Skip to content

MyCareLink™ 24950L (UDI 00643169552272)Recalls & reported problems

Medtronic, Inc. · MON 24950L USA MKT M10.2 USA MKT · Class 3

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This MyCareLink 24950L remote monitor is a bedside component used with Medtronic pacemakers and ICDs. Most of the 56 reports describe the monitor or its reader getting too hot, smoking, or throwing sparks; a few describe a burning smell or shocks patients felt were delivered. No deaths and no injuries were reported. There are no recalls for this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
56
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 56 adverse event reports that list this exact device (DI 00643169552272), received between Aug 28, 2019 and Jun 19, 2026 — classified as 56 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (49 reports).

10 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Overheating of Device49
  • Device Emits Odor4
  • Inappropriate/Inadequate Shock/Stimulation3

Patient problems

  • No Clinical Signs, Symptoms or Conditions44
  • No Consequences Or Impact To Patient9
  • No Known Impact Or Consequence To Patient4
  • Burn(s)1
  • Burning Sensation1

Recent adverse event reports

The 25 most recent of 56 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 19, 2026Inappropriate/Inadequate Shock/Stimulation

    IT WAS REPORTED THAT THE PATIENT WAS BEING SHOCKED AND ALLEGED THE MONITOR CAUSED IT. NO TROUBLESHOOTING TAKEN. THE PATIENT WAS ADVISED THAT THE BEDSIDE MONITOR DOES NOT INDUCE ANY THERAPIES BUT RECOMMENDED GETTING THE DEVICE CHECKED AT THE CLINIC. THE MONITOR REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25560133 ↗
  • MalfunctionMar 29, 2024Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR STARTED TO SMOKE AFTER THE PATIENT PLUGGED IT BACK IN. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19006917 ↗
  • MalfunctionAug 29, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17644426 ↗
  • MalfunctionMar 24, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16612495 ↗
  • MalfunctionFeb 24, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR EXPLODED OR THE PATIENT SAW SOME SPARKS FROM IT. THE REMOTE MONITOR WAS DISCONNECTED AND CALLED THE CLINIC. NO FURTHER TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16438746 ↗
  • MalfunctionFeb 10, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR OVERHEATED. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16352893 ↗
  • MalfunctionJan 3, 2023Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS HOT THAT IT ALMOST DAMAGED THE WOOD TABLE. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16086461 ↗
  • MalfunctionDec 1, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE READER OF THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15894983 ↗
  • MalfunctionOct 11, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. THE PATIENT UNPLUGGED THE REMOTE MONITOR AND NO FURTHER TROUBLESHOOTING WAS MADE DURING THE CALL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15579538 ↗
  • MalfunctionAug 16, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR FELT WARM THAT THE PATIENT THOUGHT IT WAS A FIRE HAZARD. NO TROUBLESHOOTING WAS TAKEN. THE PATIENT WANTED TO SEE IF THE MONITOR BEHAVIOR WOULD CHANGE OVER THE NEXT FEW DAYS. THE MONITOR REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15237143 ↗
  • MalfunctionJul 27, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE POWER CORD OF THE REMOTE MONITOR WAS VERY HOT AND THE CALLER THOUGHT IT WAS GOING TO START A FIRE. IT WAS NOTED THAT THE MONITOR WAS UNPLUGGED. NO FURTHER TROUBLESHOOTING WAS PERFORMED. THE POWER CORD WILL BE REPLACED. THE MONITOR REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15114700 ↗
  • MalfunctionJul 22, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING AND THE PATIENT UNPLUGGED IT. NO TROUBLESHOOTING TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15088229 ↗
  • MalfunctionJun 1, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS HOT . THE PATIENT ALSO EXPERIENCED SPARKING FROM THE REMOTE MONITOR THAT MAY HAVE BEEN COMING FROM ITS POWER CORD. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE REMOTE MONITOR IS EXPECTED TO BE RETURNED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14569426 ↗
  • MalfunctionJun 1, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE READER OF THE REMOTE MONITOR WAS BURNING THE PATIENT'S SKIN. THE CALLER WAS ASKED TO UNPLUG THE MONITOR. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Burning Sensation

    MAUDE report 14569134 ↗
  • MalfunctionMay 11, 2022Device Emits Odor; Device Emits Odor

    IT WAS REPORTED THAT THE REMOTE MONITOR SMELLED BURNT. NO TROUBLESHOOTING WAS INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14364320 ↗
  • MalfunctionApr 15, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS ASSOCIATED IN A HOUSE FIRE. NO TROUBLESHOOTING TAKEN. THE MONITOR WAS REPLACED AND THE REMOTE MONITOR RETURN STATUS WAS UNKNOWN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14118181 ↗
  • MalfunctionMar 8, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THERE WERE SPARKS THAT CAME OUT OF THE REMOTE MONITOR WHICH LANDED ON THE BED AND BURNED THE HOUSE DOWN. NO TROUBLESHOOTING WAS PERFORMED. THE MONITOR WILL NOT BE RETURNED THIS TIME. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13694328 ↗
  • MalfunctionFeb 14, 2022Overheating of Device; Overheating of Device

    THE MONITOR WAS RETURNED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13524907 ↗
  • MalfunctionFeb 9, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE POWER CORD OF THE REMOTE MONITOR GOT A CUT OR NICK IN IT AND THE CALLER STATED THAT THERE WERE SPARKS WHEN IT WAS PLUGGED IN. THERE WAS NO KNOWN CAUSE FOR THE CUT OR NICK. EXPLAINED TO THE CALLER TO LEAVE THE REMOTE MONITOR UNPLUGGED. A POWER CORD WILL BE SENT. THE MONITOR REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13502832 ↗
  • MalfunctionJan 13, 2022Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS GETTING WARM OR OVERHEATING. PATIENT HAD UNPLUGGED THE REMOTE MONITOR. NO FURTHER TROUBLESHOOTING WAS TAKEN. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13250165 ↗
  • MalfunctionNov 19, 2021Device Emits Odor; Device Emits Odor

    THE MONITOR WAS RETURNED AND LOST IN SHIPPING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12849130 ↗
  • MalfunctionOct 26, 2021Overheating of Device; Overheating of Device

    THE MONITOR WAS REPLACED AND RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12701908 ↗
  • MalfunctionSep 30, 2021Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. THE MONITOR WAS UNPLUGGED. NO FURTHER TROUBLESHOOTING WAS PERFORMED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12555514 ↗
  • MalfunctionSep 22, 2021Overheating of Device; Overheating of Device

    THE MONITOR WAS RETURNED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12511927 ↗
  • MalfunctionAug 30, 2021Overheating of Device; Overheating of Device

    THE REMOTE MONITOR WAS REPLACED AND RETURNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12387822 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.

  • Z-2168-2025 · Jun 18, 2025 · Open, Classified
    Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
  • Z-2169-2025 · Jun 18, 2025 · Open, Classified
    Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-2505-2020 · Jun 4, 2020 · Open, Classified
    CareLink SmartSync Device Manager, Model Number 24970A
  • Z-2506-2020 · Jun 4, 2020 · Open, Classified
    Patient Connector, Model Number 24967
  • Z-0845-2019 · Jan 17, 2019 · Terminated
    Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
  • Z-0399-2018 · Jul 14, 2017 · Terminated
    MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with
  • Z-0400-2018 · Jul 14, 2017 · Terminated
    MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
  • Z-2209-2010 · Jun 11, 2010 · Terminated
    Medtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w

Other Pacemaker/icd/crt non-implanted components devices

All Pacemaker/icd/crt non-implanted components devices →