t:slim X2 Insulin Pump with Basal-IQ 1000912Recalls & reported problems
Tandem Diabetes Care, Inc. · t:slim X2 Insulin Pump with Basal-IQ, Refurbished, 6.0 · Class 3
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This is a refurbished Tandem t:slim X2 insulin pump with Basal-IQ. Its reports mostly describe pump stoppages, charging trouble, cracked or unresponsive screens and wrong messages. There are 30,220 reports, including 2 deaths and 2,720 injuries. One recall, started in May 2022, is open: memory faults, an inaccurate battery display and a touchscreen that stays on. Reports dropped to 41 in the last 12 months, from 162 before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30,219
- Deaths reported
- 2
- Injuries reported
- 2,720
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 30,219 adverse event reports that list this exact device (DI 00853052007325), received between Nov 8, 2018 and Jul 7, 2026 — classified as 2 deaths, 2,720 injuries, 27,497 malfunctions.
The most frequently reported specific device problem is “Pumping Stopped” (6,484 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- May 24, 2022Open, Classified
Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.
Z-1811-2022 · Root cause: Device Design · FDA record ↗
Most reported problems
Device problems
- Pumping Stopped6,484
- Charging Problem2,983
- Incorrect Measurement2,956
- Device Displays Incorrect Message2,911
- Obstruction of Flow2,906
- Failure to Sense2,535
- Battery Problem2,432
- Adverse Event Without Identified Device or Use Problem1,719
Patient problems
- No Clinical Signs, Symptoms or Conditions12,560
- No Consequences Or Impact To Patient10,813
- No Known Impact Or Consequence To Patient4,197
- Hyperglycemia1,455
- Hypoglycemia1,106
- Elevated Ketones/ Diabetic Ketoacidosis302
- Insufficient Information42
- No Patient Involvement39
Recent adverse event reports
The 25 most recent of 30,219 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 7, 2026Fracture
IT WAS REPORTED THAT THE PUMP TOUCH SCREEN WAS CRACKED AND UNRESPONSIVE AND UNREADABLE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD IN INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25736690 ↗MalfunctionJul 7, 2026False Alarm
IT WAS REPORTED THAT AN ALTITUDE ALARM OCCURRED WHILE THE CUSTOMER WAS NOT OUTSIDE OF LABELED OPERATING ALTITUDE RANGE. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER CLEANED THE SPEAKER HOLES AND CLEARED THE ALARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25735039 ↗MalfunctionJul 1, 2026Break; Charging Problem
IT WAS REPORTED THAT INSULIN PUMP WAS DAMAGED AROUND CHARGING PORT, WHICH CAUSED INTERMITTENT TROUBLE WITH CHARGING. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED FURTHER FOLLOW UP, WAS OUT OF WARRANTY, AND WAS IN PROCESS OF PUMP RENEWAL, AND NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING OUTCOME OF EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25677606 ↗MalfunctionJun 2, 2026Pumping Stopped
IT WAS REPORTED THAT AN UNSPECIFIED MALFUNCTION ALARM OCCURRED. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION COULD BE OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25358680 ↗MalfunctionMay 28, 2026Fracture
IT WAS REPORTED THAT PUMP SCREEN WAS CRACKED AND CONSIDERED A MALFUNCTION, AND CUSTOMER DECLINED TO SPEAK WITH CUSTOMER TECHNICAL SUPPORT. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS RECEIVED REGARDING ANY ACTIONS TAKEN BY THE CUSTOMER OR TECHNICAL SUPPORT, AND NO FURTHER OUTCOME DETAILS WERE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25322267 ↗MalfunctionMay 26, 2026Program or Algorithm Execution Problem
