Sprint Quattro Secure S® 6947M72Recalls & reported problems

Medtronic, Inc. · LEAD 6947M72 SPRINT US DF4 MCRD NG ACI · Class 3 · Implant

All models of Medtronic Sprint Quattro Lead (67) · Sprint Quattro Secure S (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This is a Medtronic Sprint Quattro Secure S 6947M72 electrode, an implanted wire lead for permanent heart pacing. Most of the 30 reports describe the lead misfiring at the sensing and pacing side: picking up noise, missing beats, sudden jumps in electrical resistance, and cracks or fractures in the wire, sometimes prompting shocks or removal and replacement. Others mention bloodstream infection or the lead twisting loose in the chest. There are no deaths and 20 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
30
Deaths reported
0
Injuries reported
20
Recalls naming it
0

Summary

The FDA's MAUDE database contains 30 adverse event reports that list this exact device (DI 00643169550964), received between Feb 9, 2017 and Jul 10, 2026 — classified as 20 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Over-Sensing” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Over-Sensing9
  • Fracture8
  • High impedance6
  • High Capture Threshold5
  • Signal Artifact/Noise5
  • Inappropriate/Inadequate Shock/Stimulation4
  • Under-Sensing4
  • Adverse Event Without Identified Device or Use Problem3

Patient problems

  • No Clinical Signs, Symptoms or Conditions12
  • No Known Impact Or Consequence To Patient7
  • Shock from Patient Lead(s)4
  • Unspecified Infection3
  • Twiddlers Syndrome2
  • Bacteremia1
  • Dizziness1
  • Dyspnea1

Recent adverse event reports

The 25 most recent of 30 reports, in the reporter's own words as published by the FDA.

  • InjuryJul 10, 2026Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW SUPERIOR VENA CAVA (SVC) COIL IMPEDANCE. IT WAS FURTHER REPORTED THAT DURING THE COMPETITOR DEVICE CHANGE OUT PROCEDURE, THE RV LEAD COULD NOT BE REMOVED DUE TO THE COMPETITOR DEVICE SETSCREW APPEARED TO BE STRIPPED AND WOULD NOT ENGAGE WITH THE WRENCH TOOL. AS A RESULT, THE PHYSICIAN WAS UNABLE TO SEPARATE THE RV LEAD FROM THE COMPETITOR DEVICE. THE RV LEAD WAS CUT AND CAPPED IN THE POCKET AND A NEW RV LEAD WAS IMPLANTED WITHOUT ANY KNOWN ISSUES. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25780146 ↗
  • MalfunctionMar 31, 2026Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS). IN ADDITION, THE PATIENT EXPERIENCED SHORTNESS OF BREATH. THE RV LEAD REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Dyspnea

    MAUDE report 24744957 ↗
  • MalfunctionFeb 20, 2026Under-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED UNDERSENSING. THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24401302 ↗
  • InjurySep 10, 2025Signal Artifact/Noise; Fracture; Over-Sensing; Signal Artifact/Noise; Fracture; Over-Sensing

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO HIGH RATE NON-SUSTAINED (HRNS) EPISODES AND SENSING INTEGRITY COUNTER (SIC). IN ADDITION, A LEAD NOISE WARNING TRIGGERED AND FRACTURE WAS SUSPECTED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23017476 ↗
  • InjuryJul 28, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY TWO MONTHS POST IMPLANT PROCEDURE, THE PATIENT DEVELOPED AN INFECTION. THE SOURCE OF THE INFECTION WAS NOTED TO BE BACTEREMIA. THE IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) SYSTEM WAS REMOVED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bacteremia; Unspecified Infection

