MOSAIC ULTRA® 305U225 (UDI 00643169595026)Recalls & reported problems
Medtronic, Inc. · VLV 305U225 MOSAIC AORULT CII NNAUS PEEK · Class 3 · Implant
All models of MOSAIC ULTRA (40)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic Mosaic Ultra aortic valve is a tissue heart valve implanted to replace a diseased aortic valve. The 45 reports mostly describe the valve narrowing or leaking over time, sometimes with heavy calcium buildup, and patients being evaluated for a valve-in-valve replacement years after implantation; other reports mention endocarditis or the need for a pacemaker. Reports include 1 death and 41 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 45
- Deaths reported
- 1
- Injuries reported
- 41
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 45 adverse event reports that list this exact device (DI 00643169595026), received between Mar 14, 2017 and Apr 17, 2026 — classified as 1 death, 41 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (25 reports).
3 other recalls cover devices cleared under the same FDA submission (P990064); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information25
- Device Stenosis7
- Intravalvular regurgitation7
- Patient Device Interaction Problem6
- Adverse Event Without Identified Device or Use Problem5
- Gradient Increase4
- Perivalvular Leak4
- Calcified3
Patient problems
- Aortic Valve Insufficiency/ Regurgitation12
- Aortic Valve Stenosis11
- Insufficient Information9
- No Information8
- Endocarditis4
- No Known Impact Or Consequence To Patient4
- Calcium Deposits/Calcification3
- Non specific EKG/ECG Changes3
Recent adverse event reports
The 25 most recent of 45 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 17, 2026Gradient Increase; Patient Device Interaction Problem; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSESOPHAGEAL ECHOCARDIOGRAM (TEE) AND CARDIAC CATHETERIZATION PERFORMED NOTED INCREASING GRADIENTS, MODERATE CORONARY ARTERY DISEASE AND SEVERE AORTIC STENOSIS. AN ELECTROCARDIOGRAM PERFORMED SHOWED NORMAL SINUS RHYTHM WITH NON-SPECIFIC ST AND T WAVE ABNORMALITY. NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Non specific EKG/ECG Changes
MAUDE report 24913760 ↗InjuryJan 16, 2026Device Stenosis; Intravalvular regurgitation
IT WAS REPORTED THAT 9 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC INSUFFICIENCY AND AORTIC STENOSIS. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 24092598 ↗InjuryDec 3, 2025Patient Device Interaction Problem; Patient Device Interaction Problem
IT WAS REPORTED THAT 6 YEARS AND 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SINUS BRADYCARDIA-TACHYCARDIA SYNDROME. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Bradycardia; Tachycardia
MAUDE report 23703622 ↗MalfunctionDec 2, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
IT WAS REPORTED THAT 7 YEARS AND 8 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 23690809 ↗InjuryNov 24, 2025Calcified; Device Stenosis; Calcified; Device Stenosis
ADDITIONAL INFORMATION WAS RECEIVED THAT REASON FOR INTERVENTION WAS STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Calcium Deposits/Calcification
MAUDE report 23639580 ↗InjuryNov 18, 2025Calcified; Gradient Increase; Thickening of Material; Device Stenosis; Intravalvular regurgitation; Calcified; Gradient Increase; Thickening of Material; Device Stenosis; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM MOSAIC AORTIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS MODERATE CALCIFICATION AND THICKENING OF THE AORTIC VALVE, MODERATE TO SEVERE AORTIC VALVE STENOSIS, HIGH PEAK AORTIC VELOCITY (4.05 M/S), HIGH CALCULATED PEAK GRADIENT (63 MMHG), HIGH MEAN GRADIENT (43.0 MMHG), REDUCED AORTIC VALVE AREA (0.9 CM²), TRACE (TRIVIAL) AORTIC VALVE REGURGITATION AND SHORTNESS OF BREATH. TRANSTHORACIC ECHOCARDIOGRAM WAS CONDUCTED TO ASSESS THESE FINDINGS. NO EVIDENCE OF THROMBUS WAS OBSERVED ON THE BIOPROSTHETIC VALVE. INTERVENTION IS REQUIRED AND NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis; Calcium Deposits/Calcification; Dyspnea