IT WAS REPORTED THAT THE PUMP SOFTWARE DID NOT ACCURATELY ADJUST THE AMOUNT OF INSULIN DELIVERED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25284237 ↗MalfunctionMay 26, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25280709 ↗MalfunctionMay 26, 2026Key or Button Unresponsive/not Working
IT WAS REPORTED THAT THE WAKE BUTTON ON THE PUMP WAS UNRESPONSIVE INTERMITTENTLY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED TO TROUBLESHOOT THE ISSUE WITH TANDEM TECHNICAL SUPPORT, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25280365 ↗MalfunctionMay 22, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25257470 ↗MalfunctionMay 5, 2026Failure to Charge
IT WAS REPORTED THAT THE CUSTOMER WAS UNABLE TO CHARGE THE PUMP USING THE TANDEM-PROVIDED USB CHARGING CABLE AND WALL POWER ADAPTER. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER WAS ABLE TO SUCCESSFULLY CHARGE THE PUMP USING A SECONDARY USB CABLE AND WALL POWER ADAPTER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25081317 ↗MalfunctionApr 28, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP EXPERIENCED MALFUNCTION CODE 9 WITH NO NOTABLE EVENTS OCCURRING BEFOREHAND. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER CONTACTED TECHNICAL SUPPORT, RECEIVED INSTRUCTIONS TO RESET PUMP, SUCCESSFULLY RESET MALFUNCTION WITHOUT RECURRENCE, AND WAS ADVISED TO CONTINUE USING PUMP AND TO CALL BACK IF MALFUNCTION CODE APPEARED AGAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25015327 ↗InjuryApr 21, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CUSTOMER EXPERIENCED AN ELEVATED BLOOD GLUCOSE (BG) LEVEL OF 600 MG/DL. CAUSE WAS NOT KNOWN. CUSTOMER CHANGED SUPPLIES TO ADDRESS BG. RECOMMENDATION WAS MADE TO CONSULT WITH A HEALTHCARE PROVIDER REGARDING DIABETES MANAGEMENT.
Patient problems: Hyperglycemia
MAUDE report 24941890 ↗MalfunctionApr 13, 2026Failure to Charge
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A POWER SOURCE ALERT AFTER PLUGGING THE PUMP INTO A WALL OUTLET. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. MULTIPLE CONTACT ATTEMPTS WERE MADE BY TANDEM TECHNICAL SUPPORT TO OBTAIN ADDITIONAL INFORMATION REGARDING THE ISSUE; HOWEVER, NO RESPONSE WAS RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24856040 ↗MalfunctionApr 6, 2026Pumping Stopped
IT WAS REPORTED THAT PUMP MALFUNCTION OCCURRED INTERMITTENTLY. CUSTOMER¿S BLOOD GLUCOSE (BG) LEVEL WAS DISPLAYED AS "HIGH"; HOWEVER, SPECIFIC VALUE WAS NOT PROVIDED. ELEVATED BG WAS ADDRESS BY A MANUAL INSULIN INJECTION AND DID NOT REQUIRE ASSISTANCE FROM A THIRD PARTY. REPORTEDLY, THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY.
Patient problems: Hyperglycemia
MAUDE report 24792022 ↗MalfunctionFeb 6, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER DECLINED ASSISTANCE FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE. NO ADDITIONAL PATIENT OR EVENT INFORMATION WAS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24279012 ↗MalfunctionFeb 6, 2026Premature Discharge of Battery
CUSTOMER REPORTED BATTERY DEPLETED TOO FAST. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. CUSTOMER DECLINED FOLLOW-UP OR TO PROVIDE ADDITIONAL DETAILS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24278648 ↗MalfunctionFeb 5, 2026Failure to Charge
IT WAS REPORTED THAT THE PUMP BATTERY COULD NOT BE CHARGED. THERE WAS NO REPORTED IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. NO ADDITIONAL INFORMATION WAS PROVIDED AND THE CUSTOMER DECLINED TROUBLESHOOTING WITH TANDEM TECHNICAL SUPPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24273388 ↗MalfunctionFeb 5, 2026Tactile Prompts/Feedback
