Sprint Quattro Secure S® 6947M72Recalls & reported problems
Medtronic, Inc. · LEAD 6947M72 SPRINT US DF4 MCRD NG ACI · Class 3 · Implant
All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S (19)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This is a Medtronic Sprint Quattro Secure S 6947M72 electrode, an implanted wire lead for permanent heart pacing. Most of the 30 reports describe the lead misfiring at the sensing and pacing side: picking up noise, missing beats, sudden jumps in electrical resistance, and cracks or fractures in the wire, sometimes prompting shocks or removal and replacement. Others mention bloodstream infection or the lead twisting loose in the chest. There are no deaths and 20 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 30
- Deaths reported
- 0
- Injuries reported
- 20
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00643169550964), received between Feb 9, 2017 and Jul 10, 2026 — classified as 20 injuries, 10 malfunctions.
The most frequently reported specific device problem is “Over-Sensing” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Over-Sensing9
- Fracture8
- High impedance6
- High Capture Threshold5
- Signal Artifact/Noise5
- Inappropriate/Inadequate Shock/Stimulation4
- Under-Sensing4
- Adverse Event Without Identified Device or Use Problem3
Patient problems
- No Clinical Signs, Symptoms or Conditions12
- No Known Impact Or Consequence To Patient7
- Shock from Patient Lead(s)4
- Unspecified Infection3
- Twiddlers Syndrome2
- Bacteremia1
- Dizziness1
- Dyspnea1
Recent adverse event reports
The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.
InjuryJul 10, 2026Low impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE. IT WAS FURTHER REPORTED THAT DURING THE COMPETITOR DEVICE CHANGE OUT PROCEDURE, THE RV LEAD COULD NOT BE REMOVED DUE TO THE COMPETITOR DEVICE SETSCREW APPEARED TO BE STRIPPED AND WOULD NOT ENGAGE WITH THE WRENCH TOOL. AS A RESULT, THE PHYSICIAN WAS UNABLE TO SEPARATE THE RV LEAD FROM THE COMPETITOR DEVICE. THE RV LEAD WAS CUT AND CAPPED IN THE POCKET AND A NEW RV LEAD WAS IMPLANTED WITHOUT ANY KNOWN ISSUES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25780146 ↗MalfunctionMar 31, 2026Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). IN ADDITION, THE PATIENT EXPERIENCED SHORTNESS OF BREATH. THE RV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Dyspnea
MAUDE report 24744957 ↗MalfunctionFeb 20, 2026Under-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24401302 ↗InjurySep 10, 2025Signal Artifact/Noise; Fracture; Over-Sensing; Signal Artifact/Noise; Fracture; Over-Sensing
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO HIGH RATE NON-SUSTAINED (HRNS) EPISODES AND SENSING INTEGRITY COUNTER (SIC). IN ADDITION, A LEAD NOISE WARNING TRIGGERED AND FRACTURE WAS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23017476 ↗InjuryJul 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST IMPLANT PROCEDURE, THE PATIENT DEVELOPED AN INFECTION. THE SOURCE OF THE INFECTION WAS NOTED TO BE BACTEREMIA. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Bacteremia; Unspecified Infection
MAUDE report 22635446 ↗InjuryFeb 29, 2024Fracture; High impedance; Impedance Problem; Fracture; High impedance; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED VARIABLE PACING IMPEDANCE TO HIGH/UNDEFINED LEVELS. A LOW VOLTAGE FRACTURE WAS SUSPECTED. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18808147 ↗InjuryApr 28, 2023Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold; Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS OVERSENSING NOISE AND DELIVERED INAPPROPRIATE THERAPY. THE LEAD HAD HIGH RV THRESHOLDS AND WAS UNABLE TO CAPTURE DUE TO A CONFIRMED FRACTURE. IT WAS NOTED THAT THERE WAS AN OUT OF RANGE (OOR) IMPEDANCE WARNING FOR HIGH/UNDEFINED RVTIP TO RVCOIL PACING IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 16832902 ↗InjuryMar 3, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD BECOME DISLODGED DUE TO TWIDDLER'S SYNDROME. IT WAS NOTED THAT WHEN THE PATIENT TURNED THE DEVICE OR LEAD WOULD MOVE, THE PATIENT TRIED TO STOP IT BY TWIRLING THE LEADS. THE LEAD WAS EXPLANTED AND REPLACE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Twiddlers Syndrome
MAUDE report 16478673 ↗InjuryNov 1, 2022Signal Artifact/Noise; Fracture; Impedance Problem; Signal Artifact/Noise; Fracture; Impedance Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE ON THE PACE SENSE PORTION OF THE LEAD. IN ADDITION A JUMP IN PACING IMPEDANCE WAS NOTED. A LEAD FRACTURE WAS CONFIRMED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15709372 ↗InjuryAug 15, 2022Failure to Capture; Under-Sensing; Device Dislodged or Dislocated; High Capture Threshold; Failure to Capture; Under-Sensing; Device Dislodged or Dislocated; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD DISLODGED SEEN ON FLUOROSCOPY POST IMPLANT. THE LEAD HAD HIGH THRESHOLDS WITH NO CAPTURE. THE LEAD HAD UNDERSENSING. THE LEAD WAS REVISED AND REPOSITIONED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15230181 ↗InjuryFeb 23, 2022Fracture; High impedance; Fracture; High impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH UNDEFINED SUPERIOR VENA CAVA (SVC) IMPEDANCE. SVC FRACTURE WAS CONFIRMED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13585428 ↗MalfunctionDec 16, 2021Break; Break
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXTREMELY TWISTED AND KINKED BETWEEN THE TWO COILS. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13019216 ↗InjuryOct 15, 2021Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS FURTHER REPORTED THAT THE SENSING PARAMETERS WERE REPROGRAMMED.