MAUDE report 23581100 ↗MalfunctionJun 30, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 9 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS NOT REPORTED. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 22360169 ↗InjuryFeb 25, 2025Insufficient Device Problem Information; Device Stenosis; Insufficient Device Problem Information; Device Stenosis
ADDITIONAL INFORMATION WAS RECEIVED THAT REPORTED SUBSEQUENTLY, 24 DAYS LATER, THE VALVE WAS EXPLANTED AND REPLACED WITH AN UNKNOWN PRODUCT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Insufficient Information
MAUDE report 21465454 ↗InjuryOct 31, 2024Material Split, Cut or Torn; Intravalvular regurgitation; Material Split, Cut or Torn; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 8 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED. SUBSEQUENTLY, 1 DAY LATER, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. IT WAS ALSO REPORTED THAT THE RIGHT CORONARY CUSP (RCC) LEAFLET WAS PROLAPSED AND THE NON-CORONARY CUSP (NCC) LEAFLET WAS TORN. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Cusp Tear; Prolapse
MAUDE report 20582527 ↗InjuryOct 30, 2024Calcified; Patient Device Interaction Problem; Intravalvular regurgitation; Calcified; Patient Device Interaction Problem; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 6 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS SEVERE AORTIC REGURGITATION (AR) AND AN INFECTION. IT WAS STATED THAT VEGETATION WAS OBSERVED ON THE VALVE AND THAT THE PATIENT WAS HOSPITALIZED AND ANTIBIOTICS WERE ADMINISTERED. IT WAS STATED THAT A POSSIBLE PSEUDOANEURYM WAS OBSERVED CLOSE TO THE AREA OF THE NON-CORONARY CUSP (NCC) AND CALCIFICATION WAS OBSERVED BELOW THE LEFT CORONARY CUSP (LCC) EXTENDING INTO THE LEFT VENTRICULAR OUTFLOW TRACT (LVOT). NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Calcium Deposits/Calcification; Unspecified Infection
MAUDE report 20565017 ↗InjuryOct 10, 2024Insufficient Device Problem Information; Patient Device Interaction Problem; Intravalvular regurgitation; Insufficient Device Problem Information; Patient Device Interaction Problem; Intravalvular regurgitation
MEDTRONIC RECEIVED ADDITIONAL INFORMATION WHICH STATED THAT THE REASON FOR THE REPLACEMENT WAS AORTIC VALVE ENDOCARDITIS. IT WAS STATED THAT MILD TO MODERATE AORTIC REGURGITATION (AR) WAS ALSO PRESENT. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Endocarditis; Insufficient Information
MAUDE report 20418937 ↗InjurySep 9, 2024Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 10 DAYS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A PERMANENT PACEMAKER WAS IMPLANTED. THE REASON FOR THE PERMANENT PACEMAKER WAS REPORTED AS SINUS NODE DYSFUNCTION. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Non specific EKG/ECG Changes
MAUDE report 20176605 ↗InjurySep 3, 2024Intravalvular regurgitation; Intravalvular regurgitation
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 10 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 20126521 ↗InjuryJun 17, 2024Gradient Increase; Device Stenosis; Gradient Increase; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 8 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS (AS) WITH A MEAN GRADIENT OF 43MMHG. IT HAD ALSO BEEN REPORTED THAT THE PATIENT HAD BEEN EXPERIENCING SHORTNESS OF BREATH (SOB). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis; Dyspnea
MAUDE report 19545930 ↗InjuryJun 11, 2024Device Stenosis; Device Stenosis
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 9 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS SEVERE AORTIC STENOSIS (AS). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 19510291 ↗InjuryApr 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 7 YEARS AND 7 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM AORTIC BIOPROSTHETIC VALVE OF A DIFFERENT MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 19106758 ↗InjuryFeb 15, 2024Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 5 YEARS AND 10 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH AN UNKNOWN DEVICE. THE REASON FOR THE REPLACEMENT WAS REPORTED AS THE VALVE WAS NON FUNCTIONING. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 18709855 ↗InjuryDec 18, 2023Insufficient Device Problem Information; Intravalvular regurgitation; Insufficient Device Problem Information; Intravalvular regurgitation
