NA 5492ALRecalls & reported problems
Medtronic, Inc. · CABLE 5492AL LONG US MKT · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00643169889668), received between Jun 22, 2022 and May 19, 2025 — classified as 10 malfunctions.
The most frequently reported specific device problem is “Material Integrity Problem” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Material Integrity Problem10
- Failure to Capture3
Patient problems
- No Clinical Signs, Symptoms or Conditions10
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 19, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT CABLE WAS FRAYED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22042408 ↗MalfunctionMay 19, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT CABLE WAS FRAYED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22042392 ↗MalfunctionMay 19, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT CABLE WAS FRAYED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22042340 ↗MalfunctionMay 19, 2025Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT CABLE WAS FRAYED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22042323 ↗MalfunctionJul 5, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT CABLE FAILED TO CAPTURE AND PACING DID NOT OCCUR WHEN IN USE WITH THE EXTERNAL PULSE GENERATOR (EPG) CONNECTED TO THE PATIENT POST-OPERATIVELY. IT WAS NOTED THAT THE ISSUE WAS EXPERIENCED WHEN THE EPG WAS MOVED FROM THE PATIENT'S CHEST TO HANG NEXT TO THE BEDSIDE. IT WAS FURTHER NOTED THAT THE PACING WIRES WERE APPROPRIATELY SCREWED INTO THE LEAD PORTS. THE PATIENT WAS NOT PACING DEPENDENT AND HAD A SLOW ESCAPE. THE ISSUE WAS EXPERIENCED WITH TEN PATIENT CABLES IN TOTAL WITH MULTIPLE EPG'S USED. THE CUSTOMER SWITCHED TO USING DIFFERENT MODEL PATIENT CABLES WHICH RESOLVED THE ISSUE AND PACING RESUMED. UPON FURTHER INVESTIGATION, THE CUSTOMER CONSIDERED THAT THE WEAKNESS WAS NEAR THE STRAIN RELIEF PORTION OF THE PATIENT CABLES WITH LOSS OF CONNECTION IN OR NEAR THE SOCKET. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14912715 ↗MalfunctionJul 5, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THAT THE PATIENT CABLE FAILED TO CAPTURE AND PACING DID NOT OCCUR WHEN IN USE WITH THE EXTERNAL PULSE GENERATOR (EPG) CONNECTED TO THE PATIENT POST-OPERATIVELY. IT WAS NOTED THAT THE ISSUE WAS EXPERIENCED WHEN THE EPG WAS MOVED FROM THE PATIENT'S CHEST TO HANG NEXT TO THE BEDSIDE. IT WAS FURTHER NOTED THAT THE PACING WIRES WERE APPROPRIATELY SCREWED INTO THE LEAD PORTS. THE PATIENT WAS NOT PACING DEPENDENT AND HAD A SLOW ESCAPE. THE ISSUE WAS EXPERIENCED WITH TEN PATIENT CABLES IN TOTAL WITH MULTIPLE EPG'S USED. THE CUSTOMER SWITCHED TO USING DIFFERENT MODEL PATIENT CABLES WHICH RESOLVED THE ISSUE AND PACING RESUMED. UPON FURTHER INVESTIGATION, THE CUSTOMER CONSIDERED THAT THE WEAKNESS WAS NEAR THE STRAIN RELIEF PORTION OF THE PATIENT CABLES WITH LOSS OF CONNECTION IN OR NEAR THE SOCKET. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14912605 ↗MalfunctionJul 5, 2022Material Integrity Problem; Material Integrity Problem
IT WAS REPORTED THAT THE PATIENT CABLE FAILED TO CAPTURE AND PACING DID NOT OCCUR WHEN IN USE WITH THE EXTERNAL PULSE GENERATOR (EPG) CONNECTED TO THE PATIENT POST-OPERATIVELY. IT WAS NOTED THAT THE ISSUE WAS EXPERIENCED WHEN THE EPG WAS MOVED FROM THE PATIENT'S CHEST TO HANG NEXT TO THE BEDSIDE. IT WAS FURTHER NOTED THAT THE PACING WIRES WERE APPROPRIATELY SCREWED INTO THE LEAD PORTS. THE PATIENT WAS NOT PACING DEPENDENT AND HAD A SLOW ESCAPE. THE ISSUE WAS EXPERIENCED WITH TEN PATIENT CABLES IN TOTAL WITH MULTIPLE EPG'S USED. THE CUSTOMER SWITCHED TO USING DIFFERENT MODEL PATIENT CABLES WHICH RESOLVED THE ISSUE AND PACING RESUMED. UPON FURTHER INVESTIGATION, THE CUSTOMER CONSIDERED THAT THE WEAKNESS WAS NEAR THE STRAIN RELIEF PORTION OF THE PATIENT CABLES WITH LOSS OF CONNECTION IN OR NEAR THE SOCKET. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14912592 ↗MalfunctionJun 22, 2022Failure to Capture; Material Integrity Problem; Failure to Capture; Material Integrity Problem
