Affinity FusionRecalls & reported problems across all models
Medtronic, Inc. · 10 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
4 open recalls name at least one Affinity Fusion model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 206
- Deaths reported
- 2
- Injuries reported
- 10
- Recalls
- 5
We found 10 models of Affinity Fusion by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 206 adverse event reports in the FDA's MAUDE database, received between Mar 25, 2019 and Sep 30, 2026 — classified as 2 deaths, 10 injuries, 194 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (81 reports).
5 FDA recall records name at least one of these models by its device identifier, 4 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Affinity Fusion recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1026-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1025-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1024-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Dec 12, 2022Open, Classified
An upward trend of complaints in which the Temperature Monitoring Adapter (TMA), located on the Affinity Fusion Oxygenator, had come loose from the oxygenator either during pre-procedure perfusion set-up or post-procedure when disassembling the perfusion circuit.
Z-1023-2023 · Root cause: Process change control · Recall details · FDA record ↗
- Nov 16, 2016Terminated
Medtronic is initiating a voluntary product recall for specific lot numbers of the Affinity Fusion Oxygenators with Balance1 Biosurface. These are distributed as stand-alone devices, or as a combination unit with the Affinity Fusion Cardiotomy/Venous Reservoir, or as part of tubing packs. Medtronic has identified an out-of-specification condition exhibiting additional plastic (flash) in the arterial sampling port adjacent to the arterial outlet port of the oxygenator.
Z-0826-2017 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
Most reported Affinity Fusion problems
Device problems
- Fluid/Blood Leak81
- Material Separation74
- Delivered as Unsterile Product31
- Air/Gas in Device11
- Crack10
- Gas Output Problem8
- Material Puncture/Hole5
- Obstruction of Flow5
Patient problems
- No Clinical Signs, Symptoms or Conditions188
- Hemorrhage/Blood Loss/Bleeding25
- Hypoxia2
- Infarction, Cerebral2
- Abnormal Blood Gases1
- Ischemia1
- Ischemia Stroke1
- Reduced Blood Flow1
Affinity Fusion models
10 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Affinity Fusion® BB811OXYGENATOR, CARDIOPULMONARY BYPASS2 recalls90 reports
- Affinity Fusion® BB841RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS2 recalls36 reports
- Affinity Fusion™ CB811OXYGENATOR, CARDIOPULMONARY BYPASS1 recall4 reports
- Affinity Fusion™ CB841OXYGENATOR, CARDIOPULMONARY BYPASS1 recall0 reports
- Affinity Fusion™ BB841Reservoir, blood, cardiopulmonary bypass44 reports
- Affinity Fusion™ CB811OXYGENATOR, CARDIOPULMONARY BYPASS13 reports
- Affinity Fusion™ BB811OXYGENATOR, CARDIOPULMONARY BYPASS2 deaths11 reports
- Affinity Fusion™ RCL841CATHETER, CANNULA AND TUBING, VASCULAR, CARDIOPULMONARY BYPASS5 reports
- Affinity Fusion® BB841RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS2 reports
- Affinity Fusion® BB841Reservoir, blood, cardiopulmonary bypass1 report
Recent Affinity Fusion adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Fluid/Blood Leak
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 50946068 ↗MalfunctionAug 6, 2026Fluid/Blood Leak
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50159968 ↗MalfunctionJul 8, 2026Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR IT WAS REPORTED BLOOD LEAKED, AND SEVERAL PRODUCTS LEAKED BLOOD CONTINUOUSLY. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE (EX. DEVICES IN A TRAY OR CUSTOM PACK). THE EXACT NUMBER OF DEVICES WHICH LEAKED WAS UNKNOWN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25749836 ↗MalfunctionJul 8, 2026Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR IT WAS REPORTED BLOOD LEAKED, AND SEVERAL PRODUCTS LEAKED BLOOD CONTINUOUSLY. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE (EX. DEVICES IN A TRAY OR CUSTOM PACK). THE EXACT NUMBER OF DEVICES WHICH LEAKED WAS UNKNOWN.