MEDTRONIC RECEIVED ADDITIONAL INFORMATION WHICH CONFIRMED THAT THE REASON FOR INTERVENTION WAS SEVERE AORTIC INSUFFICIENCY (AI). NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Insufficient Information
MAUDE report 18347005 ↗InjuryNov 15, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 2 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A FAMILY MEMBER STATED THE PATIENT IS BEING EVALUATED FOR A SEPSIS INFECTION AND THE PATIENT HAD MANY UNSPECIFIED "TESTS AND PROCEDURES AND LOST ALMOST 50 POUNDS" AND "FLATLINED TWICE". NO INTERVENTION OR ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Sepsis; Weight Changes
MAUDE report 18139684 ↗InjurySep 7, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 12 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, THE PATIENT IS BEING EVALUATED FOR A TRANSCATHETER VALVE-IN-VALVE REPLACEMENT. THE REASON FOR EVALUATION WAS REPORTED AS AORTIC STENOSIS AND SEVERE AORTIC REGURGITATION. NO INTERVENTION OR ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 17703254 ↗InjuryJan 9, 2023Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 6 YEARS AND 4 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A 26MM TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 16122122 ↗InjuryJan 26, 2022Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT IN APPROXIMATELY 5 YEARS AND 6 MONTHS POST IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, A TR ANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS AORTIC STENOSIS AND MILD/MODERATE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation; Aortic Valve Stenosis
MAUDE report 13361197 ↗InjuryAug 9, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 4 YEARS AND 11 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE DUE TO SEVERE STENOSIS. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Stenosis
MAUDE report 12295426 ↗InjuryJun 23, 2021Insufficient Device Problem Information; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT 3 YEARS AND 4 MONTHS POST IMPLANT OF THIS AORTIC BIOPROSTHETIC VALVE, A TRANSCATHETER VALVE WAS IMPLANTED VALVE-IN-VALVE. THE REASON FOR INTERVENTION WAS REPORTED AS MODERATE TO SEVERE AORTIC INSUFFICIENCY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Aortic Valve Insufficiency/ Regurgitation
MAUDE report 12050943 ↗InjuryMar 10, 2021Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information; Inadequacy of Device Shape and/or Size; Insufficient Device Problem Information
MEDTRONIC RECEIVED INFORMATION THAT DURING THE IMPLANT OF THIS 25MM AORTIC BIOPROSTHETIC VALVE, IT WAS EXPLANTED AND REPLACED WITH A 23MM VALVE OF THE SAME MODEL. THE REASON FOR THE REPLACEMENT WAS NOT REPORTED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Patient problems: Insufficient Information
MAUDE report 11449692 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P990064. One submission often covers many models, so check the product and lot details in each record.
- Z-1881-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C31 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1880-2019 · May 21, 2019 · TerminatedMosaic 310 Bioprosthesis, Mitral, REF 310C29 - Product Usage: Replacement of malfunctioning native or prosthetic aortic and/or mitral valves.
- Z-1225-2009 · Jan 26, 2009 · TerminatedMedtronic Mosaic Porcine Bioprostheses with Cinch, Heart Valve; Model 305; Catalog #s 305C21AA, 305C2701, 305C2301, 305C2501, and 305C2901; Sterile LC; Rx only; Medtronic Inc 710 M
Other heart-valve, non-allograft tissue devices
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue2 deaths404 reports
- INSPIRIS RESILIA AORTIC VALVE 11500AEdwards Lifesciences LLC · Heart-Valve, Non-Allograft Tissue385 reports
- PERCEVAL PVS23Corcym Canada Corp · heart-valve, non-allograft tissue26 deaths337 reports
- MOSAIC® 305C23Medtronic, Inc. · heart-valve, non-allograft tissue327 reports
- MOSAIC® 305C25Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths324 reports
- FREESTYLE® FR995-27Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths318 reports
- FREESTYLE® FR995-29Medtronic, Inc. · heart-valve, non-allograft tissue288 reports
- FREESTYLE® FR99525Medtronic, Inc. · heart-valve, non-allograft tissue2 deaths280 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page