Sprint Quattro Secure® 6947M55 (UDI 00643169309715)Recalls & reported problems

Medtronic, Inc. · LEAD 6947M55 SPRINT USA DF4 MCRD NG ACI · Class 3 · Implant

All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure (10)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Sprint Quattro Secure pacing lead has 390 reports in the FDA database, with 229 injuries and 4 deaths. The reports often describe the lead detecting too much electrical noise, unusually high electrical resistance, and fractured wires. Some reports describe infections or shocks felt from the lead. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
391
Deaths reported
4
Injuries reported
229
Recalls naming it
0

Summary

The FDA's MAUDE database contains 391 adverse event reports that list this exact device (DI 00643169309715), received between Feb 10, 2015 and Sep 1, 2026 — classified as 4 deaths, 229 injuries, 158 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (154 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 1 reportsJan 2024: 3 reportsFeb 2024: 5 reportsMar 2024: 3 reportsApr 2024: 8 reportsMay 2024: 1 reportsJun 2024: 4 reportsJul 2024: 3 reportsAug 2024: 4 reportsSep 2024: 2 reportsOct 2024: 3 reportsNov 2024: 3 reportsDec 2024: 1 reportsJan 2025: 8 reportsFeb 2025: 2 reportsMar 2025: 2 reportsApr 2025: 2 reportsMay 2025: 3 reportsJun 2025: 2 reportsJul 2025: 7 reportsAug 2025: 2 reportsSep 2025: 3 reportsOct 2025: 3 reportsNov 2025: 4 reportsDec 2025: 6 reportsJan 2026: 2 reportsFeb 2026: 3 reportsMar 2026: 2 reportsApr 2026: 2 reportsMay 2026: 0 reportsJun 2026: 4 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports2024202520268
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing154
  • High impedance107
  • Fracture78
  • Impedance Problem73
  • Signal Artifact/Noise56
  • Adverse Event Without Identified Device or Use Problem47
  • High Capture Threshold46
  • Low impedance39

Patient problems

  • No Clinical Signs, Symptoms or Conditions177
  • No Known Impact Or Consequence To Patient106
  • Unspecified Infection38
  • Shock from Patient Lead(s)23
  • No Consequences Or Impact To Patient17
  • Chest Pain5
  • Dyspnea5
  • Sepsis5

Recent adverse event reports

The 25 most recent of 391 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 1, 2026Low impedance

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50539693 ↗
  • MalfunctionAug 14, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50259962 ↗
  • MalfunctionJun 15, 2026Low impedance

    IT WAS REPORTED THAT ON THE DAY OF A GENERATOR EXCHANGE, THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR LOW PACING IMPEDANCE. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25505010 ↗
  • MalfunctionJun 10, 2026Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED PACING IMPEDANCE WARNING FOR LOW  IMPEDANCE.  THE PATIENT HAD EXPERIENCED HYPERVENTILATION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Hyperventilation

    MAUDE report 25462273 ↗
  • MalfunctionJun 9, 2026High impedance; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A WARNING FOR UNDEFINED HIGH PACING IMPEDANCES AND UNDEFINED HIGH RV DEFIBRILLATION COIL IMPEDANCES. THE RIGHT ATRIAL (RA) LEAD TRIGGERED A WARNING FOR UNDEFINED HIGH PACING IMPEDANCES. THE LEFT VENTRICULAR (LV) LEAD TRIGGERED A WARNING FOR UNDEFINED HIGH PACING IMPEDANCES. IN ADDITION, THE RV LEAD EXHIBITED UNDERSENSING AND THE RA LEAD EXHIBITED OVERSENSING. THE RV, RA AND LV LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25446535 ↗
  • MalfunctionJun 2, 2026Over-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT UNDERSENSING AND OVERSENSING. THE RV LEAD REMAINS IN US E. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25368874 ↗
  • InjuryApr 28, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 25023142 ↗
  • MalfunctionApr 8, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING AND A BLANKED BEAT. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24821234 ↗
  • InjuryMar 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED A POCKET INFECTION. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Medical device site infection

    MAUDE report 24549954 ↗
  • MalfunctionMar 5, 2026Fracture; Over-Sensing; Impedance Problem; Unstable Impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A SUSPECTED FRACTURE. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR PACING IMPEDANCE VARIATION AND SENSING INTEGRITY COUNTER (SIC). IT WAS FURTHER REPORTED THAT THE RV LEAD EXHIBITED PACING IMPEDANCE SPIKES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24523659 ↗
  • MalfunctionFeb 26, 2026Low impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN IMPEDANCE ALERT FOR LOW PACING IMPEDANCES. THE RV LEAD PACING IMPEDANCES WERE TRENDING CONSISTENT WITH HISTORICAL VALUES. IN ADDITION, THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED INVALID DATA WITH THE P WAVE TRENDS. THE RV LEAD AND ICD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24457962 ↗
  • InjuryFeb 19, 2026Signal Artifact/Noise; Fracture; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A SUSPECTED FRACTURE ON THE LOW VOLTAGE PORTION OF THE LEAD. IT WAS FURTHER REPORTED THAT MULTIPLE LEAD INTEGRITY ALERTS (LIA) WERE TRIGGERED TO DUE HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). IT WAS FURTHER REPORTED THAT THE LEAD EXHIBITED NOISE, OVERSENSING AND T-WAVE OVERSENSING (TWOS). THE LEAD WAS REPROGRAMMED AND REMAINS IN USE. THE PATIENT WAS PLACED ON A LIFE-VEST. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24393015 ↗
  • InjuryFeb 9, 2026Fracture; High impedance

