MiniMed® 530G MMT-551PNAHRecalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-551PNAH PRDGM V3.0 PK US PW · Class 3
All models of MiniMed 530G (71)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00643169507272), received between Jun 21, 2017 and Jan 29, 2019 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Device Displays Incorrect Message” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Device Displays Incorrect Message2
- Air Leak1
- Bent1
- Failure to Deliver1
- No Apparent Adverse Event1
- Patient Device Interaction Problem1
Patient problems
- No Consequences Or Impact To Patient3
- Hyperglycemia2
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 29, 2019Patient Device Interaction Problem; Patient Device Interaction Problem
CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP BATTERY COMPARTMENT AREA WAS CRACK. CUSTOMER BLOOD GLUCOSE LEVEL WAS UNKNOWN. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE CUSTOMER WAS ADVISED THE PUMP WILL BE REPLACED AS PER TROUBLESHOOT. THE DEVICE WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8284251 ↗InjuryOct 15, 2018No Apparent Adverse Event; No Apparent Adverse Event
CUSTOMER REPORTED VIA PHONE CALL THAT THE CUSTOMER EXPERIENCED HYPERGLYCEMIA. THE CUSTOMER BLOOD GLUCOSE LEVEL WAS 520 MG/DL, OVER 400 MG/DL AND 560 MG/DL AT THE TIME OF INCIDENT. THE CUSTOMER WAS ASSISTED WITH TROUBLESHOOTING. THE CUSTOMER WAS TREATED WITH MANUAL INJECTION FOR HIGH BLOOD GLUCOSE LEVEL. CUSTOMER STATED THAT THE INSULIN PUMP HAD CRACK ON THE TOP RIGHT HAND CORNER FRONT TO THE SIDE 1/2 INCH, UNDER ESC BUTTON INTO THE WINDOW WHERE 1 INCH, BATTERY COMPARTMENT ON THE THREAD PART OF THAT OPENING. CUSTOMER DECLINE TROUBLESHOOTING FOR HIGH BLOOD GLUCOSE. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 7964576 ↗MalfunctionJun 18, 2018Failure to Deliver; Device Displays Incorrect Message
THE CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED MOTOR ERROR DURING BOLUS DELIVERY. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS 230 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS FLUSH. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP IS BEING REPLACED AND IS EXPECTED TO RETURN FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7608987 ↗InjuryJun 21, 2017Air Leak; Bent
THE CUSTOMER REPORTED VIA PHONE CALL THAT SHE EXPERIENCED A HIGH BLOOD GLUCOSE LEVEL OF 400 MG/DL. THE CUSTOMER'S CURRENT BLOOD GLUCOSE WAS 398 MG/DL AND IT WAS ALSO THE SAME READING AT THE TIME OF THE INCIDENT. CUSTOMER STATED THAT SHE HAS NOT TREATED YET. THE RECENT HIGH BLOOD GLUCOSE EVENT BEGAN ON (B)(6) 2017. TROUBLESHOOTING WAS PERFORMED AND AIR BUBBLES WERE FOUND ALONG THE TUBING LENGTH. CUSTOMER STATED THAT THE INSULIN EXITED AT THE QUICK RELEASE. CUSTOMER WAS ADVISED THAT POSITIONING IN THE MOTOR MAY HAVE SHIFTED. THE AIR BUBBLES WERE NOT SUCCESSFULLY REMOVED. CUSTOMER WAS ADVISED TO DO A COMPLETE SET CHANGE TO RESOLVE THE NO DELIVERY AND AIR BUBBLE ISSUE. CUSTOMER ALSO REPORTED THAT SHE HAD A BENT SENSOR. CUSTOMER WAS ADVISED THAT THE SENSOR MAY NOT BE WORKING OR MAY BE DAMAGED. THE RESERVOIR AND THE SENSOR WILL BE RETURNED FOR ANALYSIS. THE INSULIN PUMP WILL NOT BE RETURNED.
Patient problems: Hyperglycemia
MAUDE report 6658946 ↗MalfunctionJun 21, 2017Device Displays Incorrect Message
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE INSULIN PUMP HAD LOW BATTERY AND SENSOR END ALERTS. THE CUSTOMER¿S BLOOD GLUCOSE WAS 150 MG/DL AT THE TIME OF THE INCIDENT. THE CUSTOMER STATES THEY WERE TRYING TO REWIND BUT THE INSULIN PUMP BEGAN TO ALARM. THE CUSTOMER STATED THE BUTTONS WERE HARD TO PUSH AND NOW THEY ARE NOT WORKING. THE CUSTOMER WAS ADVISED TO OBSERVE TIME CLOCK FOR ONE MINUTE TO VERIFY IF TIME ADVANCES, FOUND TIME IS STILL ADVANCING. CUSTOMER WAS ALSO CALLING BACK AFTER LETTING THE INSULIN PUMP REST FOR TWO HOURS DUE TO A CONSTANT TONE. CUSTOMER STATED THE BUTTONS WERE NOT FUNCTIONAL. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6658422 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.
- Z-1751-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Artificial pancreas device system, threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.