NA 5487Recalls & reported problems
Medtronic, Inc. · CABLE 5487 DISP 6FT STER ROHS 5PK · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169307339), received between Aug 7, 2018 and Oct 29, 2020 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Material Integrity Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Integrity Problem5
- Failure to Capture4
- Capturing Problem1
- Connection Problem1
- Over-Sensing1
- Pacing Problem1
Patient problems
- No Consequences Or Impact To Patient4
- Cardiac Arrest2
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryOct 29, 2020Failure to Capture; Material Integrity Problem; Failure to Capture; Material Integrity Problem
IT WAS REPORTED THAT WHEN USING A PACING CABLE IT SHOWED SIGNS OF NO CAPTURE AND WAS UNRELIABLE AFTER TWO MINUTES OF USE. THE LEAD WAS CONNECTED TO AN EXTERNAL CARDIAC DEVICE INSTEAD TO OBTAIN A MORE RELIABLE CONNECTION. THE PACING CABLE WAS DISCARDED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10753766 ↗InjuryOct 13, 2020Failure to Capture; Material Integrity Problem; Failure to Capture; Material Integrity Problem
IT WAS REPORTED THAT WHEN USING THE PACING CABLES WITH AN EXTERNAL PULSE GENERATOR THERE WAS NO CAPTURE WHEN THE LEAD WAS CONNECTED INTO THE CABLE HEAD. IT WAS NOTED THAT AN EXPLANTED DEVICE THAT CONNECTED DIRECTLY VIA IS-1 PORT AND LEFT EXTERNAL WAS THEN USED INSTEAD OF THE EXTERNAL PULSE GENERATOR TO RESOLVE THE ISSUE. THE PACING CABLES WERE DISCARDED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10670337 ↗InjuryOct 13, 2020Failure to Capture; Material Integrity Problem; Failure to Capture; Material Integrity Problem
IT WAS REPORTED THAT WHEN USING THE PACING CABLES WITH AN EXTERNAL PULSE GENERATOR THERE WAS NO CAPTURE WHEN THE LEAD WAS CONNECTED INTO THE CABLE HEAD. IT WAS NOTED THAT AN EXPLANTED DEVICE THAT CONNECTED DIRECTLY VIA IS-1 PORT AND LEFT EXTERNAL WAS THEN USED INSTEAD OF THE EXTERNAL PULSE GENERATOR TO RESOLVE THE ISSUE. THE PACING CABLES WERE DISCARDED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10670331 ↗InjurySep 18, 2020Capturing Problem; Material Integrity Problem; Capturing Problem; Material Integrity Problem
IT WAS REPORTED THAT DURING A REVISION PROCEDURE, TWO DAYS AFTER A PATIENT WAS GIVEN A TEMPORARY EXTERNAL PACER, THE CABLES DID NOT WORK AS PACING FROM THE EXTERNAL TEMPORARY PACER DID NOT OCCUR. IT WAS NOTED THAT WHILST TESTING THE LEAD WITH ALLIGATOR CLIP CABLES TO A PROGRAMMER, THE IMPEDANCE AND THRESHOLD READINGS WERE NORMAL HOWEVER WHEN CONNECTING THESE CABLES, THE PATIENT EXPERIENCED ASYSTOLE. THE EXTERNAL TEMPORARY PACER WAS AT MAX OUTPUT AT 100 BEATS PER MINUTE (BPM) ALTHOUGH THERE WAS NO CAPTURE. THE CABLE WAS CHANGED OUT FOR AN ALTERNATIVE CABLE FROM THE SAME LOT WHICH GAVE THE SAME RESULT. A DIFFERENT EXTERNAL TEMPORARY PACER WITH A SECOND SET OF WIRES WAS USED AND STILL SHOWED NO CAPTURE. IN THE EVENT THE LEAD THAT WAS PREVIOUSLY IMPLANTED WAS USED TO CONTINUE TO PACE THE UNIT UNTIL THE CLINICIAN WAS READY TO IMPLANT ON THE OTHER SIDE. THE CABLES HAVE BEEN DISCARDED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Cardiac Arrest
MAUDE report 10552689 ↗MalfunctionJul 6, 2020Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE DISPOSABLE TEMPORARY PACING LEAD FAILED TO PACE THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10233065 ↗InjuryAug 7, 2018Failure to Capture; Over-Sensing; Connection Problem; Material Integrity Problem; Over-Sensing; Connection Problem; Material Integrity Problem
IT WAS REPORTED THAT, WHILE BEING PACED WITH AN EXTERNAL PULSE GENERATOR (EPG) DURING A DEVICE REPLACEMENT PROCEDURE, THE PATIENT WENT INTO ASYSTOLE, WITH LOSS OF CAPTURE AND OVERSENSING EXHIBITED. IT WAS NOTED THAT THE ISSUE AROSE WHEN THE REPLACEMENT LEAD AND PATIENT CABLE WERE BEING TAPED DOWN. THE PATIENT WAS PACED WITH AN EXTERNAL DEFIBRILLATOR AND THE CONNECTION BETWEEN THE PATIENT CABLE AND LEAD WAS ADJUSTED, BUT THE ISSUE PERSISTED. THE LEAD WAS CONNECTED TO A PACING ANALYZER AND SHOWN TO FUNCTION NORMALLY. THE PATIENT CABLE WAS REPLACED, AND THE PATIENT WAS ABLE TO BE PACED WITH THE EPG AND LEAD WITHOUT ISSUE. IT WAS NOTED THAT THE IMPLANTING PHYSICIAN ATTRIBUTED THE ISSUE TO THE CONNECTOR BLOCK OF THE PATIENT CABLE. THE CABLE WAS RETURNED FOR ANALYSIS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Cardiac Arrest
MAUDE report 7756658 ↗
FDA clearance and classification
Cleared through 510(k) K960446 on May 22, 1996 (decision: Substantially Equivalent). Submitted by Medtronic Vascular as “Model 5471 & 5471l Sterile Disposable Patient Cable Assemblies”. The FDA review took 111 days. FDA record ↗
The FDA regulates product code DSA (cable, transducer and electrode, patient, (including connector)) under 21 CFR 870.2900 as a Class II device. 21 CFR 870.2900 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 5487
Has the NA 5487 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169307339), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 5487?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Aug 7, 2018 and Oct 29, 2020. In those reports, the most common specific device problems are “Material Integrity Problem” (5), “Failure to Capture” (4) and “Capturing Problem” (1); the most common patient problem is “Cardiac Arrest” (2). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 5487?
None of the reports are classified as deaths, 5 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 5487 FDA approved?
It was cleared through 510(k) K960446 on May 22, 1996. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA 5487, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 5487). Its record shows it as no longer in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR) devices
- IntraSight 797403Philips Image Guided Therapy Corporation · Cable, Transducer And Electrode, Patient, (Including Connector)81 reports
- ASSY, CABLE, ECG, 15 LEAD, 3 WIRE PRECOR 11111Physio-Control, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)1 recall0 reports
- MR Patient Care Short Wide ECG 2.0 Cable AAMI CAB,4 LD,CV MRI ECGPhilips Ds North America LLC · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)12 reports
- NA 5492ALMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)10 reports
- NA 5492VLMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)10 reports
- NA 5487LMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)8 reports
- CORE M2 Vascular System 400-0100.17Volcano Corporation · Cable, Transducer And Electrode, Patient, (Including Connector)6 reports
- Multi-Link™ 2090101-110AirLife Finland Oy · Cable, Transducer And Electrode, Patient, (Including Connector)4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page