Advisa SR MRI™ SureScan™ A3SR01Recalls & reported problems
Medtronic, Inc. · IPG A3SR01 ADVISA SR MRI US MKT · Class 3 · Implant
All models of Advisa SR MRI SureScan (2)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Advisa SR MRI pacemaker has 229 adverse event reports, including 120 injuries and 16 deaths. The reports mostly describe infections around the device, pacing that falls below the set rate or slows the heartbeat, and batteries running down early. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 230
- Deaths reported
- 16
- Injuries reported
- 120
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 230 adverse event reports that list this exact device (DI 00643169402669), received between Feb 9, 2017 and Sep 16, 2026 — classified as 16 deaths, 120 injuries, 94 malfunctions.
The most frequently reported specific device problem is “Pacing Problem” (46 reports).
16 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem78
- Pacing Problem46
- Appropriate Device Problem Term/Code Not Available35
- Premature Discharge of Battery16
- Communication or Transmission Problem8
- Failure to Interrogate8
- Mechanical Problem8
- Migration or Expulsion of Device7
Patient problems
- Unspecified Infection59
- No Known Impact Or Consequence To Patient45
- No Clinical Signs, Symptoms or Conditions39
- Bradycardia25
- Death11
- Sepsis7
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available6
- Complaint, Ill-Defined6
Recent adverse event reports
The 25 most recent of 230 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 16, 2026Premature Discharge of Battery
It was reported that the Right Ventricular (RV) lead exhibited high thresholds leading to battery depletion of the Implantable Pulse Generator (IPG). The lead and IPG remain in use. No patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50787855 ↗InjuryAug 20, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient experienced a pocket infection with discomfort. The Implantable Pulse Generator (IPG) system was explanted a replaced with a Leadless implantable pulse generator (IPG). No further patient complications have been reported as a result of this event.
Patient problems: Discomfort; Unspecified Infection
MAUDE report 50342673 ↗MalfunctionJun 4, 2026Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA AND THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE LOWER RATE. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 25388144 ↗InjuryNov 28, 2025Pacing Problem; Pacing Problem
IT WAS FURTHER NOTED THAT THE RV LEAD AND IPG WERE EXPLANTED AND REPLACED.
Patient problems: Chest Pain; Dyspnea
MAUDE report 23670614 ↗InjuryNov 25, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT REPORTED THAT THEIR IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED DUE TO INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 23645757 ↗DeathNov 12, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS LEFT IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 23530967 ↗MalfunctionSep 4, 2025Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT THE PATIENTS REMOTE MONITOR DOES NOT ALLOW THEM TO COMPLETE THERE TRANSMISSION, AND EACH TIME IS ATTEMPTED IT SHOWS ERROR CODE. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22962521 ↗InjuryAug 29, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT WAS IN A SERIOUS ACCIDENT AND BELIEVED THEIR IMPLANTABLE PULSE GENERATOR (IPG) WAS NOT WORKING DURING THE ACCIDENT. IT WAS ALSO NOTED THE PATIENT PASSED OUT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 22921038 ↗DeathAug 22, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT DIED. IT WAS FURTHER NOTED THAT THE "DEVICE WAS NOT SATISFACTORY AND CONTRIBUTED TO THE PATIENTS PASSING". THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN PATIENT.
