IntraSight 797403Recalls & reported problems
Philips Image Guided Therapy Corporation · IntraSight System · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This Philips IntraSight cable, transducer and patient electrode has 91 adverse event reports describing cracked monitors with sharp edges and system faults that delayed procedures. There were no deaths, 2 injuries and 89 malfunctions. No recalls name this device. Reports rose from 19 in the prior 12 months to 37 in the last 12 months.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 81
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 81 adverse event reports that list this exact device (DI 00845225012878), received between May 5, 2021 and Jun 16, 2026 — classified as 1 injury, 80 malfunctions.
The most frequently reported specific device problem is “Sharp Edges” (55 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Sharp Edges55
- Crack10
- No Device Output7
- Visual Prompts will not Clear5
- Smoking4
- Computer Operating System Problem3
- Overheating of Device2
- Communication or Transmission Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions80
- Abrasion1
Recent adverse event reports
The 25 most recent of 81 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 16, 2026Overheating of Device
IT WAS REPORTED THAT THE INTRASIGHT SYSTEM EXPERIENCED A HARDWARE ERROR CODE DURING SET UP. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. A PHILIPS FIELD SERVICE ENGINEER WAS ONSITE TO INSPECT THE INTRASIGHT AND FOUND DAMAGED/BROKEN CONNECTIONS ON THE ACE BOARD, ALONG WITH A BURNT PC WIRE (CONFIRMED FROM PHOTO RECEIVED). THIS PRODUCT PROBLEM IS BEING REPORTED DUE TO A BURNT COMPONENT; POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25527311 ↗MalfunctionMay 13, 2026No Device Output
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN A PROCEDURE. DURING USE, THE AO PRESSURE WAS NOT SHOWING ON THE MONITOR. THE PROCEDURE WAS ABORTED AND RESCHEDULED FOR A LATER DATE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED; RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25164964 ↗MalfunctionMay 1, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR HAS PHYSICAL DAMAGE AND DOES NOT WORK. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25060644 ↗MalfunctionMay 1, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED AND BROKEN. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25060301 ↗MalfunctionApr 30, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED ON THE CORNER AND THE CABLE BROKE AT THE CONNECTION POINT. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR AND CABLE WERE RETURNED FOR EVALUATION. VISUAL INSPECTION OF THE MONITOR CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. FOR THE CABLE, DAMAGE WAS CONFIRMED ON THE CONNECTOR. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25051196 ↗MalfunctionApr 22, 2026No Device Output
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN A PERIPHERAL PROCEDURE. DURING USE, AN IVUS CATHETER WAS PLUGGED INTO THE PIM, BUT COULD NOT BE RECOGNIZED. THE PROCEDURE WAS ABORTED AND RESCHEDULED FOR A LATER DATE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED; RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24955172 ↗MalfunctionMar 20, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS BROKEN. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24656209 ↗MalfunctionMar 20, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24654860 ↗MalfunctionMar 13, 2026No Device Output
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN A CORONARY PROCEDURE. DURING USE, THE YELLOW WAVE FORM (PD) WOULD NOT APPEAR ON THE SYSTEM. THE CASE WAS ABORTED AND THE PATIENT WILL BE RESCHEDULED FOR A LATER DATE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE INTRASIGHT SYSTEM COULD NOT BE USED; RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24597959 ↗MalfunctionMar 6, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS DROPPED AND CRACKED DURING EQUIPMENT MOVEMENT. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH SHARP EDGES OBSERVED. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24536152 ↗MalfunctionMar 5, 2026Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR BROKE. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24524929 ↗MalfunctionMar 5, 2026Sharp Edges
IT WAS REPORTED THAT THE INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24521179 ↗MalfunctionFeb 19, 2026Computer Operating System Problem
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN AN EMERGENT PROCEDURE. DURING USE, THE HEMO SIGNAL WAS UNSTABLE, AND WHEN RECONNECTING THE FM-PIM, THE SYSTEM HAD A "CONNECT FM-PIM" MESSAGE; THEREFORE, USE OF IFR WAS ABORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED DURING AN EMERGENT CASE, RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24390961 ↗MalfunctionFeb 18, 2026Sharp Edges
IT WAS REPORTED THAT THE INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24379851 ↗MalfunctionFeb 18, 2026Sharp Edges
IT WAS REPORTED THAT THE INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED ON THE EDGE. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378492 ↗MalfunctionFeb 18, 2026Sharp Edges
IT WAS REPORTED THAT THE INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED IN THE UPPER LEFT CORNER. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378223 ↗MalfunctionFeb 18, 2026Sharp Edges
IT WAS REPORTED THAT THE INTRASIGHT SYSTEM TOUCHSCREEN MONITOR NOT RESPONDING TO TOUCH. ADDITIONALLY, WHEN PHILIPS FSE WAS ONSITE, A CRACKED SCREEN WAS OBSERVED. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24378211 ↗MalfunctionFeb 10, 2026Computer Operating System Problem
