INFINITY™ Occipitocervical Upper Thoracic System 3604012Recalls & reported problems
Medtronic Sofamor Danek, Inc. · MULTI AXIAL SCREW 3604012 4.0 X 12MM · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169554627), received between Feb 26, 2020 and Feb 26, 2020 — classified as 6 injuries.
The most frequently reported specific device problem is “Migration” (6 reports).
5 other recalls cover devices cleared under the same FDA submission (K163375); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Migration6
Patient problems
- Injury6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryFeb 26, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR CERVICAL FUSION (PCF) FROM C3 TO T2. POST-OP, THE LATERAL MASS SCREWS PULLED OUT OF C3, C4, AND C5. THIS LEFT THE PATIENT'S NECK UNSTABLE. THE SURGEON SAID THAT THE PATIENT DID NOT FALL AND HAS BEEN WEARING A COLLAR THROUGHOUT POST-OP RECOVERY. THE PATIENT WILL REQUIRE REVISION SURGERY AS A RESULT OF THIS EVENT ON A LATER DATE.
Patient problems: Injury
MAUDE report 9754863 ↗InjuryFeb 26, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR CERVICAL FUSION (PCF) FROM C3 TO T2. POST-OP, THE LATERAL MASS SCREWS PULLED OUT OF C3, C4, AND C5. THIS LEFT THE PATIENT'S NECK UNSTABLE. THE SURGEON SAID THAT THE PATIENT DID NOT FALL AND HAS BEEN WEARING A COLLAR THROUGHOUT POST-OP RECOVERY. THE PATIENT WILL REQUIRE REVISION SURGERY AS A RESULT OF THIS EVENT ON A LATER DATE.
Patient problems: Injury
MAUDE report 9754856 ↗InjuryFeb 26, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR CERVICAL FUSION (PCF) FROM C3 TO T2. POST-OP, THE LATERAL MASS SCREWS PULLED OUT OF C3, C4, AND C5. THIS LEFT THE PATIENT'S NECK UNSTABLE. THE SURGEON SAID THAT THE PATIENT DID NOT FALL AND HAS BEEN WEARING A COLLAR THROUGHOUT POST-OP RECOVERY. THE PATIENT WILL REQUIRE REVISION SURGERY AS A RESULT OF THIS EVENT ON A LATER DATE.
Patient problems: Injury
MAUDE report 9754849 ↗InjuryFeb 26, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR CERVICAL FUSION (PCF) FROM C3 TO T2. POST-OP, THE LATERAL MASS SCREWS PULLED OUT OF C3, C4, AND C5. THIS LEFT THE PATIENT'S NECK UNSTABLE. THE SURGEON SAID THAT THE PATIENT DID NOT FALL AND HAS BEEN WEARING A COLLAR THROUGHOUT POST-OP RECOVERY. THE PATIENT WILL REQUIRE REVISION SURGERY AS A RESULT OF THIS EVENT ON A LATER DATE.
Patient problems: Injury
MAUDE report 9754848 ↗InjuryFeb 26, 2020Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT POSTERIOR CERVICAL FUSION (PCF) FROM C3 TO T2. POST-OP, THE LATERAL MASS SCREWS PULLED OUT OF C3, C4, AND C5. THIS LEFT THE PATIENT'S NECK UNSTABLE. THE SURGEON SAID THAT THE PATIENT DID NOT FALL AND HAS BEEN WEARING A COLLAR THROUGHOUT POST-OP RECOVERY. THE PATIENT WILL REQUIRE REVISION SURGERY AS A RESULT OF THIS EVENT ON A LATER DATE.
Patient problems: Injury
MAUDE report 9754833 ↗InjuryFeb 26, 2020Migration; Migration
THE REVISION SURGERY TOOK PLACE ON (B)(6) 2020. THE REVISION SURGERY WENT WELL, AND THE SURGEON WAS ABLE TO ANCHOR NEW SCREWS FOR A STRONGER CONSTRUCT. WE HAVE NOT HEARD OF ANY COMPLICATIONS SINCE THE REVISION.
Patient problems: Injury
MAUDE report 9754687 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K163375. One submission often covers many models, so check the product and lot details in each record.
- Z-1410-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 32mm, REF 3604532; for immobilization and stabilization of spinal segment as an adjunct to fusion
- Z-1412-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 26mm, REF 3604026; for immobilization and stabilization of spinal segment as an adjunct to fusion
- Z-1409-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 24mm, REF 3604024; for immobilization and stabilization of spinal segment as an adjunct to fusion
- Z-1411-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, PARTIALLY THREADED MULTI AXIAL SCREW, Size: 4.0mm x 22mm, REF 360PT4022; for immobilization and stabilization of spinal segment as
- Z-1413-2023 · Feb 16, 2023 · Open, ClassifiedINFINITY Occipitocervical Upper Thoracic System, MULTI AXIAL SCREW, Size: 4.0mm x 34mm, REF 3604034; for immobilization and stabilization of spinal segment as an adjunct to fusion
Other Orthosis, cervical pedicle screw spinal fixation devices
- INFINITY™ Occipitocervical Upper Thoracic System 3604026Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation2 recalls1 report
- CONNECTOR 79-2003Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604024Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- CONNECTOR 79-2002Orthofix US LLC · Posterior cervical screw system1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 3604532Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- INFINITY™ Occipitocervical Upper Thoracic System 360PT4022Medtronic Sofamor Danek, Inc. · Orthosis, cervical pedicle screw spinal fixation1 recall0 reports
- OASYS 48551000Stryker Corporation · Orthosis, cervical pedicle screw spinal fixation45 reports
- SYMPHONY 1020-35-112Medos International Sàrl · Posterior cervical screw system6 deaths15 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page