MyCareLink™ 24950GRecalls & reported problems
Medtronic, Inc. · MON 24950G MYCARELINK/SVC USA · Class 3
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 17
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 17 adverse event reports that list this exact device (DI 00643169552104), received between Sep 22, 2017 and Sep 15, 2022 — classified as 17 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (14 reports).
1 FDA recall record name this device's identifier.
9 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jul 14, 2017Terminated
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
Z-0399-2018 · Root cause: Error in labeling · FDA record ↗
Most reported problems
Device problems
- Overheating of Device14
- Mechanical Problem2
- Inappropriate/Inadequate Shock/Stimulation1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- No Known Impact Or Consequence To Patient5
- No Consequences Or Impact To Patient4
- Swollen Lymph Nodes/Glands1
Recent adverse event reports
The 17 most recent of 17 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 15, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH WHENEVER THE PATIENT USES IT. NO TROUBLESHOOTING WAS TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15432482 ↗MalfunctionJul 14, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR BECAME SO HOT THAT IT MELTED PLASTIC. NO TROUBLESHOOTING WAS TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15020027 ↗MalfunctionJun 28, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Swollen Lymph Nodes/Glands
MAUDE report 14843609 ↗MalfunctionSep 24, 2021Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12525632 ↗MalfunctionJul 23, 2021Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS VERY WARM. THE PATIENT WAS ADVISED TO UNPLUG IT DUE TO BEING A FIRE HAZARD. THE REMOTE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12218559 ↗MalfunctionMay 18, 2021Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS HOT AND HAS BEEN BURNING UP SINCE YESTERDAY. IT WAS NOTED THAT THE NURSE WROTE DOWN ALL INFORMATION OF THE MONITOR SO NOT TO TOUCH THE UNIT. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11845314 ↗MalfunctionApr 9, 2021Overheating of Device; Overheating of Device
THE MONITOR WAS REPLACED AND RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11640370 ↗MalfunctionFeb 18, 2021Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING. NO TROUBLESHOOTING INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11345968 ↗MalfunctionOct 13, 2020Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10669196 ↗MalfunctionAug 30, 2020Overheating of Device; Overheating of Device
THE MONITOR WAS REPLACED AND RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10469733 ↗MalfunctionJul 2, 2020Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT WAS SHOCKED THREE TIMES. THE PATIENT WAS ADVISED TO GO TO THE EMERGENCY ROOM. THE REMOTE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10222963 ↗MalfunctionJun 5, 2020Overheating of Device; Overheating of Device
THE REMOTE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10122243 ↗MalfunctionOct 31, 2019Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THERE WAS A WHITE POWDER SUBSTANCE LEAKING FROM THE REMOTE MONITOR WHICH WAS BELIEVED TO BE A BATTERY ACID. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9261335 ↗MalfunctionNov 12, 2018Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE PATIENT SAW A "GOOPY" SUBSTANCE UNDER THE REMOTE MONITOR AND IT WAS ALLEGED THAT IT WAS THE BATTERY THAT WAS LEAKING. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED.NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8060174 ↗MalfunctionOct 22, 2018Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 7988608 ↗MalfunctionAug 2, 2018Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR'S READER WAS OVERHEATING. THE REMOTE MONITOR WAS POWER CYCLED. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7741718 ↗MalfunctionSep 22, 2017Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR DISPLAYED AN ERROR CODE INDICATING THE READER IS OFF THE MONITOR BASE AND HAS POWERED DOWN, IS OFF THE BASE STATION AND HAS NO BATTERY LIFE REMAINING OR THE READER IS OUT OF RANGE OF THE MONITOR. IT WAS ALSO REPORTED THAT THE READER WAS VERY HOT AND TOO HOT TO PLACE ON SKIN. THE MONITOR WAS REPLACED AND RETURNED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6888613 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.
- Z-2169-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
- Z-2168-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2506-2020 · Jun 4, 2020 · Open, ClassifiedPatient Connector, Model Number 24967
- Z-2505-2020 · Jun 4, 2020 · Open, ClassifiedCareLink SmartSync Device Manager, Model Number 24970A
- Z-0845-2019 · Jan 17, 2019 · TerminatedMedtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Z-0400-2018 · Jul 14, 2017 · TerminatedMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
- Z-2209-2010 · Jun 11, 2010 · TerminatedMedtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w
Other Pacemaker/icd/crt non-implanted components devices
- Gallant™ CDDRA500QSt. Jude Medical, Inc. · Pacemaker/icd/crt non-implanted components2 recalls33 deaths3,028 reports
- NA 229047Medtronic, Inc. · Pacemaker/icd/crt non-implanted components1 recall662 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls559 reports
- CareLink Encore™ 29901AMedtronic, Inc. · Pacemaker/icd/crt non-implanted components2 recalls474 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls422 reports
- Medtronic CareLink® Programmer 2090Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls151 reports
- Medtronic CareLink® Programmer 2090AA4Medtronic, Inc. · Pacemaker/icd/crt non-implanted components3 recalls82 reports
- Entrant™ CDDRA300QSt. Jude Medical, Inc. · Pacemaker/icd/crt non-implanted components1 recall2 deaths158 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page