CapSure Sense MRI™ SureScan™ 407458 (UDI 00643169448865)Recalls & reported problems

Medtronic, Inc. · LEAD 407458 MRI US BI MVC RCMRCD · Class 3 · Implant

All models of CapSure Sense MRI SureScan (8)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic CapSure Sense MRI lead has 758 reports, with 8 deaths and 469 injuries. Many reports describe high pacing thresholds, leads moving out of place, and failure to pace the heart, and some mention lead fractures and infections. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
766
Deaths reported
8
Injuries reported
471
Recalls naming it
0

Summary

The FDA's MAUDE database contains 766 adverse event reports that list this exact device (DI 00643169448865), received between Aug 9, 2017 and Sep 30, 2026 — classified as 8 deaths, 471 injuries, 287 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (215 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 10 reportsDec 2023: 5 reportsJan 2024: 7 reportsFeb 2024: 11 reportsMar 2024: 13 reportsApr 2024: 4 reportsMay 2024: 5 reportsJun 2024: 9 reportsJul 2024: 5 reportsAug 2024: 14 reportsSep 2024: 11 reportsOct 2024: 9 reportsNov 2024: 16 reportsDec 2024: 4 reportsJan 2025: 9 reportsFeb 2025: 10 reportsMar 2025: 10 reportsApr 2025: 5 reportsMay 2025: 12 reportsJun 2025: 12 reportsJul 2025: 3 reportsAug 2025: 8 reportsSep 2025: 10 reportsOct 2025: 6 reportsNov 2025: 8 reportsDec 2025: 7 reportsJan 2026: 13 reportsFeb 2026: 11 reportsMar 2026: 7 reportsApr 2026: 8 reportsMay 2026: 6 reportsJun 2026: 9 reportsJul 2026: 6 reportsAug 2026: 2 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold215
  • Device Dislodged or Dislocated143
  • Adverse Event Without Identified Device or Use Problem138
  • Under-Sensing138
  • Failure to Capture131
  • High impedance69
  • Over-Sensing55
  • Intermittent Capture54

Patient problems

  • No Clinical Signs, Symptoms or Conditions336
  • No Known Impact Or Consequence To Patient135
  • Unspecified Infection83
  • Bradycardia40
  • No Consequences Or Impact To Patient24
  • Dizziness22
  • Syncope/Fainting22
  • Chest Pain14

Recent adverse event reports

The 25 most recent of 766 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited occasional undersensing resulting in unnecessary pacing at times. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50942868 ↗
  • InjurySep 16, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced a pocket infection. Staphylococcus aureus and redness was noted. The Right Ventricular (RV) lead was explanted and replaced with a leadless Implantable Pulse Generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Bacterial Infection; Erythema; Medical device site infection; Unspecified Infection

    MAUDE report 50785648 ↗
  • MalfunctionSep 15, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited undersensing which resulted in over pacing. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50775970 ↗
  • MalfunctionSep 14, 2026Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited undersensing. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50752947 ↗
  • InjurySep 11, 2026Failure to Capture; Over-Sensing; Pacing Problem; Under-Sensing; Device Dislodged or Dislocated

    It was reported that the Right Ventricular (RV) lead exhibited loss of capture, dislodgement, oversensing and under sensing resulting in inappropriate pacing, and pre-termination of ventricular arrhythmia. The lead was repositioned and remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50728610 ↗
  • MalfunctionSep 3, 2026Pacing Problem; Under-Sensing

    It was reported that the Right Ventricular (RV) lead exhibited undersensing resulting in inappropriate pacing at times. The RV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50589287 ↗
  • MalfunctionSep 1, 2026Pacing Problem; Under-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50539972 ↗
  • InjuryAug 26, 2026High Capture Threshold

    It was reported that the Right Ventricular (RV) lead exhibited high thresholds. The RV lead was capped and replaced. No patient comp lications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50452872 ↗
  • InjuryAug 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection

    MAUDE report 50278400 ↗
  • InjuryJul 29, 2026Device Dislodged or Dislocated

    It was reported that the patient experienced bradycardia. The Right Ventricular (RV) lead exhibited dislodgement. The Implantable Pulse Generator (IPG) system was explanted and replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Bradycardia

    MAUDE report 50077655 ↗
  • MalfunctionJul 27, 2026Over-Sensing

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50016875 ↗
  • MalfunctionJul 20, 2026Human-Device Interface Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25869964 ↗
  • InjuryJul 9, 2026Device Dislodged or Dislocated; High Capture Threshold

    IT WAS REPORTED THAT ONE WEEK POST IMPLANT, THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED DISLODGEMENT AND HIGH THRESHOLDS. THE LEAD WAS REPOSITIONED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25759835 ↗
  • MalfunctionJul 7, 2026Electrical /Electronic Property Problem; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLDS AND A POLARITY SWITCH. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25731171 ↗
  • MalfunctionJul 2, 2026Break

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED A FRACTURE AND OVERSENSING. THE RA LEAD WAS EXPLANTED AND REPLACED. THE RIGHT VENTRICULAR (RV) LEAD WAS RETURNED TO THE MANUFACTURER AFTER BEING REPLACED DUE TO MEDICAL JUDGMENT AND SUBSEQUENTLY TESTED OUT OF SPECIFICATION DURING MANUFACTURER'S ANALYSIS. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25689421 ↗
  • InjuryJun 24, 2026Impedance Problem; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED RISING IMPEDANCE ALONG WITH HIGH THRESHOLDS. THE RV LEAD WAS CAPPED AND REPLACED AS PART OF THE SYSTEM UPGRADE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25615762 ↗
  • InjuryJun 22, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED DISCOMFORT, FEVER AND SWELLING. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED WITH NO REPLACEMENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Discomfort; Fever; Swelling/ Edema; Unspecified Infection

    MAUDE report 25577511 ↗
  • MalfunctionJun 18, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) STORED EPISODES RESULTING IN PRE-TERMINATION OF AT/AF DETECTED. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25544402 ↗
  • MalfunctionJun 16, 2026Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PACING LEAD WAS OPENED/ NOT USED FOR AN UNKNOWN REASON. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25515899 ↗
  • MalfunctionJun 15, 2026Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A SPIKE IN BIPOLAR IMPEDANCE. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25499464 ↗
  • MalfunctionJun 11, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25467713 ↗
  • MalfunctionJun 10, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING RESULTING IN INAPPROPRIATE PACING. THE LEAD REMAINS IN U SE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25455324 ↗
  • InjuryJun 9, 2026Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DEVELOPED AN INFECTION. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 25443461 ↗
  • MalfunctionJun 5, 2026Capturing Problem

    IT WAS REPORTED THAT THERE WAS A RISE IN RV (RIGHT VENTRICULAR) CAPTURE THRESHOLD. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25418134 ↗
  • MalfunctionMay 28, 2026Pacing Problem; Under-Sensing

    IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING ALONG WITH UNNECESSARY PACING AT TIMES AND INAPPROPRIATE PACING AT TIMES. THE PATIENT EXPERIENCED DIZZINESS/NEAR SYNCOPE. THE LEADS REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Presyncope

    MAUDE report 25316042 ↗

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →