Medtronic, Inc. · LEAD 439888 MRI STRAIGHT US · Class 3 · Implant

All models of Attain Performa Straight MRI SureScan (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Attain Performa Straight MRI SureScan is a left ventricular pacing lead, a wire placed in the heart to help it beat. Its 1,576 reports describe many cases of the lead failing to reliably trigger heartbeats, and others mention pocket infections, dislodged leads or nerve stimulation. The reports mention 26 deaths and 1,068 injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
1,595
Deaths reported
26
Injuries reported
1,078
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1,595 adverse event reports that list this exact device (DI 00643169601857), received between Dec 9, 2016 and Sep 30, 2026 — classified as 26 deaths, 1,078 injuries, 491 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (393 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 13 reportsDec 2023: 20 reportsJan 2024: 15 reportsFeb 2024: 22 reportsMar 2024: 23 reportsApr 2024: 17 reportsMay 2024: 16 reportsJun 2024: 16 reportsJul 2024: 13 reportsAug 2024: 18 reportsSep 2024: 10 reportsOct 2024: 16 reportsNov 2024: 15 reportsDec 2024: 6 reportsJan 2025: 20 reportsFeb 2025: 13 reportsMar 2025: 14 reportsApr 2025: 9 reportsMay 2025: 12 reportsJun 2025: 11 reportsJul 2025: 16 reportsAug 2025: 20 reportsSep 2025: 10 reportsOct 2025: 10 reportsNov 2025: 15 reportsDec 2025: 15 reportsJan 2026: 17 reportsFeb 2026: 14 reportsMar 2026: 13 reportsApr 2026: 8 reportsMay 2026: 10 reportsJun 2026: 8 reportsJul 2026: 8 reportsAug 2026: 11 reportsSep 2026: 19 reportsOct 2026: 0 reports20242025202623
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem668
  • High Capture Threshold393
  • Failure to Capture195
  • Capturing Problem174
  • Device Dislodged or Dislocated133
  • High impedance77
  • Appropriate Device Problem Term/Code Not Available63
  • Activation, Positioning or Separation Problem61

Patient problems

  • No Clinical Signs, Symptoms or Conditions545
  • Unspecified Infection352
  • No Known Impact Or Consequence To Patient181
  • Undesired Nerve Stimulation120
  • Device Overstimulation of Tissue113
  • Muscle Stimulation67
  • No Consequences Or Impact To Patient32
  • Sepsis32

Recent adverse event reports

The 25 most recent of 1,595 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Insufficient Device Problem Information

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952438 ↗
  • InjurySep 29, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50936943 ↗
  • MalfunctionSep 25, 2026Capturing Problem

    It was reported that the Left Ventricular (LV) lead and Right Ventricular (RV) lead exhibited rising pacing thresholds. The leads re main in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899304 ↗
  • MalfunctionSep 25, 2026Capturing Problem

    It was reported that the Left Ventricular (LV) lead exhibited rising thresholds. The LV lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50897441 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that approximately one month post replacement procedure, the patient developed a pocket infection. It was noted that the patient developed purulent discharge/pus. The Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted and replaced with a Leadless Implantable Pulse Generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Unspecified Infection

    MAUDE report 50877091 ↗
  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced bacteremia. The Cardiac Resynchronization Therapy Defibrillator (CRT-D) and leads were explanted and replaced with a leadless Implantable Pulse Generator (IPG). No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia

    MAUDE report 50868656 ↗
  • InjurySep 22, 2026Extracardiac stimulation

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50859213 ↗
  • InjurySep 18, 2026Fracture; Capturing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50829309 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50827665 ↗
  • InjurySep 17, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50811601 ↗
  • MalfunctionSep 17, 2026Low impedance

    It was reported that the Left Ventricular (LV) lead exhibited low pacing impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50803840 ↗
  • MalfunctionSep 15, 2026High Capture Threshold

    It was reported that the Left Ventricular (LV) lead exhibited high thresholds. It was also reported that the Right Atrial (RA) lead had position check failure. Both LV and RA lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50766879 ↗
  • MalfunctionSep 14, 2026Impedance Problem

    It was reported that the Right Ventricular (RV) lead exhibited an increase in pacing impedance and possible loss of capture, and the Left Ventricular (LV) lead exhibited an increase in pacing impedance. The leads remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50754905 ↗
  • MalfunctionSep 10, 2026Intermittent Capture; Pacing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50721374 ↗
  • MalfunctionSep 8, 2026High impedance; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50680306 ↗
  • InjurySep 8, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection

    MAUDE report 50679306 ↗
  • InjurySep 8, 2026Extracardiac stimulation

    Patient problems: Device Overstimulation of Tissue

    MAUDE report 50679157 ↗
  • MalfunctionSep 4, 2026Electrical /Electronic Property Problem; Impedance Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50613766 ↗
  • InjurySep 3, 2026Extracardiac stimulation

    It was reported by the patient that their Cardiac Resynchronization Therapy Pacemaker (CRT-P) has ¿faster than expected battery depletion¿ and they are experiencing dizziness and lightheadedness. Additionally, the patient noted that they previously had diaphragmatic stimulation and adjustments were made. The patient inquired if the adjustments may have affected their CRT-P and health. The CRT-P and Left Ventricular (LV) lead remain in use. No further patient complications have been reported as a result of this event.

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50590506 ↗
  • InjuryAug 31, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed a pocket infection approximately one month post-Cardiac Resynchronization Therapy Defibri llator (CRT-D) implant. The patient had also experienced purulent discharge/pus, redness and swelling. The organism was identified as Staphylococcus aureus. The CRT-D system was removed and a leadless pacemaker was implanted. No further patient complications have been reported as a result of this event.

    Patient problems: Bacterial Infection; Erythema; Medical device site infection; Purulent Discharge; Swelling/ Edema

    MAUDE report 50530530 ↗
  • MalfunctionAug 31, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50518627 ↗
  • InjuryAug 30, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Erosion; Unspecified Infection

    MAUDE report 50513032 ↗
  • InjuryAug 25, 2026Extracardiac stimulation

    It was reported that the patient was experiencing phrenic nerve stimulation from the Left Ventricular (LV) lead. The lead was explanted and replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50440709 ↗
  • InjuryAug 24, 2026Failure to Capture

    It was reported that the patient experienced syncope, a fall and was in a car accident. It was further reported that the Cardiac Res ynchronization Therapy Pacemaker (CRT-P) exhibited a pacing issue, the Right Ventricular (RV) lead exhibited high thresholds and intermittent capture  and that the Left Ventricular (LV) lead exhibited no capture.  The CRTP system remains in use. No further patient complications have been reported as a result of this event.

    Patient problems: Fall; Syncope/Fainting

    MAUDE report 50409273 ↗
  • InjuryAug 18, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient experienced a pocket infection that presented with discomfort, redness, and purulent discharge/pus. The Cardiac Resynchronization Therapy Pacemaker (CRT-P) system was explanted and a leadless pacemaker was implanted. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Erythema; Medical device site infection; Purulent Discharge

    MAUDE report 50308399 ↗

Other Drug eluting permanent left ventricular (lv) pacemaker electrode devices

All Drug eluting permanent left ventricular (lv) pacemaker electrode devices →