Divergence® Anterior Cervical Fusion System G6626527Recalls & reported problems

Medtronic Sofamor Danek, Inc. · IMPLANT G6626527 NP LORDO 7MMX15MMX12MM · Class 2 · Implant

All models of Divergence Anterior Cervical Fusion System (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169337435), received between Jun 23, 2017 and Mar 2, 2025 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture1
  • Mechanics Altered1
  • Migration or Expulsion of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • No Code Available1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 2, 2025Break; Mechanics Altered; Break; Mechanics Altered

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY FOR CERVICAL MYELOPATHY. IT WAS REPORTED THAT, IMPLANT ROTATED WHILE BEING IMPLANTED. UPON FURTHER SEATING WITH A TAMP, THE IMPLANT CRACKED. THERE WERE NO FRAGMENTS OF BROKEN CAGE REMAINS IN PATIENT. THE LEVELS PLANNED FOR IMPLANTATION WERE C6-C7 AND THE PROCEDURE OR TECHNIQUE PERFORMED WAS ANTERIOR CERVICAL DISCECTOMY AND FUSION. THERE WERE NO PATIENT SYMPTOMS REPORTED, NO ADDITIONAL TREATMENT OR SURGERY PERFORMED AS A RESULT. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21500930 ↗
  • MalfunctionFeb 25, 2025Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH A SPACER HAVING SPINAL THERAPY. IT WAS REPORTED THAT THE CAGE BROKE DURING ATTEMPTING TO IMPLANT. WHEN THE SURGEON WAS TAMPING THE CAGE IN PLACE, THE CAGE BROKE INTO MULTIPLE PIECES. SO ALL THE PIECES OF THE CAGE WERE EXPLANTED AND REPLACED WITH A NEW CAGE. THERE WAS A DELAY OF LESS THAN 60 MINS IN THE OVERALL PROCEDURE. PROCEDURE INVOLVED IN THIS EVENT WAS C6-7 ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF). THERE WAS NO REVISION SURGERY PLANNED/ OCCURRED AS A RESULT OF THE EVENT. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21456827 ↗
  • MalfunctionOct 7, 2024Fracture; Fracture

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT WHILE MALLETING A TAMP ONTO THE IMPLANT DURING IMPLANTATION, THE CAGE FRACTURED. THE SUPERIOR EDGE OF THE IMPLANT FRACTURED AND THE LOCKING MECHANISM CAME LOOSE AND WAS ABLE TO EXTRACT THE PIECES. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT EVENT OCCURRED DURING NORMAL USAGE OF IMPLANT. PROCEDURE INVOLVED IN THE EVENT WAS C6-7 ANTERIOR CERVICAL DISCECTOMY AND FUSION (ACDF).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20381665 ↗
  • MalfunctionSep 6, 2024Break; Break

    INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING SPINAL PRODUCT THAT WILL BE USED IN UNKNOWN SPINAL THERAPY FOR CERVICAL DISC HERNIATION. IT WAS REPORTED THAT DURING ANTERIOR CERVICAL FIXATION, A FRACTURE IN THE MEDIAL-LATERAL DIRECTION OF THE CAGE WAS OBSERVED AFTER SCREW INSERTION. THERE WERE NO BROKEN FRAGMENTS LEFT INSIDE THE PATIENT AND NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THE CRACK WAS FOUND AFTER THE CAGE WAS INSERTED. THE ISSUE OCCURRED DURING NORMAL USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20155057 ↗
  • InjuryJun 23, 2017Migration or Expulsion of Device; Migration or Expulsion of Device

    PRE-OP DIAGNOSIS: CERVICAL HERNIA. PROCEDURE: CERVICAL ANTERIOR FUSION IT WAS REPORTED THAT ON UNKNOWN DATE, POST-OP, THE IMPLANT GOT DEVIATED IN THE PATIENT'S BODY. A REVISION SURGERY WAS PERFORMED FOR FIXATION.

    Patient problems: No Code Available

    MAUDE report 6662687 ↗

Other Intervertebral fusion device with integrated fixation, cervical devices

All Intervertebral fusion device with integrated fixation, cervical devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page