    MAUDE report 22635446 ↗
  • InjuryFeb 29, 2024Fracture; High impedance; Impedance Problem; Fracture; High impedance; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED VARIABLE PACING IMPEDANCE TO HIGH/UNDEFINED LEVELS. A LOW VOLTAGE FRACTURE WAS SUSPECTED. THE RV LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18808147 ↗
  • InjuryApr 28, 2023Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold; Signal Artifact/Noise; Failure to Capture; Fracture; High impedance; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS OVERSENSING NOISE AND DELIVERED INAPPROPRIATE THERAPY. THE LEAD HAD HIGH RV THRESHOLDS AND WAS UNABLE TO CAPTURE DUE TO A CONFIRMED FRACTURE. IT WAS NOTED THAT THERE WAS AN OUT OF RANGE (OOR) IMPEDANCE WARNING FOR HIGH/UNDEFINED RVTIP TO RVCOIL PACING IMPEDANCE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 16832902 ↗
  • InjuryMar 3, 2023Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD BECOME DISLODGED DUE TO TWIDDLER'S SYNDROME. IT WAS NOTED THAT WHEN THE PATIENT TURNED THE DEVICE OR LEAD WOULD MOVE, THE PATIENT TRIED TO STOP IT BY TWIRLING THE LEADS. THE LEAD WAS EXPLANTED AND REPLACE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Twiddlers Syndrome

    MAUDE report 16478673 ↗
  • InjuryNov 1, 2022Signal Artifact/Noise; Fracture; Impedance Problem; Signal Artifact/Noise; Fracture; Impedance Problem

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED NOISE ON THE PACE SENSE PORTION OF THE LEAD. IN ADDITION A JUMP IN PACING IMPEDANCE WAS NOTED. A LEAD FRACTURE WAS CONFIRMED. THE LEAD WAS CAPPED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15709372 ↗
  • InjuryAug 15, 2022Failure to Capture; Under-Sensing; Device Dislodged or Dislocated; High Capture Threshold; Failure to Capture; Under-Sensing; Device Dislodged or Dislocated; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD DISLODGED SEEN ON FLUOROSCOPY POST IMPLANT. THE LEAD HAD HIGH THRESHOLDS WITH NO CAPTURE. THE LEAD HAD UNDERSENSING. THE LEAD WAS REVISED AND REPOSITIONED. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15230181 ↗
  • InjuryFeb 23, 2022Fracture; High impedance; Fracture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD TRIGGERED AN ALERT FOR HIGH UNDEFINED SUPERIOR VENA CAVA (SVC) IMPEDANCE. SVC FRACTURE WAS CONFIRMED. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13585428 ↗
  • MalfunctionDec 16, 2021Break; Break

    IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD WAS EXTREMELY TWISTED AND KINKED BETWEEN THE TWO COILS. THE LEAD WAS NOT USED AND ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13019216 ↗
  • InjuryOct 15, 2021Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS FURTHER REPORTED THAT THE SENSING PARAMETERS WERE REPROGRAMMED.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 12641292 ↗
  • InjuryOct 11, 2021High impedance; Over-Sensing; Impedance Problem; High impedance; Over-Sensing; Impedance Problem

    IT WAS FURTHER REPORTED THAT THE LEAD HAD UNDEFINED HIGH PACING IMPEDANCE. THE LEAD WAS CAPPED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12605666 ↗
  • MalfunctionMar 31, 2021Disconnection; Disconnection

    IT WAS REPORTED THAT DURING THE IMPLANT PROCEDURE WHEN THE PHYSICIAN WAS TUNNELING THE RIGHT VENTRICULAR (RV) LEAD, THE PIN BROKE. ANOTHER LEAD WAS USED TO COMPLETE THE PROCEDURE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11595030 ↗
  • InjuryJan 25, 2021Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT BOTH THE RIGHT ATRIAL AND RIGHT VENTRICULAR LEADS DISLODGED DUE TO TWIDDLER'S SYNDROME. THE LEFT SUBCLAVIAN VEIN WAS NOTED TO BE OCCLUDED. THE LEADS WERE NOTED TO HAVE BEEN ABANDONED. THE LEADS WERE EXPLANTED AND REPLACED ON THE OPPOSITE SIDE A FEW DAYS LATER. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Obstruction/Occlusion; Twiddlers Syndrome

