INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Recalls & reported problems
Medtronic Sofamor Danek, Inc. · PARTIAL THREAD MAS 360PT4022 4.0 X 22MM · Class 2 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
1 FDA recall record name this device's identifier, 1 still open.
4 other recalls cover devices cleared under the same FDA submission (K163375); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Feb 16, 2023Open, Classified
Incorrect type of titanium was used which reduces the gripping strength of the screw head of 5 batches of Infinity Multi-Axial Screws
Z-1411-2023 · Root cause: Under Investigation by firm · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K163375. One submission often covers many models, so check the product and lot details in each record.
- Z-1410-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion
- Z-1409-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion
- Z-1413-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion
- Z-1412-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion
Other Orthosis, cervical pedicle screw spinal fixation devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- SYMPHONY 1020-35-112Medos International Sàrl · Posterior cervical screw system6 deaths15 reports
- YUKON OCT Spinal System 7601-05030VB Spine LLC · Orthosis, cervical pedicle screw spinal fixation15 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.