Skip to content

N/A 3032028Recalls & reported problems

Medtronic Sofamor Danek, Inc. · DTS 3032028 ZEVO DISTAL 28 DEGREE · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169342385), received between Jul 14, 2025 and Jul 14, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Failure to Align” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Failure to Align1
  • Mechanical Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 14, 2025Mechanical Problem; Failure to Align; Mechanical Problem; Failure to Align

    INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING C4/5/6 OSSIFICA TION OF ANTERIOR LONGITUDINAL LIGAMENT AND CERVICAL ANTERIOR FUSION. IT WAS REPORTED THAT WHEN ATTEMPTING TO GRASP THE ZEVO PLATE USING THE DTS GUIDE, IT COULD NOT BE SECURELY HELD, AND THE LOOSENESS WAS SIGNIFICANT. THE PROTRUSION OF THE DTS GUIDE DID NOT FIT INTO THE SLOT LOCATED ON THE SIDE OF THE PLATE. THE SAME ISSUE OCCURRED EVEN WHEN USING THEZV DTS ADJUSTABLE CLAMP. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22491411 ↗