CapSure Sense® 407458 (UDI 00643169411081)Recalls & reported problems
Medtronic, Inc. · LEAD 407458 US BI RCMCRD · Class 3 · Implant
All models of CAPSURE SENSE (12)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic CapSure Sense lead, an implanted wire that delivers pacing signals to the heart's right ventricle or right atrium. The 34 reports mostly describe the lead failing to sense or capture the heartbeat properly, with rising pacing thresholds, and some mention infection or growth on the lead tip requiring removal of the pacemaker system. There are no deaths,28 injuries, and 6 malfunctions. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 34
- Deaths reported
- 0
- Injuries reported
- 28
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 34 adverse event reports that list this exact device (DI 00643169411081), received between Dec 9, 2016 and Jul 1, 2026 — classified as 28 injuries, 6 malfunctions.
The most frequently reported specific device problem is “High Capture Threshold” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- High Capture Threshold11
- Adverse Event Without Identified Device or Use Problem9
- Failure to Capture5
- Under-Sensing5
- Capturing Problem4
- High impedance4
- Intermittent Capture4
- Over-Sensing3
Patient problems
- No Known Impact Or Consequence To Patient11
- Unspecified Infection6
- No Clinical Signs, Symptoms or Conditions5
- Bradycardia2
- Dizziness2
- Fatigue2
- Syncope/Fainting2
- Weakness2
Recent adverse event reports
The 25 most recent of 34 reports, in the reporter's own words as published by the FDA.
InjuryJul 1, 2026Capturing Problem; High Capture Threshold
IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH AND RISING THRESHOLDS AND THE RIGHT ATRIAL (RA) LEAD EXHIBITED OUT OF RANGE LOW IMPEDANCE AND UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) ST ORED EPISODES RESULTING IN TERMINATION OF AT/AF DETECTED EVENTS IN PROGRESS, LACK OF DETECTION OF AT/AF, AND INAPPROPRIATE PACING. IT WAS ALSO NOTED THAT THERE WERE LOW AMPLITUDE P WAVES. BOTH LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Syncope/Fainting
MAUDE report 25678363 ↗InjuryApr 22, 2026High impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE AND HIGH THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24950841 ↗InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED VEGETATION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. A TEMPORARY IMPLANTABLE PULSE GENERATOR (IPG) WAS PLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Medical device site infection
MAUDE report 24814973 ↗InjuryAug 13, 2025High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) MAY HAVE EXHIBITED PREMATURE BATTERY DEPLETION. IT WAS ALSO NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED ELEVATED THRESHOLDS. THE IPG WAS REMOVED AND REPLACED AND THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22792152 ↗InjuryOct 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A SYSTEMIC INFECTION. THE SOURCE OF THE INFECTION IS REPORTED TO BE THE LEG, WHICH WAS AMPUTATED. VEGETATION WAS ALSO NOTED ON THE TIP OF THE RIGHT VENTRICULAR (RV) LEAD. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 20520427 ↗MalfunctionAug 15, 2024Device Sensing Problem; Unstable Capture Threshold; Device Sensing Problem; Unstable Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING AND APPEARED TO RESULT IN INAPPROPRIATE ATRIAL PACING. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED VARIABLE THRESHOLDS AND R WAVE MEASUREMENTS. THE RA LEAD AND RV LEAD REMAIN IN USE. NO PATIENT C OMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19991334 ↗InjuryMay 12, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMP LICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 16917442 ↗MalfunctionNov 23, 2022Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15850610 ↗InjuryNov 10, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THERE WAS EROSION AT THE IMPLANT SITE. THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED. A NEW IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS IMPLANTED 6 DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Pocket Erosion
MAUDE report 15775371 ↗InjuryAug 10, 2022Intermittent Capture; Capturing Problem; High Capture Threshold; Intermittent Capture; Capturing Problem; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING AND HIGH THRESHOLDS AND INTERMITTENT LOSS OF CAPTURE AT MAXIMUM OUTPUTS. THE RV LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15201631 ↗InjuryAug 26, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS FURTHER REPORTED THAT CULTURES WERE TAKEN AND NO BACTERIA WAS IDENTIFIED.
Patient problems: Sepsis; Unspecified Infection
MAUDE report 12374962 ↗InjuryApr 26, 2021Over-Sensing; Over-Sensing
IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE.