    MAUDE report 11227265 ↗
  • MalfunctionJan 4, 2021Under-Sensing; Under-Sensing

    IT WAS REPORTED THAT THE PATIENT PRESENTED TO THE EMERGENCY DEPARTMENT AFTER HAVING RECEIVED A SHOCK. A REMOTE TRANSMISSION INDICATED A VENTRICULAR UNDERSENSED BEAT NOTED ON ELECTROGRAM (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11109148 ↗
  • MalfunctionDec 10, 2020Over-Sensing; High Capture Threshold; Over-Sensing; High Capture Threshold

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A LEAD INTEGRITY ALERT (LIA) TRIGGERED BY HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). IN ADDITION, THE LEAD HAD HIGH THRESHOLDS AND T-WAVE OVERSENSING (TWOS) AND OVERSENSING WAS NOTED ON STORED ELECTROGRAMS (EGM). THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10987834 ↗
  • InjuryJul 1, 2020Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation; Fracture; High impedance; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A CONFIRMED FRACTURE. THE LEAD HAD HIGH AND UNDEFINED PACING IMPEDANCE AND DELIVERED AN INAPPROPRIATE SHOCK. THE LEAD WAS INITIALLY PROGRAMMED OFF AND SUBSEQUENTLY EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 10220539 ↗
  • InjuryMar 23, 2020Fracture; Fracture

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED A POSSIBLE FRACTURE. THE LEAD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9868499 ↗
  • MalfunctionMar 11, 2020Low impedance; Low impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD HAD A LOW IMPEDANCE OUT OF RANGE ALERT. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9819828 ↗
  • InjuryJan 22, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS INFECTED. THE CRT-D SYSTEM REMAINS IN USE, BUT AN EXTRACTION PROCEDURE WAS PLANNED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 9616067 ↗
  • InjuryFeb 19, 2019Signal Artifact/Noise; Fracture; High impedance; Signal Artifact/Noise; Fracture; High impedance

    IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH IMPEDANCE MEASUREMENTS FOR BOTH THE SUPERIOR VENA CAVA (SVC) AND RIGHT VENTRICULAR (RV) COILS. IT WAS FURTHER REPORTED THAT THE SVC COIL WAS FRACTURED AND HAD HIGH UNDEFINED IMPEDANCE. ISOMETRICS SHOWED THAT THE SVC TO RV COIL WAS NOISY. THE LEAD WAS REPROGRAMMED AND PARTIALLY REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8351296 ↗
  • InjuryDec 5, 2018Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation; Signal Artifact/Noise; Over-Sensing; Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT A CAREALERT INDICATED THAT THE PATIENT RECEIVED INAPPROPRIATE THERAPY AFTER OVERSENSING NOISE. THERE WERE OTHER EPISODES THAT INDICATED ONE TO TWO SECONDS OF OVERSENSING. IT WAS ALSO NOTED THAT APPROXIMATELY A WEEK EARLIER THE RIGHT VENTRICULAR (RV) LEAD HAD A LEAD INTEGRITY WARNING TRIGGERED DUE TO HIGH RATE NON-SUSTAINED EPISODES AND SENSING INTEGRITY COUNTER (SIC). THE RV LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Shock from Patient Lead(s)

    MAUDE report 8134970 ↗
  • InjurySep 17, 2018High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THERE WAS HIGH THRESHOLD ON THE RIGHT VENTRICULAR (RV) LEAD. THE LEAD WAS EXPLANTED AND REPLACED. IT WAS ALSO REPORTED DURING THE RV LEAD REPLACEMENT, THE RIGHT ATRIAL (RA) LEAD DISLODGED. IT WAS NOTED THE LEAD WAS PREVIOUSLY TURNED OFF DUE TO ATRIAL STANDSTILL. THE LEAD WAS REMOVED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7882864 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →