MiniMed® 530G MMT-551NAH (UDI 00643169404267)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-551NAH PRDGM 1.8ML V1.1F PK EN · Class 3
All models of MiniMed 530G (71)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic MiniMed insulin pump is an insulin pump system. Of its 163 adverse event reports, 58 are injuries and 104 are malfunctions, with 1 death. Reports describe incorrect messages, motor and battery alarms, and cracked or broken parts. Reports also mention high blood sugar (hyperglycemia) and low blood sugar (hypoglycemia). No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 163
- Deaths reported
- 1
- Injuries reported
- 58
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 163 adverse event reports that list this exact device (DI 00643169404267), received between Jun 1, 2017 and Aug 7, 2020 — classified as 1 death, 58 injuries, 104 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (32 reports).
2 other recalls cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message32
- Patient Device Interaction Problem21
- Break20
- Adverse Event Without Identified Device or Use Problem18
- Mechanical Problem14
- Crack13
- Device Difficult to Program or Calibrate11
- No Display/Image11
Patient problems
- No Consequences Or Impact To Patient84
- Hyperglycemia60
- Hypoglycemia14
- No Known Impact Or Consequence To Patient4
- Blood Loss2
- Death1
- Elevated Ketones/ Diabetic Ketoacidosis1
- Hematoma1
Recent adverse event reports
The 25 most recent of 163 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2020Power Problem; Patient Device Interaction Problem; Power Problem; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP RECEIVED A BATTERY OUT LIMIT ALARM. THE CUSTOMER STATED THEY WERE ABLE TO CLEAR THE ALARM AND DID NOT RECEIVE REQUEST TO COMPLETE THE REWIND PROCESS. IT WAS ADVISED THAT THE CAUSE OF THE ALARM WAS DUE TO A DAMAGED BATTERY CAP. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10377505 ↗InjuryAug 3, 2020Visual Prompts will not Clear; Visual Prompts will not Clear
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED LOW BLOOD GLUCOSE AND HIGH BLOOD GLUCOSE. THE CUSTOMER LOW BLOOD GLUCOSE LEVEL WAS 42 MG/DL AND HIGH BLOOD GLUCOSE LEVEL WAS 320 MG/DL. CUSTOMER'S CURRENT BLOOD GLUCOSE LEVEL WAS 193 MG/DL. CUSTOMER STATED THAT THEY INSTALLING NEW BATTERY, MONITORING INSULIN PUMP FOR 2 HOURS AND FROZEN DISPLAY RECURRED. CUSTOMER WAS USING INSULIN PUMP SYSTEM WITHIN 48 HOURS OF REPORTED HIGH BLOOD GLUCOSE EVENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING AND DECLINED FOR LOW BLOOD GLUCOSE AND HIGH BLOOD GLUCOSE. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia; Hypoglycemia
MAUDE report 10355731 ↗InjuryApr 9, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
THE CUSTOMER¿S MOTHER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE OF OVER 400 MG/DL. THE CUSTOMER TREATED WITH THE INSULIN PUMP. TROUBLESHOOTING WAS DECLINED FOR HIGH BLOOD GLUCOSE. THE DRIVE SUPPORT CAP APPEARS NORMAL. THE CUSTOMER WAS NEITHER IN THE EMERGENCY ROOM, NOR ADMITTED INTO HOSPITAL AS A RESULT OF HIGH BLOOD GLUCOSE. BASED ON CUSTOMER REPORT CUSTOMER DOES NOT ALLEGE PUMP WAS UNDER DELIVERING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 9942619 ↗MalfunctionApr 4, 2020Obstruction of Flow; Obstruction of Flow
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A STUCK IN NO DELIVERY WHEN TRYING TO CHANGE THE INFUSION SET. CUSTOMER STATED THAT THE INSULIN DID EXIT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9925964 ↗MalfunctionMar 30, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM. INSULIN PUMP HISTORY SHOWED MOTION SENSOR TEST FAILURE ALARM. THE CUSTOMER WAS ABLE TO CLEAR THE ALARM AND ABLE TO REWIND INSULIN PUMP AS WELL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9896726 ↗MalfunctionFeb 12, 2020Premature Discharge of Battery; Mechanical Problem; Premature Discharge of Battery; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR AND UNEXPECTED RESTART ALARM. CUSTOMER REPORTED THEY WERE ABLE TO CLEAR THE BOTH ALARMS. CUSTOMER STATED THEY ARE UNABLE TO REWIND THE INSULIN PUMP. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9703615 ↗MalfunctionJan 27, 2020Display or Visual Feedback Problem; Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem; Display or Visual Feedback Problem; Mechanical Problem; Obstruction of Flow; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP RECEIVED MOTOR ERROR ALARM, NO DELIVERY ALARM AND CRACKED AROUND THE RESERVOIR CASING. CUSTOMER STATED THAT THE RESERVOIR WAS LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. CUSTOMER ALSO STATED THAT BACK LIGHT DID NOT WORK. CUSTOMER STATED THE INSULIN PUMP HAD COSMETIC DAMAGE. CUSTOMER STATED THE INFUSION SET AND RESERVOIR DID NOT SHOW SIGNS OF DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9630835 ↗InjuryJan 26, 2020Mechanical Problem; Obstruction of Flow; No Apparent Adverse Event; Mechanical Problem; Obstruction of Flow; No Apparent Adverse Event