    IT WAS FURTHER REPORTED THAT THE RV LEAD WAS SUSPECTED TO BE FRACTURED. THE RV LEAD WAS CAPPED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24293845 ↗
  • MalfunctionJan 22, 2026High impedance; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A WARNING FOR HIGH/UNDEFINED BIPOLAR, RVTIP TO RVCOIL IMPEDANCE AND SENSING INTEGRITY COUNTER (SIC).  THE RV LEAD REMAINS IN USE.  NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24151868 ↗
  • InjuryJan 16, 2026Pacing Problem; Defibrillation/Stimulation Problem; Under-Sensing; High Capture Threshold

    IT WAS REPORTED THAT A STORED EPISODE NOTED RIGHT VENTRICULAR (RV) UNDERSENSING DURING ARRYTHMIA RESULTING IN DELAY OF VENTRICULAR T ACHYCARDIA (VT) EPISODE DETECTION, HIGH VOLTAGE THERAPY AND ANTI-TACHYCARDIA PACING (ATP). IT WAS ALSO NOTED THAT THERE WERE HIGH RV THRESHOLDS. THE RV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Tachycardia

    MAUDE report 24104120 ↗
  • MalfunctionDec 22, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED INTERMITTENT LOSS OF CAPTURE. ADDITIONALLY, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED POTENTIAL T-WAVE OVERSENSING (TWOS). THE LV AND RV LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23872477 ↗
  • InjuryDec 19, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED ALERTS FOR HIGH, OUT-OF-RANGE SUPERIOR VENA CAVA (SVC) DEFIBRILLATION IMPEDANCE. THE SVC COIL WAS PROGRAMMED OFF. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23850788 ↗
  • InjuryDec 18, 2025Failure to Capture; Fracture; High impedance; Failure to Capture; Fracture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A SUSPECTED LOW VOLTAGE FRACTURE. IT WAS ALSO NOTED THAT THE LEAD EXHIBITED HIGH IMPEDANCE AND NO CAPTURE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23843713 ↗
  • InjuryDec 15, 2025Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Appropriate Device Problem Term/Code Not Available; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS CAPPED AND REPLACED DUE TO AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Shock from Patient Lead(s)

    MAUDE report 23802494 ↗
  • InjuryDec 8, 2025Signal Artifact/Noise; Break; Collapse; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Break; Collapse; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS FURTHER REPORTED THAT THE LOW VOLTAGE PORTION OF THE LEAD WAS FRACTURED AND EXHIBITED SIGNS OF CLAVICULAR LEAD CRUSH.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 23745057 ↗
  • InjuryDec 1, 2025Over-Sensing; Pacing Problem; Over-Sensing; Pacing Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS PROGRAMMED TO EXTENDED BIPOLAR AND WAS SENSING ATRIAL ACTIVITY CAUSING OVERSENSING AND PAUSES. THE PATIENT WAS SYNCOPAL, SO A TEMPORARY PACING WIRE WAS IMPLANTED. A DEVICE CHANGEOUT PROCEDURE WAS DONE AND THE PARAMETERS ON THE LEAD WERE REPROGRAMED. THE LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Non specific EKG/ECG Changes; Syncope/Fainting

    MAUDE report 23687125 ↗
  • MalfunctionNov 24, 2025Over-Sensing; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23633608 ↗
  • MalfunctionNov 12, 2025High impedance; Over-Sensing; Low impedance; High impedance; Over-Sensing; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING DURING ARRHYTHMIA DETECTION, HIGH PACING IMPEDANCE, LOW DEFIBRILLATION IMPEDANCE, AND A LEAD INTEGRITY ALERT (LIA) WAS TRIGGERED DUE TO SENSING INTEGRITY COUNTER (SIC) AND HIGH-RATE NON-SUSTAINED EPISODES. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23533349 ↗
  • InjuryNov 11, 2025Impedance Problem; High Capture Threshold; Impedance Problem; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND RISING PACING IMPEDANCE. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23528344 ↗
  • MalfunctionNov 3, 2025High impedance; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH UNDEFINED PACING IMPEDANCE. THE LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING, TRIGGERING FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) WITH INAPPROPRIATE ATRIAL PACING AND IMPACTING MODE SWITCH. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23450488 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →