CapSure Sense MRI™ SureScan™ 457445 (UDI 00643169410954)Recalls & reported problems
Medtronic, Inc. · LEAD 457445 MRI US BI RCMCRD · Class 3 · Implant
All models of CapSure Sense MRI SureScan (8)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 15
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 00643169410954), received between Apr 10, 2018 and Mar 27, 2024 — classified as 6 injuries, 9 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Under-Sensing6
- Device Dislodged or Dislocated5
- High Capture Threshold4
- Failure to Capture3
- Capturing Problem2
- Over-Sensing2
- Device Sensing Problem1
- High impedance1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Known Impact Or Consequence To Patient6
- Dizziness1
Recent adverse event reports
The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2024High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD TRIGGERED AN ALERT FOR HIGH ATRIAL THRESHOLDS. FOUR MONTHS LATER, THE RIGHT VENTRICULAR (RV) LEAD DISPLAYED UNDERSENSING ON ONE BEAT ON THE CURRENT ELECTROGRAM (EGM). THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18986324 ↗MalfunctionMar 19, 2024Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING NOTED ON CURRENT ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18932959 ↗MalfunctionNov 7, 2023Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT UNDERSENSING/ATRIAL EVENTS FALLING IN BLANKING ON STORED ELECTROGRAMS (EGM), HOWEVER, THE DEVICE IS STAYING MODE SWITCHED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18081993 ↗MalfunctionApr 4, 2023Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED INTERMITTENT UNDERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16673424 ↗MalfunctionOct 2, 2022Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING RESULTING IN A RATE RESPONSIVE PACING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15526137 ↗MalfunctionSep 7, 2022Under-Sensing; Under-Sensing
IT WAS REPORTED THAT INTERMITTENT ATRIAL UNDERSENSING BY THE RIGHT ATRIAL (RA) LEAD WAS SUSPECTED DURING RECORDED EPISODES. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) FOR HIGH-RATE, NON-SUSTAINED (HRNS) EPISODES AND INCREASED SENSING INTEGRITY COUNTER (SIC). THE LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15374196 ↗InjuryNov 11, 2021High impedance; Capturing Problem; Device Dislodged or Dislocated; High impedance; Capturing Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH IMPEDANCE, AND A GRADUAL RISE IN THRESHOLDS. IT WAS FURTHER NOTED THAT UNDER FLUOROSCOPY THE RA LEAD HAD VERY LITTLE SLACK AND HAD MOST LIKELY PULLED BACK FROM ITS ORIGINAL POSITION. THE RA LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12800913 ↗MalfunctionNov 24, 2020Impedance Problem; High Capture Threshold; Impedance Problem; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED IMPEDANCE WARNING AND HIGH THRESHOLDS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10895028 ↗InjurySep 5, 2019Capturing Problem; Device Dislodged or Dislocated; Capturing Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE DAY AFTER THE IMPLANT PROCEDURE, THE RIGHT ATRIAL (RA) LEAD DISLODGED DUE TO PATIENT ANATOMY, WITH A RISE IN THRESHOLDS. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8964232 ↗MalfunctionJul 22, 2019Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING AND NOISE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8810867 ↗InjuryMay 22, 2019Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT ONE DAY POST REPLACEMENT PROCEDURE, THE RIGHT ATRIAL (RA) LEAD HAD NO CAPTURE. IT WAS NOTED THAT A CHEST X-RAY REVEALED THE RA LEAD DISLODGED. DURING LEAD REVISION, AFTER THE LEAD WAS REVISED THE PHYSICIAN WENT TO REATTACH THE LEAD INTO THE HEADER AND THE WRENCH WOULD NOT ENGAGE INTO THE SET SCREW. THE PHYSICIAN COULD NOT SCREW THE LEAD BACK INTO THE SET SCREW. THE DEVICE WAS EXPLANTED AND REPLACED, AND THE RA LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8633662 ↗InjuryAug 8, 2018Failure to Capture; Under-Sensing; High Capture Threshold
IT WAS LATER REPORTED THAT ATRIAL OUTPUT HAD BEEN INCREASED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7764090 ↗MalfunctionJul 5, 2018Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS OBSERVED WITH FUNCTIONAL UNDERSENSING WITH ATRIAL EVENTS FALLING INTO THE BLANKING AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7664604 ↗InjuryApr 30, 2018Failure to Capture; Device Dislodged or Dislocated; High Capture Threshold; Failure to Capture; Device Dislodged or Dislocated; High Capture Threshold
IT WAS REPORTED THAT FIVE DAYS POST IMPLANT, THE PATIENT EXPERIENCED DIZZINESS AND THE RIGHT ATRIAL (RA) LEAD HAD HIGH THRESHOLDS, NO CAPTURE AND HAD DISLODGED. THE LEAD WAS REVISED AND REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness
MAUDE report 7473779 ↗InjuryApr 10, 2018Failure to Capture; Device Sensing Problem; Device Dislodged or Dislocated; Failure to Capture; Device Sensing Problem; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD DISLODGED. THE RA WAS NOT SENSING OR CAPTURING, AND X-RAY CONFIRMED THE LEAD HAD PULLED BACK. THE ATRIAL LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7418802 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page