Skip to content

Attain Ability™ Straight MRI SureScan™ 439678 (UDI 00643169601802)Recalls & reported problems

Medtronic, Inc. · LEAD 439678 MRI STRAIGHT US · Class 3 · Implant

All models of Attain Ability Straight MRI SureScan (4)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169601802), received between Apr 6, 2017 and Aug 18, 2025 — classified as 1 injury, 5 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • High Capture Threshold3
  • Device Dislodged or Dislocated1
  • Failure to Capture1
  • Fracture1
  • Intermittent Capture1
  • Pacing Problem1
  • Unstable Capture Threshold1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Complaint, Ill-Defined1
  • No Clinical Signs, Symptoms or Conditions1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2025Intermittent Capture; Intermittent Capture

    IT WAS REPORTED THAT THE LEFT VENTRICULAR (LV) LEAD EXHIBITED INTERMITTENT LOSS OF CAPTURE. THE LEAD REMAINS IN USE. NO PATIENT COMP LICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22835292 ↗
  • MalfunctionJul 13, 2020High Capture Threshold; High Capture Threshold

    IT WAS REPORTED THAT THERE WAS A HIGH THRESHOLD ON THE LEFT VENTRICULAR (LV) LEAD. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 10264925 ↗
  • MalfunctionApr 25, 2018Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT DURING AN IMPLANT PROCEDURE, THE LEFT VENTRICULAR LEAD DISLODGED. THE LEAD WAS NOT USED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7459996 ↗
  • InjuryAug 9, 2017Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT WAS SYMPTOMATIC AND WAS SEEN IN THE CLINIC. INTERROGATION OF THE DEVICE FOUND ANODAL STIMULATION. PROGRAMMING CHANGES WERE MADE AND THE LEFT VENTRICULAR LEAD REMAINS IN USE. THE PATIENT IS A PARTICIPANT IN THE PRODUCT SURVEILLANCE REGISTRY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Complaint, Ill-Defined

    MAUDE report 6783330 ↗
  • MalfunctionJun 8, 2017Failure to Capture; High Capture Threshold; Unstable Capture Threshold; Failure to Capture; High Capture Threshold; Unstable Capture Threshold

    IT WAS REPORTED THAT THE LEFT VENTRICULAR LEAD HAD HIGH THRESHOLD. THE LEAD WAS EXPLANTED AND DURING THE REPLACEMENT PROCEDURE, TWO LEADS WERE ATTEMPTED AND NOT USED. BOTH HAD NO CAPTURE, AND HIGH AND UNSTABLE THRESHOLD. ANOTHER LEAD WAS IMPLANTED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6627631 ↗
  • MalfunctionApr 6, 2017Fracture; High Capture Threshold; Fracture; High Capture Threshold

    IT WAS REPORTED THE LEFT VENTRICULAR (LV) LEAD WAS FRACTURED WHILE TYING IT DOWN AT THE SUTURE COLLAR. IT WAS FURTHER REPORTED THE CAPTURED THRESHOLD INCREASED AFTER TYING IT DOWN. THE LEAD WAS REMOVED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6469625 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →

Sources and updates