Patient problems: Shock from Patient Lead(s)
MAUDE report 12641292 ↗InjuryOct 11, 2021High impedance; Over-Sensing; Impedance Problem; High impedance; Over-Sensing; Impedance Problem
IT WAS FURTHER REPORTED THAT THE LEAD HAD UNDEFINED HIGH PACING IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12605666 ↗MalfunctionMar 31, 2021Disconnection; Disconnection
IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE WHEN THE PHYSICIAN WAS TUNNELING THE RIGHT VENTRICULAR (RV) LEAD, THE PIN BROKE. ANOTHER LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11595030 ↗InjuryJan 25, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated
IT WAS REPORTED THAT BOTH THE RIGHT ATRIAL AND RIGHT VENTRICULAR LEADS DISLODGED DUE TO TWIDDLER'S SYNDROME. THE LEFT SUBCLAVIAN VEIN WAS NOTED TO BE OCCLUDED. THE LEADS WERE NOTED TO HAVE BEEN ABANDONED. THE LEADS WERE EXPLANTED AND REPLACED ON THE OPPOSITE SIDE A FEW DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Obstruction/Occlusion; Twiddlers Syndrome
MAUDE report 11227265 ↗MalfunctionJan 4, 2021Under-Sensing; Under-Sensing
IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT AFTER HAVING RECEIVED A SHOCK. A REMOTE TRANSMISSION INDICATED A VENTRICULAR UNDERSENSED BEAT NOTED ON ELECTROGRAM (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11109148 ↗MalfunctionDec 10, 2020Over-Sensing; High Capture Threshold; Over-Sensing; High Capture Threshold
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A LEAD INTEGRITY ALERT (LIA) TRIGGERED BY HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). IN ADDITION, THE LEAD HAD HIGH THRESHOLDS AND T-WAVE OVERSENSING (TWOS) AND OVERSENSING WAS NOTED ON STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10987834 ↗InjuryJul 1, 2020Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation; Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A CONFIRMED FRACTURE. THE LEAD HAD HIGH AND UNDEFINED PACING IMPEDANCE AND DELIVERED AN INAPPROPRIATE SHOCK. THE LEAD WAS INITIALLY PROGRAMMED OFF AND SUBSEQUENTLY EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 10220539 ↗InjuryMar 23, 2020Fracture; Fracture
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9868499 ↗MalfunctionMar 11, 2020Low impedance; Low impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A LOW IMPEDANCE OUT OF RANGE ALERT. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9819828 ↗InjuryJan 22, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS INFECTED. THE CRT-D SYSTEM REMAINS IN USE, BUT AN EXTRACTION PROCEDURE WAS PLANNED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 9616067 ↗InjuryFeb 19, 2019Signal Artifact/Noise; Fracture; High impedance; Signal Artifact/Noise; Fracture; High impedance
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE MEASUREMENTS FOR BOTH THE SUPERIOR VENA CAVA (SVC) AND RIGHT VENTRICULAR (RV) COILS. IT WAS FURTHER REPORTED THAT THE SVC COIL WAS FRACTURED AND HAD HIGH UNDEFINED IMPEDANCE. ISOMETRICS SHOWED THAT THE SVC TO RV COIL WAS NOISY. THE LEAD WAS REPROGRAMMED AND PARTIALLY REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8351296 ↗InjuryDec 5, 2018Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT A CAREALERT INDICATED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY AFTER OVERSENSING NOISE. THERE WERE OTHER EPISODES THAT INDICATED ONE TO TWO SECONDS OF OVERSENSING. IT WAS ALSO NOTED THAT APPROXIMATELY A WEEK EARLIER THE RIGHT VENTRICULAR (RV) LEAD HAD A LEAD INTEGRITY WARNING TRIGGERED DUE TO HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 8134970 ↗InjurySep 17, 2018High Capture Threshold; High Capture Threshold
IT WAS REPORTED THAT THERE WAS HIGH THRESHOLD ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED DURING THE RV LEAD REPLACEMENT, THE RIGHT ATRIAL (RA) LEAD DISLODGED. IT WAS NOTED THE LEAD WAS PREVIOUSLY TURNED OFF DUE TO ATRIAL STANDSTILL. THE LEAD WAS REMOVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7882864 ↗
Other permanent pacemaker Electrode devices
- CapSureFix Novus MRI™ SureScan® 5076-45Medtronic, Inc. · permanent pacemaker Electrode136 deaths16,004 reports
- CapSureFix Novus MRI™ SureScan® 5076-58Medtronic, Inc. · permanent pacemaker Electrode124 deaths8,405 reports
- Capsurefix® Novus 5076-45Medtronic, Inc. · permanent pacemaker Electrode85 deaths7,050 reports
- CapSureFix Novus MRI™ SureScan™ 407652Medtronic, Inc. · permanent pacemaker Electrode69 deaths6,097 reports
- SelectSecure™ MRI SureScan™ 383069Medtronic, Inc. · permanent pacemaker Electrode56 deaths4,850 reports
- Sprint Quattro Secure® 694765Medtronic, Inc. · permanent pacemaker Electrode73 deaths3,852 reports
- Attain Performa™ MRI SureScan™ 429888Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,377 reports
- CapSureFix Novus MRI™ SureScan™ 407645Medtronic, Inc. · permanent pacemaker Electrode38 deaths3,370 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page