N/A 6640005Recalls & reported problems
Medtronic Sofamor Danek, Inc. · DRIVER 6640005 4.75 NON-RETAINING DRIVER · Class 1
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00643169492516), received between May 7, 2021 and Feb 12, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break4
Patient problems
- No Clinical Signs, Symptoms or Conditions4
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2025Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL (HCP) VIA A MEDTRONIC REPRESENTATIVE REGARDING A PATIENT UNDERGOING UNKNOWN SPINE SURGERY. IT WAS REPORTED THAT THE TIP OF THE SCREWDRIVER WAS BROKEN WHILE ADJUSTING A SCREW. DOCTOR WAS ABLE TO RETRIEVE THE PIECE OF THE DRIVER. NO HARM TO THE PATIENT OR DELAY TO THE CASE. NO FURTHER COMPLICATIONS REPORTED/ANTICIPATED DUE TO THIS EVENT. ADDITIONAL INFORMATION WAS RECEIVED FROM THE REP THAT PROCEDURE USED IN THIS EVENT WAS TLIF AND PRE OP DIAGNOSIS OF THIS EVENT WAS SPINAL STENOSIS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21362105 ↗MalfunctionFeb 3, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L4-1 HARDWARE R EMOVAL FOR AN INDICATION OF PSEUDOARTHROSIS. IT WAS REPORTED THAT THE TIP OF THE INTRUMENTS BROKEN OFF WHILE REMOVING THE PREVIOUS HARDWARE. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21289769 ↗MalfunctionJul 3, 2024Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING FRONT/POSTERIOR FUSION AND VERTEBRAL BODY REPLACEMENT. IT WAS REPORTED THAT THE TIP BROKE WHEN ATTEMPTING TO TIGHTEN THE SET SCREW. THERE WERE NO FRAGMENTS OF THE IMPLANTS OR INSTRUMENTS REMAINED IN THE PATIENT'S BODY. PRODUCT WILL BE RETURNED. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19663949 ↗MalfunctionMay 7, 2021Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT WITH CASE OF REDO RIGHT L2-3, LEFT L5-S1 DECOMPRESSION AND TRANSFORAMINAL LUMBAR INTERBODY FUSION AND REMOVAL OF PREVIOUS SCREWS AT L3-4-5, RESTABILIZATION AT L2-3-4-5-1 WITH POSTEROLATERAL FUSION FROM L2-S1 IT WAS REPORTED THAT WE PULLED OUT A 6.5X45 GLOBUS SCREW FROM RIGHT L5. THE SURGEON UPSIZED TO A 7.5X45, AND WENT STRAIGHT TO A VOYAGER 4.75 ATS SCREW. HE WAS STANDING ON THE PATIENTS LEFT SIDE REACHING ACROSS AND DOING THE RIGHT SIDED SCREW. THE ANGLE HE HAD IT IN WAS QUITE LATERAL. THE SCREW BECAME DIFFICULT TO ADVANCE, AND THEN ALL AT ONCE THE DRIVER JUST SPUN INSIDE THE TULIP BECAUSE THE TIP HAD BROKEN OFF IN IT. WE TRIED SEVERAL METHODS TO GET THE BROKEN SCREWDRIVER TIP OUT OF THE TULIP, AND TRIED TO HELICOPTER THE SCREW OUT ALTOGETHER USING A SHORT ROD AND SETCAP. WHAT WE ENDED UP HAVING TO DO WAS USE A METAL CUTTING BURR TO DRILL OUT THE BROKEN OFF PART OF THE SCREWDRIVER, AND THEN WE WERE ABLE TO BACK OUT THE SCREW USING THE HELICOPTER METHOD FROM ABOVE. THIS HAPPENED WITH FIRST THE NAVIGATED DRIVER, AND THEN A SHORT NON RETAINING DRIVER AS WELL BEFORE WE GOT TH
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11789002 ↗
Other ORTHOPEDIC MANUAL SURGICAL INSTRUMENT devices
- AML EXCEL 2001-42-000Depuy (Ireland) · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT2,009 reports
- NA 03.010.523Synthes GmbH · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,208 reports
- EXPRESSEW 214141Medos International Sàrl · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT1,166 reports
- N/A 9560524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT866 reports
- INHANCE 6500-11-104Depuy (Ireland) · Orthopedic manual surgical instrument831 reports
- INHANCE 6500-11-102Depuy (Ireland) · Orthopedic manual surgical instrument567 reports
- R3 71364450Smith & Nephew, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT520 reports
- N/A 9561524Medtronic Sofamor Danek, Inc. · ORTHOPEDIC MANUAL SURGICAL INSTRUMENT494 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.