Attain Performa™ S MRI SureScan™ 459888Recalls & reported problems

Medtronic, Inc. · LEAD 459888 MRI S-TIP US · Class 3 · Implant

All models of Attain Performa S MRI SureScan (2)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic Attain Performa lead is an implanted wire that carries pacing signals from a pacemaker to the heart. Its 2,362 reports often describe the lead failing to pace the heart, needing too much energy to do so, and the lead moving out of place, and mention 33 deaths and 1,448 injuries. Some also mention infection and unwanted nerve or muscle stimulation. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
2,374
Deaths reported
33
Injuries reported
1,455
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2,374 adverse event reports that list this exact device (DI 00643169601871), received between Dec 9, 2016 and Sep 30, 2026 — classified as 33 deaths, 1,455 injuries, 886 malfunctions.

The most frequently reported specific device problem is “High Capture Threshold” (554 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 26 reportsDec 2023: 24 reportsJan 2024: 26 reportsFeb 2024: 14 reportsMar 2024: 23 reportsApr 2024: 39 reportsMay 2024: 28 reportsJun 2024: 19 reportsJul 2024: 28 reportsAug 2024: 20 reportsSep 2024: 16 reportsOct 2024: 33 reportsNov 2024: 21 reportsDec 2024: 14 reportsJan 2025: 25 reportsFeb 2025: 17 reportsMar 2025: 25 reportsApr 2025: 23 reportsMay 2025: 24 reportsJun 2025: 17 reportsJul 2025: 30 reportsAug 2025: 29 reportsSep 2025: 25 reportsOct 2025: 11 reportsNov 2025: 23 reportsDec 2025: 22 reportsJan 2026: 15 reportsFeb 2026: 30 reportsMar 2026: 30 reportsApr 2026: 29 reportsMay 2026: 26 reportsJun 2026: 23 reportsJul 2026: 11 reportsAug 2026: 17 reportsSep 2026: 18 reportsOct 2026: 0 reports20242025202639
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem889
  • High Capture Threshold554
  • Failure to Capture276
  • Device Dislodged or Dislocated271
  • Capturing Problem190
  • Activation, Positioning or Separation Problem179
  • High impedance130
  • Impedance Problem129

Patient problems

  • No Clinical Signs, Symptoms or Conditions890
  • Unspecified Infection490
  • No Known Impact Or Consequence To Patient315
  • Undesired Nerve Stimulation142
  • Device Overstimulation of Tissue135
  • Muscle Stimulation106
  • No Consequences Or Impact To Patient67
  • Erythema51

Recent adverse event reports

The 25 most recent of 2,374 reports, in the reporter's own words as published by the FDA.

  • InjurySep 30, 2026Device Dislodged or Dislocated

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50952547 ↗
  • InjurySep 30, 2026Intermittent Capture; High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951877 ↗
  • InjurySep 29, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to an infection resulting in discomfort, erosion, redness, swelling and wound dehiscence. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Erosion; Erythema; Swelling/ Edema; Unspecified Infection; Wound Dehiscence

    MAUDE report 50934517 ↗
  • InjurySep 28, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50923749 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Medical device site infection

    MAUDE report 50891966 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that three months post Cardiac Resynchronization Therapy Defibrillator (CRT-D) implant, the patient experienced a device pocket infection, discomfort, pain, purulent discharge, redness, and swelling. The CRT-D system was explanted and then replaced six days later. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Erythema; Medical device site infection; Pain; Purulent Discharge; Swelling/ Edema

    MAUDE report 50888930 ↗
  • InjurySep 24, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted due to a mobile echogenic mass/vegetation attached to the Right Atrial (RA) lead. It was noted that the patient also experienced discomfort. No further patient complications have been reported as a result of this event.

    Patient problems: Discomfort; Medical device site infection

    MAUDE report 50882162 ↗
  • InjurySep 18, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Unspecified Infection

    MAUDE report 50828438 ↗
  • InjurySep 15, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Bacterial Infection

    MAUDE report 50781043 ↗
  • MalfunctionSep 15, 2026Activation, Positioning or Separation Problem; Device Dislodged or Dislocated

    It was reported that during an implant procedure a guidewire was unable to smoothly track through the Left Ventricular (LV) lead. When the wire was pulled back the lead came with it. The LV lead was removed and replaced with a new lead. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50777543 ↗
  • InjurySep 14, 2026Extracardiac stimulation

    It was reported that three transmissions were missing current electrograms (EGM). The Cardiac Resynchronization Therapy Pacemaker (CRT-P) remains in use. It was further noted that the patient experienced diaphragmatic stimulation and the Left Ventricular (LV) lead was reprogrammed. No further patient complications have been reported as a result of this event.

    Patient problems: Device Overstimulation of Tissue

    MAUDE report 50754068 ↗
  • MalfunctionSep 10, 2026Capturing Problem; Impedance Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50722193 ↗
  • MalfunctionSep 8, 2026Appropriate Device Problem Term/Code Not Available

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50679164 ↗
  • MalfunctionSep 8, 2026Low impedance; High Capture Threshold

    It was reported that the Left Ventricular (LV) lead exhibited high capture thresholds and low pacing impedance. The lead remains in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50674959 ↗
  • InjurySep 4, 2026Adverse Event Without Identified Device or Use Problem

    Patient problems: Abscess; Fever; Unspecified Infection

    MAUDE report 50618344 ↗
  • MalfunctionSep 4, 2026Capturing Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50617216 ↗
  • MalfunctionSep 2, 2026Capturing Problem

    Patient problems: Inappropriate Device Stimulation of Tissue

    MAUDE report 50571587 ↗
  • MalfunctionSep 2, 2026Extracardiac stimulation; High Capture Threshold

    Patient problems: Dyspnea; Inappropriate Device Stimulation of Tissue

    MAUDE report 50570315 ↗
  • MalfunctionAug 26, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50460482 ↗
  • MalfunctionAug 24, 2026Failure to Capture

    It was reported that there was a position check failure on the Right Atrial (RA) lead. It was also reported that the Left Ventricular (LV) lead exhibited loss of capture. The RA lead and LV lead remain in use. No patient complications have been reported as a result of this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50414028 ↗
  • MalfunctionAug 21, 2026High Capture Threshold

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50375227 ↗
  • InjuryAug 20, 2026Adverse Event Without Identified Device or Use Problem

    It was reported that the patient developed a bacteremia infection. The Cardiac Resynchronization Therapy Pacemaker (CRT-P) system was explanted and replaced. No further patient complications have been reported as a result of this event.

    Patient problems: Bacteremia

    MAUDE report 50355664 ↗
  • MalfunctionAug 19, 2026Activation, Positioning or Separation Problem; Material Integrity Problem

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50326096 ↗
  • InjuryAug 17, 2026Extracardiac stimulation

    It was reported that approximately four weeks after implant, the patient experienced phrenic nerve stimulation. The Left Ventricular (LV) lead was explanted and replaced with another model in an alternate location. No further patient complications have been reported as a result of this event.

    Patient problems: Device Overstimulation of Tissue

    MAUDE report 50288921 ↗
  • MalfunctionAug 13, 2026High Capture Threshold

    Patient problems: Fall

    MAUDE report 50237278 ↗

Other permanent pacemaker Electrode devices

All permanent pacemaker Electrode devices →