5 reportsMar 2024: 1 reportsApr 2024: 3 reportsMay 2024: 1 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 3 reportsAug 2025: 1 reportsSep 2025: 2 reportsOct 2025: 5 reportsNov 2025: 2 reportsDec 2025: 5 reportsJan 2026: 24 reportsFeb 2026: 9 reportsMar 2026: 8 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 2 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202624
Last 36 months. The most recent month may be incomplete.

Affinity Fusion recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1026-2023 · Root cause: Process change control · Recall details · FDA record ↗

  2. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1025-2023 · Root cause: Process change control · Recall details · FDA record ↗

  3. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1024-2023 · Root cause: Process change control · Recall details · FDA record ↗

  4. Dec 12, 2022Open, Classified

    An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.

    Z-1023-2023 · Root cause: Process change control · Recall details · FDA record ↗

  5. Nov 16, 2016Terminated

    Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity Fusion Oxygenators with Balance1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous Reservoir, or as part of tubing packs. Medtronic has identified an out-of-specification condition exhibiting additional plastic (flash) in the arterial sampling port adjacent to the arterial outlet port of the oxygenator.

    Z-0826-2017 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗

Most reported Affinity Fusion problems

Device problems

  • Fluid/Blood Leak81
  • Material Separation74
  • Delivered as Unsterile Product31
  • Air/Gas in Device11
  • Crack10
  • Gas Output Problem8
  • Material Puncture/Hole5
  • Obstruction of Flow5

Patient problems

  • No Clinical Signs, Symptoms or Conditions188
  • Hemorrhage/Blood Loss/Bleeding25
  • Hypoxia2
  • Infarction, Cerebral2
  • Abnormal Blood Gases1
  • Ischemia1
  • Ischemia Stroke1
  • Reduced Blood Flow1

Affinity Fusion models

10 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Affinity Fusion adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medtronic, Inc. products

All Medtronic, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page