MiniMed® 530G MMT-551NAH (UDI 00643169503632)Recalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-551NAH PRDGM 1.8 V3.0 PK EN · Class 3
All models of MiniMed 530G (71)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic MiniMed 530G insulin pump appears in 671 adverse event reports, including 2 deaths and 234 injuries. Reports mostly describe buttons that stick or stop responding, insulin squirting out during priming, unexpected no-delivery alarms, and blank displays. Some describe hospitalization for high blood sugar. There are no recalls that name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 671
- Deaths reported
- 2
- Injuries reported
- 234
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 671 adverse event reports that list this exact device (DI 00643169503632), received between May 17, 2017 and Sep 22, 2020 — classified as 2 deaths, 234 injuries, 435 malfunctions.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (137 reports).
2 other recalls cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Patient Device Interaction Problem137
- Adverse Event Without Identified Device or Use Problem82
- Device Difficult to Program or Calibrate80
- No Display/Image75
- Mechanical Problem73
- Device Displays Incorrect Message69
- Obstruction of Flow53
- No Apparent Adverse Event48
Patient problems
- No Consequences Or Impact To Patient387
- Hyperglycemia225
- Hypoglycemia63
- No Known Impact Or Consequence To Patient9
- Elevated Ketones/ Diabetic Ketoacidosis7
- Nausea4
- Vomiting4
- Death3
Recent adverse event reports
The 25 most recent of 671 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 22, 2020Mechanical Problem; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR ALARM DURING BOLUS. CUSTOMER WAS ABLE TO COMPLETE PUMP REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WAS RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10565834 ↗MalfunctionSep 2, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP'S RETAINER RING WAS LOOSENED. RESERVOIR WAS ABLE TO LOCK INTO PLACE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10486348 ↗MalfunctionJul 30, 2020Failure to Cycle; Failure to Cycle
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. CUSTOMER REPORTED DRIVE SUPPORT CAP WAS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10344046 ↗MalfunctionJul 29, 2020Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Power Problem; Priming Problem; Device Difficult to Program or Calibrate; Obstruction of Flow; Battery Problem; Power Problem; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A SQUIRTS INSULIN WHEN PRIMING, STICKY AND UNRESPONSIVE BUTTONS, UNEXPLAINED NO DELIVERY ALARMS, NO LOW BATTERY WARNING AND FAILED BATTERY TEST. THE CUSTOMER STATED THE ISSUES REMAINED UNRESOLVED. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10337041 ↗InjuryJul 23, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 476 MG/DL AND 236 MG/DL. CUSTOMER REPORTED NOW HE GOT IT DOWN BY GIVING HER UNITS OF INSULIN BUT HE FOUND THAT THERE WAS NO INSULIN ON THE INSULIN PUMP. THE PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10312959 ↗MalfunctionJul 21, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BLACK RING AROUND THE RESERVOIR COMPARTMENT WAS HANGING OUT AND THE TOP OF THE RESERVOIR COMPARTMENT WAS CRACKED OFF. CUSTOMER STATED THAT THE RESERVOIR DID LOCK IN PLACE WHEN INSERTED INTO INSULIN PUMP. CUSTOMER STATED THE INSULIN PUMP HAD COSMETIC DAMAGE. CUSTOMER STATED THE INFUSION SET AND RESERVOIR DID NOT SHOW SIGNS OF DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10300161 ↗MalfunctionJul 15, 2020Display or Visual Feedback Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Display or Visual Feedback Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE SCREEN OF INSULIN PUMP HAD BUTTON WAS HARDER TO PRESS AND INSULIN WAS SQUIRTED. CUSTOMER STATED THAT THE INSULIN PUMP HAD UNEXPLAINED NO DELIVERY ALARM AND BATTERY LIFE ABOUT A WEEK. CRACK ON INSULIN PUMP BATTERY COMPARTMENT AND LOWER DURING REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10275409 ↗MalfunctionJul 8, 2020Mechanical Problem; Device Difficult to Program or Calibrate; Priming Problem; Mechanical Problem; Device Difficult to Program or Calibrate; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP¿S BOLUS BUTTON WAS NOT WORKING, SQUIRTS INSULIN WHEN PRIMING AND WAS LOUD DURING REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10244498 ↗MalfunctionJun 24, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD CRACK AND RECEIVED MOTOR ERROR ALARM. CUSTOMER STATED THE RESERVOIR LIP RING DID NOT SHOW SIGNS OF DAMAGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10188105 ↗MalfunctionJun 23, 2020Priming Problem; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD SQUIRTS INSULIN DURING PRIMING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10183938 ↗MalfunctionJun 23, 2020Device Alarm System; Mechanical Problem; Battery Problem; Patient Device Interaction Problem; Device Alarm System; Mechanical Problem; Battery Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARM HARDER TO HEAR. CUSTOMER REPORTED THAT THE INSULIN PUMP REJECT NEW BATTERY, AND REWIND MULTIPLE TIME AND USE MORE INSULIN DURING PRIMING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10183129 ↗MalfunctionJun 22, 2020Mechanical Problem; Patient Device Interaction Problem; Mechanical Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT INSULIN PUMP HAD MOTOR ISSUE AND HAD CRACK ON SIDE AND EDGE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10177978 ↗MalfunctionJun 22, 2020Failure to Cycle; Failure to Cycle