IT WAS REPORTED THAT WHEN IT WAS ATTEMPTED TO USE THE PATIENT CABLE IT WAS UNABLE TO CAPTURE RENDERING IT UNUSABLE. IT WAS NOTED THAT THE PATIENT CABLE WAS CHANGED OUT TO A DIFFERENT PRODUCT WHICH FUNCTIONED PROPERLY TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14775013 ↗MalfunctionJun 22, 2022Failure to Capture; Material Integrity Problem; Failure to Capture; Material Integrity Problem
IT WAS REPORTED THAT WHEN IT WAS ATTEMPTED TO USE THE PATIENT CABLE IT WAS UNABLE TO CAPTURE RENDERING IT UNUSABLE. IT WAS NOTED THAT THE PATIENT CABLE WAS CHANGED OUT TO A DIFFERENT PRODUCT WHICH FUNCTIONED PROPERLY TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14774946 ↗MalfunctionJun 22, 2022Failure to Capture; Material Integrity Problem; Failure to Capture; Material Integrity Problem
IT WAS REPORTED THAT WHEN IT WAS ATTEMPTED TO USE THE PATIENT CABLE IT WAS UNABLE TO CAPTURE RENDERING IT UNUSABLE. IT WAS NOTED THAT THE PATIENT CABLE WAS CHANGED OUT TO A DIFFERENT PRODUCT WHICH FUNCTIONED PROPERLY TO RESOLVE THE ISSUE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14774653 ↗
FDA clearance and classification
Cleared through 510(k) K203556 on Jul 12, 2021 (decision: Substantially Equivalent). Submitted by Medtronic, Inc. as “Medtronic Model 5492A, 5492V, 5492AL, 5492VL Patient Cables”. The FDA review took 217 days. FDA record ↗
The FDA regulates product code DSA (cable, transducer and electrode, patient, (including connector)) under 21 CFR 870.2900 as a Class II device. 21 CFR 870.2900 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the NA 5492AL
Has the NA 5492AL been recalled?
No FDA recall record in our data names this exact device identifier (DI 00643169889668), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the NA 5492AL?
The FDA's MAUDE database lists 10 adverse event reports for this device, received between Jun 22, 2022 and May 19, 2025. In those reports, the most common specific device problems are “Material Integrity Problem” (10) and “Failure to Capture” (3). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA 5492AL?
None of the reports are classified as deaths, 0 as injuries and 10 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA 5492AL FDA approved?
It was cleared through 510(k) K203556 on Jul 12, 2021. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA 5492AL, and is it still sold?
The FDA's device database lists Medtronic, Inc. as the company responsible for it (brand NA, model 5492AL). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR) devices
- IntraSight 797403Philips Image Guided Therapy Corporation · Cable, Transducer And Electrode, Patient, (Including Connector)81 reports
- ASSY, CABLE, ECG, 15 LEAD, 3 WIRE PRECOR 11111Physio-Control, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)1 recall0 reports
- MR Patient Care Short Wide ECG 2.0 Cable AAMI CAB,4 LD,CV MRI ECGPhilips Ds North America LLC · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)12 reports
- NA 5492VLMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)10 reports
- NA 5487LMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)8 reports
- NA 5487Medtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)6 reports
- CORE M2 Vascular System 400-0100.17Volcano Corporation · Cable, Transducer And Electrode, Patient, (Including Connector)6 reports
- Multi-Link™ 2090101-110AirLife Finland Oy · Cable, Transducer And Electrode, Patient, (Including Connector)4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page