SPRINT QUATTRO® 6944-58 (UDI 00643169440852)Recalls & reported problems
Medtronic, Inc. · LEAD 6944-58 SPRINT QUATTRO US · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169440852), received between Oct 6, 2017 and Jun 24, 2024 — classified as 4 injuries.
The most frequently reported specific device problem is “Capturing Problem” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Capturing Problem1
- Device Dislodged or Dislocated1
- Fracture1
- High Capture Threshold1
- High impedance1
- Over-Sensing1
- Signal Artifact/Noise1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient1
- Unspecified Infection1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 24, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO AN INFECTION. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 19601239 ↗InjuryMay 15, 2024Signal Artifact/Noise; Over-Sensing; Signal Artifact/Noise; Over-Sensing
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE HOSPITAL DUE TO A LEAD INTEGRITY ALERT (LIA) TRIGGERED ON THE RIGHT VENTRICULAR (RV) LEAD WITH OVER ONE-HUNDRED SHORT V-V INTERVALS AND NON-SUSTAINED VENTRICULAR TACHYCARDIA EPISODES AND NOISE WAS NOTED. IT WAS FURTHER REPORTED THAT THERE WAS ELECTROMAGNETIC INTERFERENCE (EMI)/NOISE ON THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) AND A LOOSE SET SCREW ISSUE WAS SUSPECTED. THE RV LEAD AND ICD REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19322735 ↗InjuryApr 12, 2022Fracture; High impedance; Capturing Problem; Fracture; High impedance; Capturing Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD RISING THRESHOLDS AND HIGH/UNDEFINED PACING IMPEDANCE. THE LEAD IS SUSPECTED TO BE FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14084059 ↗InjuryOct 6, 2017Device Dislodged or Dislocated; High Capture Threshold; Device Dislodged or Dislocated; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXPERIENCED A MICRODISLODGEMENT AND HIGH THRESHOLDS. THE LEAD WAS REVISED, THE PACE/SENSE PORTION OF THE LEAD WAS CAPPED AND THE REST OF THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6924876 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page