Patient problems: Chest Pain; Syncope/Fainting
MAUDE report 11722785 ↗InjuryApr 23, 2021Intermittent Capture; Failure to Capture; Intermittent Capture; Failure to Capture
IT WAS REPORTED THAT THE PATIENT HAD A SUSTAINED VENTRICULAR TACHYCARDIA AND ON THE SAME WENT INTO CARDIAC ARREST, WITH POSSIBLE EXTERNAL DEFIBRILLATION ADMINISTERED. IT WAS ALSO REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED POSSIBLE NON-CAPTURE AND ALSO POSSIBLE INTERMITTENT CAPTURE, ON STORED ELECTROGRAMS (EGMS). THE PATIENT WAS RUSHED TO THE CATHETER LABORATORY, WHERE AN UNSPECIFIED SURGICAL INTERVENTION WAS CARRIED OUT. THE PATIENT WAS THEN TRANSFERRED TO THE INTENSIVE CARE UNIT (ICU). THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Cardiac Arrest; Tachycardia
MAUDE report 11715442 ↗MalfunctionNov 16, 2020Pacing Problem; Pacing Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED NO PACING SPIKES. IT WAS NOTED THAT THE PATIENT EXPERIENCED BRADYCARDIA. THE LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia
MAUDE report 10842799 ↗InjuryOct 28, 2020High impedance; High Capture Threshold; High impedance; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND HIGH IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10746203 ↗InjuryJun 25, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS REMOVED DUE TO VALVE VEGETATION. THE IPG SYSTEM WAS REPLACED APPROXIMATELY TWELVE DAYS POST EXPLANT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 10194969 ↗InjuryJun 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED A SUBCLAVIAN STENOSIS AT THE POINT OF ENTRY OF EXISTING LEADS. THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLD AND PREMATURE BATTERY DEPLETION OF THE IMPLANTABLE PULSE GENERATOR (IPG) WAS OBSERVED. THE RIGHT VENTRICULAR (RV) LEAD REMAINS IN USE, THE RA LEAD WAS REPROGRAMMED AND THE IPG WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Vascular System (Circulation), Impaired
MAUDE report 10184445 ↗MalfunctionJan 7, 2020Failure to Capture; Failure to Capture
IT WAS FURTHER REPORTED THAT THE PATIENT WAS CARDIOVERTED OUT OF ATRIAL FIBRILLATION (AF). ATRIAL SENSING RETURNED TO SUFFICIENT LEVELS AND THE RIGHT VENTRICULAR (RV) LEAD PARAMETERS WERE OPERATING NORMALLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9556762 ↗InjuryNov 7, 2019Failure to Capture; Fracture; High impedance; Under-Sensing; Failure to Capture; Fracture; High impedance; Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE, LOSS OF SENSING, LOSS OF CAPTURE AND WAS FOUND TO BE FRACTURED. THE TERMINAL SEGMENT OF THE LEAD WAS CUT AND THE LEAD WAS CAPPED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9289368 ↗InjuryMay 14, 2019Failure to Capture; High Capture Threshold; Failure to Capture; High Capture Threshold
THE PATIENT PRESENTED TO THE CLINIC WITH WEAKNESS, DIZZINESS AND SHORTNESS OF BREATH. UPON DEVICE INTERROGATION, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND LOSS OF CAPTURE. THE RV LEAD WAS CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dizziness; Dyspnea; Weakness
MAUDE report 8607015 ↗InjuryJan 9, 2019High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR LEAD THRESHOLD WAS HIGH. THE OUTPUT WAS REPROGRAMMED TO PROVIDE A BETTER SAFETY MARGIN AND THE LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL LEAD HAD POSSIBLE INTERMITTENT UNDERSENSING, AS NOTED ON PRESENTING AND STORED ELECTROGRAMS (EGM.) THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8231073 ↗InjuryDec 18, 2018Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THERE WAS ATRIAL UNDERSENSING NOTED DURING EPISODES AND ON THE CURRENT ELECTROGRAM (EGM), ALONG WITH ONE ISOLATED POSSIBLE UNDERSENSED QRS COMPLEX. THE PATIENT WAS TO BE ADMITTED TO A HOSPITAL. THE LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8174856 ↗MalfunctionSep 26, 2018Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE PATIENT EXPERIENCED BRADYCARDIA AND DIZZINESS. OVERSESNSING ON THE RIGHT VENTRICULAR (RV) LEAD WAS OBSERVED. THE LEAD IS STILL IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bradycardia; Dizziness
MAUDE report 7911978 ↗MalfunctionJul 10, 2018Over-Sensing; Capturing Problem; High Capture Threshold; Capturing Problem
IT WAS REPORTED THAT THE THRESHOLD ON THE RIGHT VENTRICULAR (RV) LEAD HAD INCREASED AND HAD BEEN HIGH. THERE WERE ALSO POSSIBLE FAR FIELD R-WAVES (FFRW) AND POSSIBLE ATRIAL OVERSENSING NOTED ON THE RIGHT ATRIAL (RA) LEAD. THE RA LEAD WAS EXPLANTED AND REPLACED. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7673830 ↗InjuryJul 6, 2018Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THERE WAS UNDERSENSING ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7668199 ↗
Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes438 deaths19,632 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes200 deaths12,708 reports
- INGEVITY™ MRI 7741Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes36 deaths7,874 reports
- INGEVITY™+ 7841Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes15 deaths7,419 reports
- Tendril™ STSSt. Jude Medical, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes190 deaths6,776 reports
- INGEVITY™+ 7842Boston Scientific Corporation · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes21 deaths5,995 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes50 deaths5,027 reports
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes40 deaths5,000 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