CapSure Sense MRI™ SureScan™ 457445 (UDI 00643169410916)Recalls & reported problems

Medtronic, Inc. · LEAD 457445 MRI US BI RCMCRD MVC · Class 3 · Implant

All models of CapSure Sense MRI SureScan (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic CapSure Sense MRI pacemaker lead has 1,400 FDA adverse event reports, with 594 injuries, 790 malfunctions and 16 deaths. The reports mostly describe the lead missing heart signals, needing high energy to trigger a heartbeat, and moving out of place. Some mention infections or chest pain. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,416
Deaths reported
16
Injuries reported
600
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,416 adverse event reports that list this exact device (DI 00643169410916), received between Aug 9, 2017 and Sep 30, 2026 — classified as 16 deaths, 600 injuries, 800 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (610 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 29 reportsDec 2023: 16 reportsJan 2024: 23 reportsFeb 2024: 13 reportsMar 2024: 21 reportsApr 2024: 24 reportsMay 2024: 15 reportsJun 2024: 18 reportsJul 2024: 14 reportsAug 2024: 12 reportsSep 2024: 7 reportsOct 2024: 21 reportsNov 2024: 29 reportsDec 2024: 9 reportsJan 2025: 21 reportsFeb 2025: 15 reportsMar 2025: 14 reportsApr 2025: 13 reportsMay 2025: 18 reportsJun 2025: 16 reportsJul 2025: 26 reportsAug 2025: 15 reportsSep 2025: 15 reportsOct 2025: 9 reportsNov 2025: 13 reportsDec 2025: 12 reportsJan 2026: 14 reportsFeb 2026: 17 reportsMar 2026: 21 reportsApr 2026: 15 reportsMay 2026: 22 reportsJun 2026: 13 reportsJul 2026: 11 reportsAug 2026: 14 reportsSep 2026: 16 reportsOct 2026: 0 reports20242025202629
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Under-Sensing610
  • High Capture Threshold310
  • Device Dislodged or Dislocated224
  • Adverse Event Without Identified Device or Use Problem162
  • Pacing Problem147
  • Over-Sensing138
  • Failure to Capture124
  • Decreased Sensitivity79

Patient problems

  • No Clinical Signs, Symptoms or Conditions896
  • No Known Impact Or Consequence To Patient159
  • Unspecified Infection118
  • No Consequences Or Impact To Patient40
  • Chest Pain24
  • Dyspnea24
  • Dizziness23
  • Bradycardia18

Recent adverse event reports

The 25 most recent of 1,416 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed a pocket infection. The Implantable Pulse Generator (IPG) sysetm was explanted and replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50945880 ↗
  • MalfunctionSep 30, 2026Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) episodes resulting in termination of AT/AF detected event(s) in progress. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50945078 ↗
  • MalfunctionSep 28, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Ventricular (RV) lead triggered multiple alerts for high RV defibrillation impedance. The Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times. The leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50921549 ↗
  • MalfunctionSep 28, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50917053 ↗
  • InjurySep 25, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced an infection. The Implantable Pulse Generator (IPG) system was removed and replaced with a leadless IPG. No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50895766 ↗
  • MalfunctionSep 18, 2026High Capture Threshold

    It was reported that the Right Atrial (RA) lead exhibited high thresholds. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50824497 ↗
  • MalfunctionSep 10, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722502 ↗
  • MalfunctionSep 9, 2026Failure to Capture; High Capture Threshold

    It was reported that the Right Atrial (RA) lead exhibited loss of capture and high thresholds. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50689444 ↗
  • InjurySep 8, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the patient experienced pneoumothorax. The leads remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: Pneumothorax

    MAUDE report 50666410 ↗
  • MalfunctionSep 8, 2026Failure to Capture; Over-Sensing

    It was reported that the Right Atrial (RA) lead exhibited over sensing and loss of capture on stored episodes. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50666169 ↗
  • MalfunctionSep 3, 2026Electrical /Electronic Property Problem; High Capture Threshold

    It was reported that approximately six weeks post implant the Right Atrial (RA) lead exhibited an Atrial Lead Position Check failure and high thresholds. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50586221 ↗
  • MalfunctionSep 2, 2026Pacing Problem; Under-Sensing

    It was also reported that the RA lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AF) events resulting in unnecessary pacing at times.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50567704 ↗
  • InjurySep 2, 2026Appropriate Device Problem Term/Code Not Available

    It was reported that the patient's transvenous Implantable Pulse Generator (IPG) system was explanted as it was causing them discomfort. A leadless IPG was implanted instead. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort

    MAUDE report 50562048 ↗
  • MalfunctionSep 2, 2026Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50558271 ↗
  • MalfunctionSep 1, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50541108 ↗
  • MalfunctionSep 1, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and termination of AT/AF detected event(s) in progress. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50538669 ↗
  • MalfunctionAug 31, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50518272 ↗
  • InjuryAug 28, 2026Under-Sensing; Decreased Sensitivity

    It was reported that the Right Atrial (RA) lead exhibited undersensing and diminished sensing. The RA lead was capped and replaced. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50493305 ↗
  • MalfunctionAug 27, 2026High Capture Threshold

    It was reported that the remote monitoring network transmission was not exported to the monitoring inbox for review. A ticket was created and the issue was under investigation. It was further reported that the Right Atrial (RA) lead displayed high capture thresholds. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50479178 ↗
  • MalfunctionAug 27, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times and resulting in termination of AT/AF detected event(s) in progress. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50468401 ↗
  • MalfunctionAug 26, 2026Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing during Atrial Tachycardia/Atrial Fibrillation (AT/AF) stored episodes resulting in termination of AT/AF detected event(s) in progress. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50453724 ↗
  • MalfunctionAug 26, 2026Signal Artifact/Noise; Fracture; High impedance

    It was reported that the Right Atrial (RA) lead exhibited noise, high impedance and was fractured. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50451258 ↗
  • MalfunctionAug 21, 2026Capturing Problem; Device Dislodged or Dislocated; Impedance Problem

    It was reported that the Right Atrial (RA) lead had exhibited an slow and steady increase in RA pacing impedance and rise in threshold. The lead was stated to have been dislodged. It was stated that the patient had a fall from their dental chair last month but was not sure if the fall was enough to impact a dislodgement of the lead.  The RA impedance alert tone was turn off  and alert set to higher limit. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: Fall

    MAUDE report 50369500 ↗
  • MalfunctionAug 20, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Atrial (RA) lead exhibited undersensing on stored electrograms (EGM) during Atrial Tachycardia/Atrial Fibrillation (AT/AF) events resulting in unnecessary pacing at times. The RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50340255 ↗
  • MalfunctionAug 18, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50297110 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources and updates