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN AN EMERGENT STEMI PROCEDURE. DURING USE, THE SYSTEM HAD AN ERROR MESSAGE (IVUS IS NOT AVAILABLE); THEREFORE, IT COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITH A MOBILE SYSTEM. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED DURING AN EMERGENT CASE, RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24308135 ↗MalfunctionJan 27, 2026Computer Operating System Problem
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN AN EMERGENT AORTIC DISSECTION PROCEDURE. DURING USE, THE SYSTEM HAD AN UNKNOWN HARDWARE ERROR; THEREFORE, IT COULD NOT BE USED. THE PROCEDURE WAS COMPLETED WITH A MOBILE C-ARM AND ANGIO. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED DURING AN EMERGENT CASE, RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24193459 ↗MalfunctionJan 23, 2026Visual Prompts will not Clear
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN AN EMERGENT PROCEDURE. DURING USE, AN ERROR MESSAGE WAS DISPLAYED ON THE MONITOR, EVEN AFTER MULTIPLE ATTEMPTS TO REBOOT THE SYSTEM. THE PROCEDURE WAS COMPLETED WITHOUT THE IVUS SYSTEM. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED DURING AN EMERGENT CASE; RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24171200 ↗MalfunctionJan 9, 2026No Device Output
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN A CORONARY PROCEDURE. DURING USE, THE PD VALUE WAS INACCURATE, VALUE WAS DRIFTING, AND UNABLE TO NORMALIZE. THE IFR PROCEDURE WAS ABORTED AND THE PATIENT WILL BE TREATED IN A DIFFERENT LAB, SCHEDULED TWO DAYS LATER. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE PROCEDURE WAS ABORTED, RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24030930 ↗MalfunctionDec 24, 2025Sharp Edges; Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23889380 ↗MalfunctionDec 24, 2025Sharp Edges; Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS CRACKED. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23888601 ↗MalfunctionDec 12, 2025Sharp Edges; Sharp Edges
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM TOUCHSCREEN MONITOR WAS BROKEN. THERE WAS NO PATIENT PRESENT AND NO USER INJURY REPORTED. THE TOUCHSCREEN MONITOR WAS RETURNED FOR EVALUATION. VISUAL INSPECTION CONFIRMED A CRACK WITH MISSING SCREEN MATERIAL, RESULTING IN SHARP EDGES. THIS PRODUCT PROBLEM IS BEING REPORTED IN AN ABUNDANCE OF CAUTION BECAUSE THE TOUCHSCREEN MONITOR HAS SHARP EDGES THAT CAN RESULT IN A POTENTIAL FOR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23794990 ↗MalfunctionDec 12, 2025No Device Output; No Device Output
IT WAS REPORTED THAT AN INTRASIGHT SYSTEM WAS USED IN AN EMERGENT CORONARY PROCEDURE. DURING USE, THE FM-PIM BOX WAS NOT SHOWING READINGS ON THE MONITOR. THE IFR ATTEMPT WAS ABORTED, AND THE PATIENT WAS MOVED TO ANOTHER ROOM. THE PROCEDURE WAS CONTINUED USING ANOTHER DEVICE. NO PATIENT INJURY REPORTED. THIS PRODUCT PROBLEM IS BEING REPORTED BECAUSE THE SYSTEM COULD NOT BE USED IN AN EMERGENT CASE, RESULTING IN A POTENTIAL FOR HARM DUE TO THE DELAY OF THE PROCEDURE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23788948 ↗
FDA clearance and classification
Cleared through Special 510(k) K190078 on Feb 11, 2019 (decision: Substantially Equivalent). Submitted by Volcano Corporation as “IntraSight”. The FDA review took 26 days. FDA record ↗
The FDA regulates product code DSA (cable, transducer and electrode, patient, (including connector)) under 21 CFR 870.2900 as a Class II device. 21 CFR 870.2900 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the IntraSight 797403
Has the IntraSight 797403 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00845225012878), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the IntraSight 797403?
The FDA's MAUDE database lists 81 adverse event reports for this device, received between May 5, 2021 and Jun 16, 2026. In those reports, the most common specific device problems are “Sharp Edges” (55), “Crack” (10) and “No Device Output” (7); the most common patient problem is “Abrasion” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the IntraSight 797403?
None of the reports are classified as deaths, 1 as an injury and 80 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the IntraSight 797403 FDA approved?
It was cleared through Special 510(k) K190078 on Feb 11, 2019. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the IntraSight 797403, and is it still sold?
The FDA's device database lists Philips Image Guided Therapy Corporation as the company responsible for it (brand IntraSight, model 797403). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Cable, Transducer And Electrode, Patient, (Including Connector) devices
- ASSY, CABLE, ECG, 15 LEAD, 3 WIRE PRECOR 11111Physio-Control, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)1 recall0 reports
- MR Patient Care Short Wide ECG 2.0 Cable AAMI CAB,4 LD,CV MRI ECGPhilips Ds North America LLC · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)12 reports
- NA 5492ALMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)10 reports
- NA 5492VLMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)10 reports
- NA 5487LMedtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)8 reports
- NA 5487Medtronic, Inc. · CABLE, TRANSDUCER AND ELECTRODE, PATIENT, (INCLUDING CONNECTOR)6 reports
- CORE M2 Vascular System 400-0100.17Volcano Corporation · Cable, Transducer And Electrode, Patient, (Including Connector)6 reports
- Multi-Link™ 2090101-110AirLife Finland Oy · Cable, Transducer And Electrode, Patient, (Including Connector)4 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.