IT WAS REPORTED THAT THE PUMP WAS NOT VIBRATING WHEN ALERTS AND ALARMS WERE DECLARED. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL. REPORTEDLY, THE CUSTOMER CONTINUED USING THE PUMP FOR INSULIN THERAPY UNTIL THE REPLACEMENT PUMP ARRIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24268646 ↗MalfunctionFeb 3, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THE PUMP BATTERY WAS DEPLETING QUICKLY. THERE WAS NO REPORTED ADVERSE IMPACT TO THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. THE CUSTOMER CONTINUED TO USE THE PUMP FOR INSULIN THERAPY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24251491 ↗MalfunctionJan 28, 2026Display or Visual Feedback Problem
IT WAS REPORTED THAT THE PUMP TOUCHSCREEN WAS ON, BUT THE DISPLAY REMAINED BLACK. THE CUSTOMER REVERTED TO AN ALTERNATE METHOD OF INSULIN THERAPY. THERE WAS NO ADVERSE IMPACT TO THE CUSTOMER'S BLOOD GLUCOSE LEVEL.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24202682 ↗MalfunctionJan 12, 2026Pumping Stopped
CUSTOMER REPORTED AN ISSUE WITH THE PUMP INDICATING A CARTRIDGE ALARM 20, BUT THERE WAS NO ADVERSE IMPACT ON THE CUSTOMER¿S BLOOD GLUCOSE LEVEL. AS THE PUMP WAS OUT OF WARRANTY, TECHNICAL SUPPORT DECIDED TO REPLACE IT AND ADVISED THE CUSTOMER TO USE MDI AS BACKUP THERAPY. THE CUSTOMER WAS INFORMED THAT THEY WOULD RECEIVE A REPLACEMENT PUMP WITH UPDATED SOFTWARE AND INSTRUCTED TO PROGRAM THEIR SETTINGS AND CONNECT THEIR CGM TO THE NEW PUMP. CUSTOMER AGREED TO SEND INSURANCE CARD PICTURES, AND ARRANGEMENTS WERE MADE TO PROVIDE AN OUT-OF-WARRANTY LOANER PUMP UNTIL THE REPLACEMENT ARRIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24036752 ↗InjuryJan 12, 2026Mechanical Problem
A CARTRIDGE ALARM WAS REPORTED BY THE CUSTOMER DURING THE LOAD PROCESS, WITH THE BLOOD GLUCOSE VALUE DISPLAYED AS "HIGH," ALTHOUGH THE SPECIFIC VALUE WAS UNKNOWN. THE CUSTOMER MANAGED THE HIGH BLOOD GLUCOSE BY ADMINISTERING A CORRECTION INJECTION USING A MULTIPLE DAILY INJECTION METHOD. THE CALLER WAS EDUCATED TO WAIT UNTIL PROMPTED DURING THE LOAD PROCESS TO REMOVE THE USED CARTRIDGE, WHICH THEY UNDERSTOOD AND ACKNOWLEDGED. TECHNICAL SUPPORT RESOLVED TO REPLACE THE PUMP, EVEN THOUGH IT WAS OUT OF WARRANTY, BUT THE CUSTOMER DECLINED THE OFFER FOR AN OUT-OF-WARRANTY LOANER PUMP.
Patient problems: Hyperglycemia
MAUDE report 24035874 ↗MalfunctionJan 8, 2026Pumping Stopped
CALLER REPORTED THAT PUMP EXPERIENCED A MALFUNCTION. THERE WAS NO REPORTED IMPACT ON THE CUSTOMER'S BLOOD GLUCOSE LEVEL AS IT WAS UNCLEAR OR NOT PROVIDED. CUSTOMER RECEIVED ASSISTANCE FROM TECHNICAL SUPPORT TO RESET THE PUMP, AFTER WHICH THE MALFUNCTION DID NOT RECUR. CALLER WAS ADVISED TO CONTINUE USING THE PUMP AND TO CALL BACK IF ANY FURTHER MALFUNCTION CODES APPEARED, WHICH THEY ACKNOWLEDGED AND UNDERSTOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24012710 ↗InjuryJan 8, 2026Adverse Event Without Identified Device or Use Problem
A CUSTOMER REPORTED EXPERIENCING HOSPITALIZATION DUE TO A SEIZURE CAUSED BY HYPOGLYCEMIA. THE CUSTOMER DECLINED FOLLOW-UP FROM TANDEM TECHNICAL SUPPORT REGARDING THE ISSUE, THEREFORE NO ADDITIONAL INFORMATION WAS OBTAINED.
Patient problems: Convulsion/Seizure; Hypoglycemia
MAUDE report 24003849 ↗MalfunctionJan 2, 2026Pumping Stopped
IT WAS REPORTED THAT A MALFUNCTION ALARM OCCURRED. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 435 MG/DL. DURING TROUBLESHOOTING WITH TECHNICAL SUPPORT, THE MALFUNCTION ALARM WAS CLEARED, AND INSULIN DELIVERY WAS RESUMED SUCCESSFULLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23943704 ↗
Other Automated Insulin Dosing , Threshold Suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000892Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall1 death5,216 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.