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD A COMPROMISED FORCE SENSOR SYSTEM ALARM. THE CUSTOMER STATED THE DEVICE WAS SQUIRTING INSULIN OUT DURING MANUAL PRIME PROCESS. THE DRIVE SUPPORT CAP APPEARS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10177940 ↗MalfunctionJun 18, 2020Failure to Cycle; Failure to Cycle
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP ALARMED COMPROMISED FORCE SENSOR SYSTEM. CUSTOMER STATED THAT HE DRIVE SUPPORT CAP APPEARS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10167670 ↗MalfunctionJun 8, 2020Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem; Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTON WAS UNRESPONSIVE. ALSO HAD CRACKS ON BATTERY COMPARTMENT, HAD RAINBOW EFFECT ON SCREEN AND SQUIRTED OUT INSULIN FROM CANNULA WHILE PRIMING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10127740 ↗MalfunctionJun 8, 2020No Display/Image; Mechanical Problem; No Display/Image; Mechanical Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD MOTOR ERROR AND WAS NOT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10127300 ↗MalfunctionMay 28, 2020Display or Visual Feedback Problem; Mechanical Problem; Date/Time-Related Software Problem; Patient Device Interaction Problem; Display or Visual Feedback Problem; Mechanical Problem; Date/Time-Related Software Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE CLOCK NEED TO BE REPROGRAMMED ONCE IN A WHILE. CUSTOMER REPORTED THAT BATTERY PORT WAS CRACKED AND REWIND WAS LOUDER AND SCREEN HAD RAINBOW EFFECT. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10097256 ↗MalfunctionMay 25, 2020Device Difficult to Program or Calibrate; Device Difficult to Program or Calibrate
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE BUTTONS WERE NOT WORKING. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10085210 ↗MalfunctionMay 12, 2020Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem; Display or Visual Feedback Problem; Device Difficult to Program or Calibrate; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. CUSTOMER BLOOD GLUCOSE AT THE TIME OF INCIDENT UNKNOWN. TROUBLESHOOT WAS NOT PERFORMED. INSULIN PUMP WILL NOT BE RETURNING FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10046208 ↗MalfunctionMay 11, 2020Patient Device Interaction Problem; Patient Device Interaction Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE PIECE OF PLASTIC WAS MISSING AND INSULIN PUMP HAD CRACK AROUND THE BATTERY. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10043743 ↗MalfunctionMay 11, 2020Failure to Cycle; No Apparent Adverse Event; Priming Problem; Failure to Cycle; No Apparent Adverse Event; Priming Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN WAS SQUIRTING OUT DURING MANUAL PRIME. THE CUSTOMER STATED THAT DRIVE SUPPORT CAP WAS NORMAL. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10042464 ↗MalfunctionMay 8, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THEY HAD HYPERGLYCEMIA AND THE CUSTOMER DECLINED TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10038112 ↗MalfunctionMay 8, 2020Failure to Cycle; Failure to Cycle
INFORMATION RECEIVED BY MEDTRONIC INDICATED THAT THE INSULIN PUMP HAD COMPROMISED FORCE SENSOR SYSTEM. CUSTOMER WAS ABLE TO CLEAR THE ALARM AND WAS ABLE TO COMPLETE REWIND. NO HARM REQUIRING MEDICAL INTERVENTION WAS REPORTED. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10037504 ↗InjuryMay 6, 2020Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event; Adverse Event Without Identified Device or Use Problem; No Apparent Adverse Event
THE CUSTOMER REPORTED VIA PHONE CALL THAT THEY WERE EXPERIENCING HIGH BLOOD GLUCOSE LEVEL. CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 596 MG/DL AT THE TIME OF INCIDENT. CUSTOMER ALSO STATED THAT THE INSULIN PUMP RECEIVED INSULIN FLOW BLOCKED ALARM AND CANNULA WAS BENT. CUSTOMER DECLINED FOR TROUBLESHOOTING. THE INSULIN PUMP WILL NOT BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 10028152 ↗MalfunctionMay 4, 2020Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem; Device Difficult to Program or Calibrate; Battery Problem; Patient Device Interaction Problem
INFORMATION INDICATED BY MEDTRONIC THAT THE INSULIN PUMP HAD KEYPAD ANOMALY. THE INSULIN PUMP DID NOT HAVE MOISTURE DAMAGE. THE ISSUE WAS NOT RESOLVED. INSULIN PUMP WILL NOT BE RETURN FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10020023 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.
- Z-1751-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Artificial pancreas device system, threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page