Sentus ProMRI OTW QP S-85/49 406082Recalls & reported problems

BIOTRONIK SE & Co. KG · Class 3 · Implant

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Biotronik Sentus heart pacing lead is a wire placed in the heart to help control its rhythm, and its 505 reports mostly describe the lead moving out of position. Some report fractures, failure to pace, and infections around the device. There is 1 death and 110 injuries reported, and no recalls name this device. Reports fell to 32 in the past year from 55 in the year before.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
511
Deaths reported
1
Injuries reported
115
Recalls naming it
0

Summary

The FDA's MAUDE database contains 511 adverse event reports that list this exact device (DI 04035479145606), received between Jun 15, 2018 and Sep 25, 2026 — classified as 1 death, 115 injuries, 395 malfunctions.

The most frequently reported specific device problem is “Device Dislodged or Dislocated” (298 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 4 reportsDec 2023: 1 reportsJan 2024: 6 reportsFeb 2024: 3 reportsMar 2024: 5 reportsApr 2024: 4 reportsMay 2024: 1 reportsJun 2024: 4 reportsJul 2024: 2 reportsAug 2024: 5 reportsSep 2024: 2 reportsOct 2024: 5 reportsNov 2024: 4 reportsDec 2024: 2 reportsJan 2025: 8 reportsFeb 2025: 5 reportsMar 2025: 4 reportsApr 2025: 6 reportsMay 2025: 6 reportsJun 2025: 2 reportsJul 2025: 2 reportsAug 2025: 11 reportsSep 2025: 3 reportsOct 2025: 2 reportsNov 2025: 7 reportsDec 2025: 3 reportsJan 2026: 6 reportsFeb 2026: 6 reportsMar 2026: 2 reportsApr 2026: 3 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202611
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Dislodged or Dislocated298
  • Adverse Event Without Identified Device or Use Problem88
  • Failure to Capture50
  • High impedance35
  • Fracture30
  • Ambient Noise Problem23
  • High Capture Threshold19
  • Over-Sensing14

Patient problems

  • No Clinical Signs, Symptoms or Conditions284
  • No Known Impact Or Consequence To Patient117
  • Unspecified Infection67
  • Undesired Nerve Stimulation14
  • Twiddlers Syndrome7
  • Muscle Stimulation6
  • Pocket Erosion6
  • Vascular Dissection4

Recent adverse event reports

The 25 most recent of 511 reports, in the reporter's own words as published by the FDA.

  • InjurySep 25, 2026Device Dislodged or Dislocated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50903028 ↗
  • InjurySep 25, 2026Device Dislodged or Dislocated

    This lead was explanted due to dislodgement. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50899520 ↗
  • InjurySep 23, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Undesired Nerve Stimulation

    MAUDE report 50874723 ↗
  • InjurySep 8, 2026Ambient Noise Problem

    This lead was capped due to noise. No adverse patient events were reported. Should additional information be received, this file will be updated.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50669906 ↗
  • MalfunctionSep 3, 2026Failure to Capture; Fracture; Ambient Noise Problem; Impedance Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50590297 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information be received, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50552160 ↗
  • InjurySep 1, 2026Adverse Event Without Identified Device or Use Problem

    System was explanted due to infection. Should additional information become available, this file will be updated.

    Patient problems: Unspecified Infection

    MAUDE report 50550243 ↗
  • InjuryAug 26, 2026Fracture

    This lead was explanted due to fracture from patient twiddling. Should additional information be received this file will be updated.

    Patient problems: Twiddlers Syndrome

    MAUDE report 50457257 ↗
  • InjuryMay 19, 2026Positioning Problem

    POSITION OF LEAD WAS NOT PROVIDING ADEQUATE CRT SYNCHRONIZATION. DUE TO LENGTH OF IMPLANT AND THE PATIENTS AGE, MD ELECTED TO CAP RATHER THAN ATTEMPT EXTRACTION. NO ADVERSE PATIENT EVENTS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25216363 ↗
  • InjuryMay 12, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25149426 ↗
  • InjuryMay 8, 2026Break

    INSULATION BREACH AT THE PIN DURING GENERATOR CHANGE. LV LEAD WAS TURNED OFF AT THIS TIME AND REMAINS IMPLANTED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25123733 ↗
  • InjuryApr 27, 2026Break

    THIS LEAD WAS CAPPED DUE TO ENTIRE PIN OF LEAD BROKE OFF DURING GEN CHANGE. INSULATION WAS PULLED BACK AND WIRES WERE EXPOSED. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25010216 ↗
  • InjuryApr 20, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24923394 ↗
  • InjuryApr 13, 2026Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO POCKET EROSION/INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Pocket Erosion; Unspecified Infection

    MAUDE report 24859158 ↗
  • InjuryMar 26, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24706949 ↗
  • InjuryMar 2, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24482778 ↗
  • InjuryFeb 24, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24429076 ↗
  • MalfunctionFeb 20, 2026High impedance; Ambient Noise Problem

    OUT OF RANGE LV PACING IMPEDANCE (GREATER THAN 3000 OHMS) SEEN ON HOME MONITORING. SENSING WAS PROGRAMMED OFF PREVIOUSLY DUE TO NOISE ON THE LV CHANNEL. LEAD TO BE EVALUATED FURTHER AND REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24400174 ↗
  • InjuryFeb 17, 2026Adverse Event Without Identified Device or Use Problem

    THIS SYSTEM WAS EXPLANTED DUE TO INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24365049 ↗
  • InjuryFeb 9, 2026Fracture; High impedance

    LEAD WAS SHOWING FLUCTUATING HIGH IMPEDANCE. THE POCKET WAS OPENED TO RULE OUT HEADER ISSUE AND THE LEAD WAS FOUND TO BE FRACTURED UNDER THE SUTURE SLEEVE. LEAD STILL FUNCTIONAL IN POLE 1 AND 3. WE MOVED SUTURE SLEEVE AND ADDED A LEAD REPAIR KIT OVER FRACTURED PORTION. LEAD REMAINS IMPLANTED AND ACTIVE. SHOULD ADDITIONAL INFORMATION BE RECEIVED THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24292432 ↗
  • InjuryFeb 9, 2026Device Dislodged or Dislocated; Migration

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT SECONDARY TO DEVICE POCKET MIGRATION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24291063 ↗
  • InjuryFeb 3, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24246045 ↗
  • InjuryJan 27, 2026Failure to Capture; High impedance

    THIS LEAD WAS CAPPED DUE TO LOSS OF CAPTURE AND PACING IMPEDANCE OUT OF RANGE. NO ADVERSE PATIENT EVENTS WERE REPORTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24190612 ↗
  • InjuryJan 23, 2026Device Dislodged or Dislocated

    THIS LEAD WAS EXPLANTED DUE TO DISLODGEMENT. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24161269 ↗
  • InjuryJan 20, 2026Adverse Event Without Identified Device or Use Problem

    SYSTEM WAS EXPLANTED DUE TO POCKET INFECTION. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

    Patient problems: Unspecified Infection

    MAUDE report 24125453 ↗

Other Drug eluting permanent left ventricular (lv) pacemaker electrode devices

All Drug eluting permanent left ventricular (lv) pacemaker electrode devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page