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 25749831 ↗MalfunctionJun 29, 2026Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION HOLLOW FIBER OXYGENATOR, IT WAS REPORTED THAT THERE WAS A FLUID LEAK ON THE BOTTOM OF THE DEVICE. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE (EX. DEVICES IN A TRAY OR CUSTOM PACK).THERE WAS NO BLOOD LOSS.THE LEAK WAS FROM THE OXYGENATOR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25652961 ↗MalfunctionApr 8, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR/CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT WHEN IT WAS OPENED, THE PACKAGING WAS FOUND TO BE DAMAGED. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE DEVICE ITSELF WAS NOT DAMAGED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24814208 ↗MalfunctionApr 1, 2026Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT AFTER UNPACKING AND DURING PRE-FLUSHING OF THE OXYGENATOR, A WATER LEAK WAS OBSERVED. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24751430 ↗MalfunctionMar 31, 2026Gas Output Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT THE DEVICE DID NOT OXYGENATED WELL. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24740975 ↗MalfunctionMar 30, 2026Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR, IT WAS REPORTED THAT DURING THE OXYGENATOR PRIMING PROCESS A SEVERE WATER LEAK WAS OBSERVED. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24726687 ↗MalfunctionMar 23, 2026Delivered as Unsterile Product
IT WAS REPORTED THAT BEFORE A SURGICAL PROCEDURE, THE FUSION OXYGENATOR/CARDIOTOMY VENOUS RESERVOIR PACKAGING WAS FOUND TO BE DAMAGED, THE STERILE PACKAGE WAS COMPROMISED, AND THE PRODUCT WAS CONTAMINATED. THE DEVICE WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24660105 ↗MalfunctionMar 23, 2026Fluid/Blood Leak
IT WAS REPORTED THAT DURING USE OF THE FUSION OXYGENATOR, BLEEDING OCCURRED. THE PRODUCT WAS REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 24659915 ↗InjuryMar 19, 2026Gas Output Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION HOLLOW FIBER OXYGENATOR, IT WAS REPORTED THAT AFTER THE START OF EXTRACORPOREAL CIRCULATION, AN OXYGEN-AIR BLENDER WAS USED TO PROVIDE ROUTINE VENTILATION TO THE DEVICE AT 2 L/MIN WITH FIO2 AT 70%. IT WAS OBSERVED THAT THE ARTERIAL BLOOD AT THE DEVICE OUTLET APPEARED DARK, AND CONTINUOUS MONITORING SHOWED A SUSTAINED DECREASE IN THE PATIENT'S SVO2 (MIXED VENOUS OXYGEN SATURATION) AND SPO2 (PERIPHERAL CAPILLARY OXYGEN SATURATION). RAPID CHECKS OF THE OXYGEN TUBING, OXYGEN-AIR BLENDER CONNECTIONS, AND BOTH AIR AND OXYGEN PRESSURES REVEALED NO ABNORMALITIES. FIO2 WAS IMMEDIATELY ADJUSTED TO 100% AND FLOW INCREASED TO 4 L/MIN, BUT THERE WAS STILL NO IMPROVEMENT. THE SURGICAL AND ANESTHESIA TEAMS WERE NOTIFIED, AND THE CIRCUIT WAS STOPPED FOR REINFUSION. AFTER SWITCHING TO AN OXYGEN CYLINDER AS THE GAS SOURCE, THE ARTERIAL BLOOD BECAME NOTICEABLY REDDER, SO EXTRACORPOREAL CIRCULATION FLOW WAS GRADUALLY INCREASED TO FULL FLOW. SUBSEQUENT BLOOD GAS ANALYSIS SHOWED A SIGNIFICANT IMPROVEMENT IN ARTERIAL OXYGEN PARTIAL PRESSURE, BUT CARBON DIOXIDE PARTIAL PRESSURE REMAINED ABNORMALLY HIGH. AFTER VENTILATING WITH PURE OXYGEN AT MORE T
Patient problems: Abnormal Blood Gases; Hypoxia
MAUDE report 24637071 ↗MalfunctionMar 18, 2026Air/Gas in Device
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A FUSION OXYGENATOR AND FUSION CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT FOAM WAS OBSERVED ON THE SURFACE OF THE FUSION CARDIOTOMY VENOUS RESERVOIR. THE USE OF THE DEVICE WAS UNSPECIFIED. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24625963 ↗MalfunctionMar 12, 2026Fluid/Blood Leak; Gas Output Problem; Air/Gas in Device
IT WAS REPORTED THAT DURING USE OF THE FUSION OXYGENATOR, THERE WERE ALLEGATIONS OF A BLOOD LEAKAGE, BUBBLING, NO GAS OUTPUT AT THE GAS OUTLET, AND POOR CARBON DIOXIDE REMOVAL. THE PRODUCT WAS REPLACED. IT WAS UNKNOWN IF THERE WAS ANY PATIENT COMPLICATIONS AS A RESULT OF THE EVENT.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 24580541 ↗MalfunctionMar 11, 2026Gas Output Problem
IT WAS REPORTED THAT DURING THE FUSION OXYGENATOR PERFORMANCE WAS SUBOPTIMAL, RESULTING IN INADEQUATE OXYGENATION AND INSUFFICIENT CARBON DIOXIDE REMOVAL AT STANDARD VENTILATION SETTINGS. THE USE OF THE DEVICE WAS UNSPECIFIED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24566911 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page