Patient problems: Insufficient Information
MAUDE report 22871379 ↗InjuryJul 3, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DEVELOPED A POCKET INFECTION WITH PURULENT DISCHARGE/PUS NOTED. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection; Purulent Discharge
MAUDE report 22392604 ↗InjuryJan 10, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED BY THE PATIENTS HUSBAND THAT THE PATIENT EXPERIENCED BRADYCARDIA AND THEIR IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE LOWER RATE. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 21122275 ↗MalfunctionJan 6, 2025Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A HEART RATE BELOW THE LOWER RATE OF THE IMPLANTABLE PULSE GENERATOR (IPG). THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 21077712 ↗InjuryDec 2, 2024Misconnection; Misconnection
IT WAS REPORTED THAT THE PATIENT EXPERIENCED ASYSTOLE AND SYNCOPE. IT WAS ALSO REPORTED THAT THE THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED INTERMITTENT LOSS OF CAPTURE AND WAS DROPPING BEATS. THE LEAD WAS REPROGRAMMED. RISING THRESHOLDS INTERMITTENT LOSS OF CAPTURE WAS STILL NOTED ON THE LEAD. DURING THE CHANGEOUT PROCEDURE, WHEN THE RV LEAD WAS TESTED, HIGH OUT OF RANGE IMPEDANCE WAS NOTED POSSIBLY DUE TO THE DEVICE EXHIBITING CONNECTOR PORT ISSUES. THE LEAD WAS CAPPED, AND DEVICE WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Asystole; No Clinical Signs, Symptoms or Conditions; Vascular Dissection
MAUDE report 20818255 ↗MalfunctionNov 12, 2024Failure to Interrogate; Failure to Interrogate
IT WAS REPORTED THAT ERROR CODE 3269 APPEARED IN THE APPLICATION INDICATING A BLUETOOTH FAILURE OCCURRED. IT WAS ALSO REPORTED THAT THERE WAS NO TELEMETRY WITH THE IMPLANTABLE PULSE GENERATOR (IPG). TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. REFERRED PATIENT TO THE CLINIC. THE SMART READER AND THE IPG REMAIN IN USE. THE SMART READER APPLICATION REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20665370 ↗MalfunctionOct 16, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED, THAT THERE WAS A CONCERN THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS NOT WORKING CORRECTLY. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20462450 ↗MalfunctionAug 27, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE LOWER LIMIT. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 20073842 ↗MalfunctionAug 23, 2024Communication or Transmission Problem; Communication or Transmission Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED LOSS OF TELEMETRY. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20048859 ↗InjuryJun 10, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT APPROXIMATELY THREE WEEKS POST IMPLANT, THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED DUE TO A POCKET INFECTION. PURULENT LIQUID WAS DISCHARGED WITH THE INITIAL EXCISION AND FOUND THROUGHOUT THE POCKET . THE INCISION WAS EXTENDED TO ALLOW EXPLORATION OF THE PUS AND PURULENT TISSUE WAS EXCISED. THE ORGANISM WAS IDENTIFIED AS BACTEREMIA. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Pocket Erosion; Purulent Discharge
MAUDE report 19496433 ↗MalfunctionMay 22, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN AND PACING BELOW THE SET LIMITS OF THE IPG WITH THE IPG NOT KICKING IN AT THE LOWER RATE. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Chest Pain
MAUDE report 19368764 ↗MalfunctionMar 18, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED WEAKNESS AND PACING BELOW THE SET LIMITS OF THE IMPLANTABLE PULSE GENERATOR (IPG). THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Muscle Weakness/Atrophy
MAUDE report 18923842 ↗InjuryMar 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN UNSPECIFIED INFECTION INVOLVING THE TRICUSPID VALVE. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED WITH A LEADLESS IPG. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 18901603 ↗InjuryFeb 14, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A METHICILLIN-SENSITIVE STAPHYLOCOCCUS AUREUS INFECTION WITH DISCOMFORT, REDNESS, TENDERNESS AND SWELLING ABOVE THE IMPLANTABLE PULSE GENERATOR (IPG) IMPLANT SITE. THE PATIENT ALSO EXPERIENCED ATRIAL FIBRILLATION (AF), HEART FAILURE WITH PRESERVED EJECTION FRACTION (HFPEF), HYPERTENSION, HYPERLIPIDEMIA, GOUT, MSSA BACTEREMIA COMPLICATED BY EPIDURAL ABSCESS, SPLENIC ABSCESS, L-S SPINE OSTEOMYELITIS AND PSOAS ABSCESS. THE IPG SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Abscess; Atrial Fibrillation; Bacteremia; Bacterial Infection; Discomfort; Erythema; Heart Failure/Congestive Heart Failure; High Blood Pressure/ Hypertension; Swelling/ Edema
MAUDE report 18701393 ↗MalfunctionFeb 12, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT STATED THE IMPLANTABLE PULSE GENERATOR (IPG) IS "TURNING ON AND OFF." THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18684028 ↗InjuryJan 18, 2024Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE PATIENT FELT LIKE PACEMAKER WAS FUNNY AND THEY HAD A SYNCOPAL EPISODE AND FALL AT HOME. THE CARELINK TRANSMISSION INDICATED THERE WAS A DECREASE IN PACING PERCENTAGE FOR THE PAST MONTH, UNKNOWN IF RELATED TO PATIENT SYMPTOMS. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 18533383 ↗DeathJan 8, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS DECEASED. THERE WAS AN ALLEGATION FROM THE PATIENT'S FAMILY MEMBER THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS RELATED TO THE PATIENT'S DEATH. A CAUSE OF DEATH WAS REQUESTED BUT WAS NOT RECEIVED.
Patient problems: Insufficient Information
MAUDE report 18469223 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-0806-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ADAPTA, Dual chamber pacemaker (DDD): (a) Model Number ADD01 (b) Model Number ADDR01 (c) Model Number ADDR03 (d) Model Number ADDR06 (e) Mo
- Z-0811-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
…and 1 more.
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page