CAPSUREFIX® NOVUS 407645 (UDI 00643169550803)Recalls & reported problems
Medtronic, Inc. · LEAD 407645 CAPSUREFIX NOVUS PACE US · Class 3 · Implant
Quick summary
This Medtronic Capsurefix Novus pacemaker lead is a wire that connects a pacemaker to the heart. Many of the 180 reports describe the lead misreading heart signals, and some describe infections requiring the device to be removed. The reports include 63 injuries, 115 malfunctions and 2 deaths. Reports from the last 12 months number 6, compared with 17 in the 12 months before. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 180
- Deaths reported
- 2
- Injuries reported
- 63
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 180 adverse event reports that list this exact device (DI 00643169550803), received between Dec 9, 2016 and Jul 10, 2026 — classified as 2 deaths, 63 injuries, 115 malfunctions.
The most frequently reported specific device problem is “Under-Sensing” (86 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Under-Sensing86
- Over-Sensing35
- Adverse Event Without Identified Device or Use Problem22
- Device Dislodged or Dislocated22
- High Capture Threshold16
- Decreased Sensitivity10
- Capturing Problem8
- Pacing Problem8
Patient problems
- No Clinical Signs, Symptoms or Conditions78
- No Known Impact Or Consequence To Patient66
- Unspecified Infection13
- Cardiac Perforation4
- Chest Pain3
- Fatigue3
- Muscle Stimulation3
- Pericardial Effusion3
Recent adverse event reports
The 25 most recent of 180 reports, in the reporter's own words as published by the FDA.
InjuryJul 10, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN INFECTION AND SEPSIS. THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Sepsis; Unspecified Infection
MAUDE report 25778340 ↗MalfunctionJul 2, 2026Over-Sensing; Under-Sensing
IT WAS REPORTED THAT THE PATIENT EXPERIENCED CHEST PAIN. IT WAS NOTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING DURING ATRIAL TACHYCARDIA/ ATRIAL FIBRILLATION (AT/AF) STORED EPISODES, RESULTING IN TERMINATION OF AT/AF DETECTED EVENT(S) IN PROGRESS. IT WAS FURTHER NOTED THAT THE RA LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE RA LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Chest Pain
MAUDE report 25688015 ↗MalfunctionMay 19, 2026Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THAT RIGHT ATRIAL (RA) LEAD UNDERSENSING WAS OBSERVED RESULTING IN UNNECESSARY PACING AND TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) DETECTED EVENTS IN PROGRESS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25222542 ↗MalfunctionMar 20, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING RESULTING IN MISCLASSIFICATION OF EPISODE DETECTION. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649826 ↗MalfunctionMar 18, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24627375 ↗MalfunctionFeb 5, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED POSSIBLE FAR FIELD R-WAVE (FFRW) OVERSENSING DURING A NON SUSTAINED VENTRICULAR TACHYCARDIA EVENT. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24269453 ↗InjuryOct 20, 2025Pacing Problem; Under-Sensing; High Capture Threshold; Pacing Problem; Under-Sensing; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED HIGH THRESHOLDS. IT WAS ALSO REPORTED THAT THE RA LEAD DISPLAYED UNDER SEN SING AND APPEARS TO RESULT IN INAPPROPRIATE ATRIAL PACING AND FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EPISODES STILL IN PROGRESS. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23336433 ↗InjurySep 5, 2025Over-Sensing; Under-Sensing; Over-Sensing; Under-Sensing
IT WAS FURTHER REPORTED THAT THE RA LEAD WAS REPROGRAMMED AGAIN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22975845 ↗InjuryAug 28, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING RESULTING IN INAPPROPRIATE MODE SWITCH AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OVERSENSING AND POSSIBLE T-WAVE OVERSENSING (TWOS). BOTH THE LEADS WERE REPROGRAMMED AND REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22910718 ↗InjuryAug 26, 2025Fracture; Fracture
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD WAS FRACTURED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22893398 ↗MalfunctionAug 18, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON STORED ELECTROGRAMS (EGM) RESULTING IN FASLSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22825115 ↗MalfunctionJul 20, 2025Electrical /Electronic Property Problem; Under-Sensing; Electrical /Electronic Property Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING NOTED ON THE PRESENTING ELECTROGRAM (EGM). IT WAS ALSO NOTED THAT THE ATRIAL POSITION CHECK HAD FAILED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22552555 ↗MalfunctionJun 13, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING ON A SINGLE BEAT DURING A STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22203291 ↗MalfunctionJun 6, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING ON STORED EPISODE ELECTROGRAMS (EGM)T TIMES. POTENTIAL FOR FALSE ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) DETECTION. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22156905 ↗MalfunctionJun 3, 2025Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED LOW IMPEDANCE WARNING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22132036 ↗MalfunctionJun 3, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF AT/AF EVENT(S) AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22130607 ↗MalfunctionMay 16, 2025Over-Sensing; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED OCCASIONAL UNDER-SENSING DURING A MONITORED VENTRICULAR TACHYCARDIA (VT) EPISODE. IT WAS ALSO REPORTED THAT THE RIGHT ATRIAL (RA) LEAD POTENTIALLY EXHIBITED FAR FIELD R-WAVE (FFRW) OVERSENSING. BOTH LEADS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22034188 ↗MalfunctionMar 12, 2025Electrical /Electronic Property Problem; Over-Sensing; Under-Sensing; Decreased Sensitivity; Electrical /Electronic Property Problem; Over-Sensing; Under-Sensing; Decreased Sensitivity
IT WAS FURTHER NOTED THAT THE RA LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENT(S) AT TIMES.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21585524 ↗MalfunctionFeb 21, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21431603 ↗MalfunctionJan 30, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING FURTHER TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACHYCARDIA (AT)/ ATRIAL FIBRILLATION (AF) EVENT(S) AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21272419 ↗MalfunctionJan 28, 2025Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD AND THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDER-SENSING. THE LEADS REMAIN IN USE . NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21253122 ↗MalfunctionDec 10, 2024Over-Sensing; Pacing Problem; Over-Sensing; Pacing Problem
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED FAR-FIELD R-WAVE OVERSENSING, WHICH CAUSED INAPPROPRIATE PACING. IT WAS ALSO REPORTED THAT THE RA LEAD EXHIBITED OVERSENSING ON STORED ELECTROGRAMS (EGM). THE PATIENT MENTIONED EXPERIENCING A SENSATION OF BEING SHOCKED, DESPITE ONLY HAVING A PACEMAKER. THE RA LEAD AND THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20890795 ↗MalfunctionNov 6, 2024Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED UNDERSENSING TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF ATRIAL TACH YCARDIA/ATRIAL FIBRILLATION (AT/AF) EVENTS AT TIMES. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20621154 ↗MalfunctionSep 20, 2024Pacing Problem; Under-Sensing; Pacing Problem; Under-Sensing
IT WAS REPORTED THAT THE RIGHT ATRIAL (RA) LEAD APPEARED TO BE UNDER SENSING DURING ATRIAL TACHYCARDIA (AT)/ATRIAL FIBRILLATION (AF) , TRIGGERING DEVICE CLASSIFIED FALSE TERMINATION OF AT/AF EVENT(S) AT TIMES. THE RA UNDER SENSING APPEARED TO RESULT IN INAPPROPRIATE ATRIAL PACING AT TIMES. THE RA LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20265036 ↗MalfunctionSep 13, 2024Break; Break
THE RIGHT ATRIAL (RA) LEAD WAS RETURNED TO THE MANUFACTURER, ANALYZED AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20210750 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.