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE HOSPITALIZED DUE TO UNKNOWN REASON ON AN UNKNOWN DATE. CUSTOMER¿S BLOOD GLUCOSE READING WAS UNKNOWN. THE CUSTOMER STATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARMS AND WOULD NOT DELIVER INSULIN AS WELL AS UNEXPLAINED NO DELIVERY ALARMS. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 9630595 ↗MalfunctionDec 6, 2019Patient Device Interaction Problem; Priming Problem; Patient Device Interaction Problem; Priming Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD SHOT INSULIN DURING PRIMING PROCESS. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. CUSTOMER ALSO REPORTED THAT THE RESERVOIR CASE WAS CHIPPED AND HAD SCRATCHES ON SCREEN. CUSTOMER DECLINED HELPLINE TRANSFER. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9421283 ↗MalfunctionNov 25, 2019Mechanical Problem; Mechanical Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP RECEIVED THE MOTOR POSITION ENCODED ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 149 MG/DL. CUSTOMER STATED THAT FORGOT TO REMOVE HER PUMP WHILE UNDERGOING MRI. CUSTOMER STATED THE DRIVE SUPPORT CAP APPEARS NORMAL. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9367804 ↗MalfunctionNov 20, 2019Mechanical Problem; Device Difficult to Program or Calibrate; Power Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Power Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE BUTTONS WERE LESS RESPONSIVE, SOMETIMES ACT BUTTON HAD TO PRESS TWICE. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ALSO REPORTED THAT INSULIN PUMP HAD REJECTED NEW BATTERY, SLOWER DURING REWIND. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9346754 ↗MalfunctionNov 18, 2019Premature Discharge of Battery; Premature Discharge of Battery
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD UNEXPECTED RESTART ALARM. THE CUSTOMER¿S BLOOD GLUCOSE WAS 106 MG/DL. THE TROUBLESHOOTING WAS PERFORMED. CUSTOMER REPORTED THAT THEY WERE ABLE TO CLEAR THE ALARM. CUSTOMER WAS ABLE TO COMPLETE INSULIN PUMP REWIND. THE CUSTOMER STATED THAT THE ALARM WAS OCCURRED SHORTLY AFTER INSERTING A NEW BATTERY. THE CUSTOMER WAS ADVISED THAT THE INSULIN PUMP WILL NEED TO BE REPLACED AND DISCONTINUE USE OF THE INSULIN PUMP AND REVERT TO BACKUP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9339479 ↗MalfunctionNov 11, 2019Device Difficult to Program or Calibrate; Visual Prompts will not Clear; Device Difficult to Program or Calibrate; Visual Prompts will not Clear
THE CUSTOMER¿S PARENT REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER¿S BLOOD GLUCOSE WAS 60 MG/DL AT THE TIME OF INCIDENT. CUSTOMER¿S PARENT STATED THAT THE INSULIN PUMP ALARMED BUTTON ERROR AND THE BUTTONS WERE UNRESPONSIVE. NO SIGNIFICANT EVENTS LEADING TO BUTTON ERROR WERE OBSERVED. BUTTON ERROR TROUBLESHOOTING WAS PERFORMED AND CONFIRMED THAT TIME IS ADVANCING. THE CUSTOMER¿S PARENT WAS ADVISED TO HAVE THE CUSTOMER DISCONTINUE USE OF PUMP AND REVERT TO BACK-UP PLAN. THE INSULIN PUMP IS BEING REPLACED AND IS EXPECTED TO RETURN FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9303555 ↗MalfunctionNov 7, 2019Communication or Transmission Problem; Communication or Transmission Problem
THE CUSTOMER INQUIRED VIA PHONE CALL ON HOW TO DETERMINE SENSITIVITY. CUSTOMER WAS ADVISED TO CONSULT THEIR HEALTHCARE PROVIDER ABOUT SENSITIVITY. CUSTOMER'S BLOOD GLUCOSE VALUE WAS NOT PROVIDED. CUSTOMER HAS RECEIVED A NEW GENERATION PUMP. THE CUSTOMER RETURNED THE OLD DEVICE FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9291571 ↗MalfunctionOct 12, 2019Patient Device Interaction Problem; Patient Device Interaction Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THE PART WHERE THE RESERVOIR SCREWS IN OF INSULIN PUMP HAD CHUNKS AND COMING OFF FROM LIP RING AREA. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. CUSTOMER STATED THAT THE RESERVOIR ABLE TO LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9183567 ↗MalfunctionSep 24, 2019Mechanical Problem; Mechanical Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD MOTOR ERROR. CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL. CUSTOMER WAS ABLE TO CLEAR THE ALARM. CUSTOMER WAS UNABLE TO COMPLETE INSULIN PUMP REWIND DUE TO INSULIN PUMP ALARM. TROUBLESHOOTING WAS PERFORMED. CUSTOMER WAS ADVISED THAT THE INSULIN PUMP NEEDED TO BE REPLACED AND TO DISCONTINUE USE OF THE INSULIN PUMP AND REFER TO A BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9108045 ↗InjurySep 5, 2019Mechanical Problem; Detachment of Device or Device Component; No Apparent Adverse Event; Mechanical Problem; Detachment of Device or Device Component; No Apparent Adverse Event
