N/A 6640008Recalls & reported problems
Medtronic Sofamor Danek, Inc. · DRIVER 6640008 4.75 SCREWDRIVER · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169492493), received between May 28, 2021 and Jun 6, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 6, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING THE IMPLANTATION OF A SCREW IN A SPINAL PROCEDURE, THE TIP OF THE SCREWDRIVER BROKE OFF IN THE SCREW HEAD AND COULD NOT BE RETRIEVED. IT WAS THEN DETERMINED THAT THE BROKEN TIP SHOULD REMAIN IN PLACE, AND RODS AND SET SCREWS WERE SUBSEQUENTLY PLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22159350 ↗MalfunctionSep 9, 2024Break; Break
INFORMATION WAS RECEIVED FROM A MANUFACTURER REPRESENTATIVE REGARDING A DEVICE USED FOR SPINAL THERAPY. IT WAS REPORTED THAT THE TIP OF THE DEVICE HAS BROKEN. THERE WAS NO PATIENT INVOLVEMENT REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20166467 ↗MalfunctionAug 10, 2022Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROVIDER (HCP) VIA MANUFACTURER REPRESENTATIVE REGARDING A SPINAL PRODUCT USED IN UNKNOWN SPINAL THERAPY. IT WAS REPORTED THAT DURING THE FIRST SCREW IMPLANTATION PROCEDURE (L4), THE SCREWDRIVER WAS BROKEN AND STUCK INTO SCREW HEAD AND CANNOT BE TAKEN OUT. WHEN ANOTHER SCREW WAS IMPLANTED, THE OTHER DRIVER WAS BROKEN. NO FURTHER COMPLICATIONS OR SYMPTOMS WERE REPORTED. ADDITIONAL INFORMATION WAS RECEIVED VIA MANUFACTURER REPRESENTATIVE THAT THERE WAS NO MALFUNCTION ASSOCIATED WITH THE SCREWS. BROKEN FRAGMENTS OF DRIVERS WAS LEFT INSIDE THE PATIENT AS SURGEON FAILED TO PULL OUT THE SCREWS DUE TO PATIENT'S HIGH BONE DENSITY. THERE IS NO REVISION SURGERY WAS SCHEDULED OR PLANNED AS THE SCREW WAS TOO HARD TO UNSCREW DURING THE SURGERY, AND THE MRI/CT IMAGE SHOWED THE SCREW AND ROD WERE STABLE (2 DAY AFTER SURGERY).
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 15202299 ↗MalfunctionMay 28, 2021Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT IMPLANTED WITH SCREWS USING A SCREWDRIVER FOR A TLIF. IT WAS REPORTED THAT TIP OF SCREW DRIVER BROKE DURING INSERTION OF SCREWS INSIDE SCREW CORE. THERE WERE NO PATIENT SYMPTOMS OR COMPLICATIONS AS A RESULT OF THE EVENT. THE PRE-OP DIAGNOSIS WAS DDD AND LEVELS IMPLANTED WERE L4 <(>&<)> L5. THE PRODUCT DID NOT COME IN CONTACT WITH THE PATIENT. THERE WAS NO FRAGMENT OF THE INSTRUMENT REMAINING IN THE PATIENT 'S BODY. THE EVENT OCCURRED DURING THE MIDDLE OF SURGERY. THERE WAS NO ADDITIONAL SURGERY OR TREATMENT/ REVISION SURGERY/ DELAY IN OVERALL PROCEDURE TIME/ HOSPITALIZATION AS A RESULT OF THE EVENT. THE PRODUCT WILL NOT BE REPLACED BY A MEDTRONIC PRODUCT IN THE FUTURE. NO FURTHER COMPLICATIONS WERE REPORTED/ ANTICIPATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11899517 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page