YUKON OCT Spinal System 7601-05030Recalls & reported problems

VB Spine LLC · Polyaxial Screw Ø5.0x30 mm · Class 2 · Implant

All models of YUKON OCT Spinal System (54)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
15
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 15 adverse event reports that list this exact device (DI 10888857350021), received between Jul 17, 2019 and Jul 10, 2025 — classified as 2 injuries, 13 malfunctions.

The most frequently reported specific device problem is “Fracture” (12 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture12
  • Material Separation2
  • Break1
  • Device Dislodged or Dislocated1
  • Insufficient Device Problem Information1

Patient problems

  • No Clinical Signs, Symptoms or Conditions10
  • No Consequences Or Impact To Patient3
  • Pain2
  • Injury1
  • No Information1

Recent adverse event reports

The 15 most recent of 15 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 10, 2025Fracture; Fracture

    NO NEW INFORMATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22470502 ↗
  • MalfunctionDec 8, 2023Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT DURING ROUTINE POST-OPERATIVE FOLLOW-UP, IT WAS DISCOVERED THAT ONE OF THE YUKON OCT POLYAXIAL SCREWS FRACTURED AT T1 APPROXIMATELY THREE MONTHS AFTER IMPLANTATION. REVISION SURGERY IS NOT CURRENTLY PLANNED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18297560 ↗
  • MalfunctionNov 9, 2022Fracture; Fracture

    A COMPANY REPRESENTATIVE REPORTED THAT THE TULIP HEAD OF A YUKON OCT POLYAXIAL SCREW DETACHED FROM THE SCREW SHANK INTRA-OPERATIVELY. THE BROKEN SCREW WAS REMOVED AND REPLACED WITH ANOTHER SCREW OF THE SAME SIZE. THE SURGERY WAS COMPLETED SUCCESSFULLY AFTER A 30-MINUTE SURGICAL DELAY. NO ADVERSE CONSEQUENCE TO THE PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15759744 ↗
  • MalfunctionOct 29, 2022Fracture; Fracture

    THIS RECORD IS A CONSOLIDATION OF Q3 2022 RECORDS SUMMARIZED AS A PART OF THE FDA VOLUNTARY MALFUNCTION SUMMARY REPORTING PROGRAM. THIS REPORT CAPTURES ONE INSTANCE OF INTRAOPERATIVE YUKON POLYAXIAL SCREW FRACTURE. THE ASSOCIATED PROCEDURE WAS COMPLETED SUCCESSFULLY BY REPLACING THE FRACTURED SCREW. THERE WAS A 30-MINUTE SURGICAL DELAY AND NO ADVERSE CONSEQUENCE TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15695227 ↗
  • MalfunctionJul 29, 2021Fracture; Fracture

    THIS REPORT SUMMARIZES TWO MALFUNCTION EVENTS WHERE YUKON OCT POLYAXIAL SCREWS FRACTURED INTRA-OPERATIVELY. SURGERIES WERE SUCCESSFULLY COMPLETED WITH A 15 MINUTE DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12247668 ↗
  • MalfunctionJan 28, 2021Fracture; Fracture

    THIS REPORT SUMMARIZES THREE MALFUNCTION EVENTS WHERE YUKON OCT POLYAXIAL SCREWS FRACTURED INTRA-OPERATIVELY. SURGERY WAS SUCCESSFULLY COMPLETED AND NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11247964 ↗
  • MalfunctionOct 30, 2020Fracture; Fracture

    THIS REPORT SUMMARIZES <NOE> 1 </NOE> MALFUNCTION EVENT WHERE A YUKON OCT POLYAXIAL SCREW BROKE UPON INSERTION INTO THE T1 PEDICLE INTRA-OPERATIVELY. SURGERY WAS SUCCESSFULLY COMPLETED WITH ANOTHER SCREW AND A FIVE MINUTE DELAY. NO ADVERSE CONSEQUENCES OR MEDICAL INTERVENTION WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10764756 ↗
  • MalfunctionOct 29, 2020Material Separation; Device Dislodged or Dislocated; Material Separation; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A YUKON POLYAXIAL SCREW TULIP DISENGAGED POST-OPERATIVELY AT T1. REVISION SURGERY IS NOT SCHEDULED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10756871 ↗
  • MalfunctionOct 19, 2020Fracture; Fracture

    A PHYSICIAN REPORTED A FRACTURED YUKON POLYAXIAL SCREW POST-OPERATIVELY. THE PATIENT HAS REPORTED "HEARING A CLICKING NOISE PERIODICALLY." REVISION SURGERY IS NOT PLANNED AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10701606 ↗
  • MalfunctionOct 8, 2020Material Separation; Material Separation

    A PHYSICIAN REPORTED THAT TWO YUKON SCREW TULIPS DISENGAGED APPROXIMATELY SIX MONTHS AFTER IMPLANTATION. REVISION STATUS IS CURRENTLY UNKNOWN. THIS REPORT REPRESENTS THE FIRST OF THE TWO SCREWS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10648759 ↗
  • MalfunctionMar 27, 2020Fracture; Fracture

    IT WAS REPORTED THAT 2 YUKON OCT SPINAL SYSTEM POLYAXIAL SCREWS FRACTURED 6 WEEKS POST-OPERATIVELY. THE DEVICE REMAINS IMPLANTED. THIS RECORD REPRESENTS SCREW 2 OF 2.

    Patient problems: No Consequences Or Impact To Patient; Pain

    MAUDE report 9891845 ↗
  • MalfunctionMar 27, 2020Fracture; Fracture

    IT WAS REPORTED THAT 2 YUKON OCT SPINAL SYSTEM POLYAXIAL SCREWS FRACTURED 6 WEEKS POST-OPERATIVELY. THE DEVICE REMAINS IMPLANTED. THIS RECORD REPRESENTS SCREW 1 OF 2.

    Patient problems: No Consequences Or Impact To Patient; Pain

    MAUDE report 9891796 ↗
  • MalfunctionMar 24, 2020Fracture; Fracture

    IT WAS REPORTED THAT A YUKON OCT SPINAL SYSTEM POLYAXIAL SCREW FRACTURED INTRA-OPERATIVELY DURING INSERTION. SURGERY WAS SUCCESSFULLY COMPLETED WITH AN UNKNOWN SURGICAL DELAY USING A REPLACEMENT SCREW.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9875046 ↗
  • InjurySep 9, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    A PHYSICIAN REPORTED THAT THE SHAFT OF A YUKON POLYAXIAL SCREW BROKE DURING IMPLANT SURGERY. THE BROKEN DEVICE WAS REMOVED AND REPLACED WITH A 15 MINUTE SURGICAL DELAY. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE. THERE WAS NO CONSEQUENCE TO THE PATIENT.

    Patient problems: No Information

    MAUDE report 8977600 ↗
  • InjuryJul 17, 2019Fracture; Break; Fracture

    IT WAS REPORTED THAT THE POLYAXIAL SCREW BROKE POST OPERATIVELY; PATIENT UNDERWENT REVISION SURGERY.

    Patient problems: Injury

    MAUDE report 8798310 ↗

Other Orthosis, cervical pedicle screw spinal fixation devices