THE CUSTOMER¿S FATHER REPORTED VIA PHONE CALL THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA; THE INSULIN PUMP HAD A MOTOR ERROR ALARM, AND THE DRIVE SUPPORT CAP WAS PUSHING OUT THE END OF THE INSULIN PUMP. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 400 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS ADVISED TO ENSURE THEY ARE DISCONTINUED FROM THE INSULIN PUMP AND FOLLOW THEIR MEDICALLY PRESCRIBED BACK UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 8963521 ↗MalfunctionJul 22, 2019Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Material Integrity Problem; Patient Device Interaction Problem; Priming Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Material Integrity Problem; Patient Device Interaction Problem; Priming Problem
THE CUSTOMER'S PARENT REPORTED VIA PHONE CALL THAT THE BUTTONS WERE LESS RESPONSIVE, SCRATCHES ON DISPLAY OF INSULIN PUMP AFFECTED ABILITY TO READ THE SCREEN. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT IN CASING AT BATTERY CAP AND RESERVOIR AREA OF INSULIN PUMP HAD CRACK, SLOWER DURING PRIME, RANDOM UNEXPLAINED NO DELIVERY ALERTS, SLOWER DURING REWIND, DIMINISHED BATTERY LIFE. THE CUSTOMER WAS ADVISED TO REPLACE AND TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 8810617 ↗MalfunctionJul 18, 2019Device Difficult to Program or Calibrate; Obstruction of Flow; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD KEYPAD ANOMALY AND FAILED BATTERY. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER REPORTED THAT THE INSULIN PUMP REJECTED NEW BATTERY, BUTTONS WERE LESS RESPONSIVE AND ALSO HAD RANDOM NO DELIVERY ALARM. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8801326 ↗MalfunctionJul 9, 2019Device Difficult to Program or Calibrate; Visual Prompts will not Clear; Device Difficult to Program or Calibrate; Visual Prompts will not Clear
THE CUSTOMER REPORTED VIA PHONE CALL THAT HER INSULIN PUMP HAD A BUTTON ERROR AND SHE COULD NOT CLEAR IT. THE CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 130 MG/DL. THE CUSTOMER WAS ASSISTED IN TROUBLESHOOTING. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8770280 ↗InjuryJul 4, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL OF 505, 469 AND 493 MG/DL. THE CUSTOMER DID NOT PROVIDE THE SYMPTOMS REGARDING HIGH BLOOD GLUCOSE. THE CUSTOMER WAS TREATED FOR THE HIGH BLOOD GLUCOSE WITH MANUAL INJECTIONS. THE CUSTOMER DECLINED TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE LEVEL. THE INSULIN PUMP WAS NOT RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 8762983 ↗MalfunctionJun 20, 2019Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A KEYPAD ANOMALY. TROUBLESHOOTING FOR BUTTON ERROR/KEYPAD ANOMALY WAS PERFORMED. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE ACT BUTTON WAS UNRESPONSIVE. THE CUSTOMER ALSO STATED THAT THERE WAS NO SIGNIFICANT EVENT LEADING TO THE KEYPAD ISSUES. THE CUSTOMER DECLINED TO CONTINUE TROUBLESHOOTING AND ADVISED THAT THEY WILL JUST MONITOR THE INSULIN PUMP. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8715439 ↗MalfunctionJun 14, 2019Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Power Problem; Insufficient Device Problem Information; Patient Device Interaction Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Power Problem; Insufficient Device Problem Information; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT BUTTONS WERE SLOW TO RESPOND TO COMMANDS AND INSULIN PUMP HAD CODE ERRORS. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE MOTOR SOUND DURING REWIND, CHANGING BATTERIES EVERY 2 WEEKS AND OFTEN REJECTS NEW BATTERIES. THE CUSTOMER ALSO STATED THAT INSULIN PUMP HAD NO DELIVERY ALARM AND LARGE CRACKS. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8698962 ↗MalfunctionJun 13, 2019No Display/Image; Patient Device Interaction Problem; No Display/Image; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD A BLANK DISPLAY AND CRACK ON THE RESERVOIR COMPARTMENT. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 220 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER WAS NOT ASSISTED WITH TROUBLESHOOTING. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8695066 ↗MalfunctionMay 27, 2019Patient Device Interaction Problem; Patient Device Interaction Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE HALF OF RESERVOIR OPENING WAS BROKEN. THE CUSTOMER¿S BLOOD GLUCOSE WAS UNKNOWN. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8643938 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.
- Z-1751-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Artificial pancreas device system, threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page