MiniMed® 530G MMT-751RNAHRecalls & reported problems
Medtronic Minimed, Inc. · PUMP MMT-751RNAH PRDGM V3.0 PK US RC · Class 3
All models of MiniMed 530G (71)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00643169507975), received between Jul 7, 2017 and Jul 24, 2018 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P120010); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
- Mechanical Problem1
Patient problems
- Hyperglycemia2
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 24, 2018Mechanical Problem; Mechanical Problem
THE CUSTOMER REPORTED VIA PHONE CALL THE INSULIN PUMP ALARMED MOTOR ERROR AND OFF NO POWER. THE CUSTOMER¿S BLOOD GLUCOSE LEVEL WAS UNKNOWN AT THE TIME OF THE INCIDENT. THE CUSTOMER STATED THAT THE DRIVE SUPPORT CAP WAS NORMAL AND THAT THE INSULIN PUMP WAS NOT EXPOSED TO HIGH MAGNETIC FIELDS AND WAS ABLE TO COMPLETE THE REWIND PROCESS. CUSTOMER ALSO REPORTED THAT THE INSULIN PUMP HAD AN OFF NO POWER ALARM WITHOUT THE LOW BATTERY ALERT WARNING. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE INSULIN PUMP AND TO REVERT TO THE BACK-UP PLAN. THE INSULIN PUMP IS BEING REPLACED AND IS EXPECTED TO RETURN FOR ANALYSIS.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7713412 ↗InjuryFeb 20, 2018Adverse Event Without Identified Device or Use Problem
THE CUSTOMER REPORTED VIA PHONE CALL THAT THE PATIENT¿S BLOOD GLUCOSE WAS HIGH. THE CUSTOMER REPORTED THAT SHE'S BEEN WAKING UP WITH HIGH BLOOD GLUCOSE AND TRIED ADJUSTING HER BASALS AND SHE'S STILL WAKING UP HIGH. PATIENT'S BLOOD GLUCOSE AT TIME OF INCIDENT WAS UNKNOWN. PATIENT'S CURRENT BLOOD GLUCOSE WAS 183 MG/DL. THE CUSTOMER HAD SOME DIFFICULTY IN BREATHING AND EXTREME THIRST. THE CUSTOMER WAS AT 461MG/DL AT 1PM AND WAS 302MG/DL WHEN SHE WOKE UP. BASED ON CUSTOMER REPORT CUSTOMER DOES NOT ALLEGE PUMP WAS UNDER DELIVERING. CUSTOMER STATES THE DRIVE SUPPORT CAP APPEARS NORMAL. CUSTOMER IS NOT CALLING BACK TO PERFORM A HIGH PRESSURE TEST. CUSTOMER FEELS PUMP IS UNDER-DELIVERING, AND WOULD LIKE IT REPLACED. CUSTOMER IS TREATING FOR HIGH BLOOD GLUCOSE BY BOLUSING WITH THE PUMP AND ADJUSTING BASALS. THE INSULIN PUMP WILL BE RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 7284698 ↗InjuryJul 7, 2017Insufficient Device Problem Information
CUSTOMER REPORTED VIA PHONE CALL THAT THEY EXPERIENCED HIGH BLOOD GLUCOSE LEVEL OF 600 MG/DL. THE CUSTOMER WAS HOSPITALIZED FOR HIGH READINGS ON (B)(6) 2017 BETWEEN 9AM AND 10AM. CUSTOMER'S BLOOD GLUCOSE VALUE WAS 387 MG/DL AT THE TIME OF THE CALL. CUSTOMER REPORTED THAT THE HIGH BLOOD GLUCOSE LEVELS BEGAN ON (B)(6) 2017. TROUBLESHOOTING WAS PERFORMED. THE DRIVE SUPPORT CAP APPEARED NORMAL. THERE WERE NO LEAKS OR AIR BUBBLES. THERE WERE NO SITE OR INSULIN ISSUES. THE DEVICE SETTINGS WERE CORRECT. THE INSULIN PUMP PASSED THE HIGH PRESSURE TEST. CUSTOMER WAS ADVISED TO CHANGE THE INFUSION SET, RESERVOIR AND INSULIN AND TO TREAT PER HEALTHCARE PROFESSIONAL'S RECOMMENDATION. THE PRODUCT WAS RETURNED FOR ANALYSIS.
Patient problems: Hyperglycemia
MAUDE report 6695641 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P120010. One submission often covers many models, so check the product and lot details in each record.
- Z-1751-2026 · Feb 13, 2026 · Open, ClassifiedMiniMed 530G Insulin Pump (MMT-551, MMT-751)
- Z-2544-2014 · Mar 13, 2014 · TerminatedMedtronic MiniMed Paradigm Insulin Infusion Pumps, Paradigm models: MMT-511, MMT-512, MMT-712, MMT-712E, MMT-515, MMT-715, MMT-522, MMT-522K, MMT-722, MMT-722K, MMT-523, MMT-523K
Other Artificial pancreas device system, threshold suspend devices
- t:slim X2 Insulin Pump with Basal-IQ 1000886Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall17 deaths108,656 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend4 recalls98 deaths35,170 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000912Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall2 deaths30,219 reports
- t:slim X2 Insulin Pump with Basal-IQ 1000898Tandem Diabetes Care, Inc. · Automated Insulin Dosing , Threshold Suspend1 recall6 deaths16,106 reports
- MiniMed® 630G MMT-1715KMMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls50 deaths9,892 reports
- MiniMed® 630G MMT-1715KMedtronic Minimed, Inc. · Automated insulin dosing , threshold suspend4 recalls36 deaths7,947 reports
- MiniMed® 530G MMT-751NASMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend36 deaths7,479 reports
- MiniMed®630G MMT-1715KMedtronic Minimed, Inc. · Artificial pancreas device system, threshold suspend2 recalls17 deaths6,645 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page