TiMesh 015-042Recalls & reported problems

Medtronic Ps Medical, Inc. · PLATE 015-042 STRAIGHT 1H X 1H 8MM · Class 2 · Implant

All models of TiMesh (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Medtronic TiMesh plate is a small titanium plate used to repair the skull. The 26 reports mostly describe the plate breaking, cracking, or arriving bent — during surgery or the cleaning process, often replaced with a backup product. There are no deaths and no injuries. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
26
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00643169620278), received between Sep 13, 2018 and Feb 12, 2026 — classified as 26 malfunctions.

The most frequently reported specific device problem is “Break” (26 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break26
  • Fracture4

Patient problems

  • No Clinical Signs, Symptoms or Conditions22
  • No Known Impact Or Consequence To Patient18

Recent adverse event reports

The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 12, 2026Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A TITANIUM PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT THE CONNECTING PLATE BROKE DURING A CRANIOTOMY. THE PROCEDURE WAS COMPLETED WITH BACK UP PRODUCTS. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24322724 ↗
  • MalfunctionSep 16, 2025Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE USED FOR CRANIAL REPAIR. IT WAS REPORTED THAT THERE WAS BREAKAGE/FRACTURE OF THE PLATE OCCURRED DURING THE DISINFECTION PROCESS. THE PROCEDURE WAS COMPLETED WITH BACK UP PRODUCTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23063843 ↗
  • MalfunctionDec 27, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT DURING INSPECTION UPON THE ARRIVAL OF THE PURCHASED GOOD, IT WAS FOUND THAT THE PRODUCT WAS KINKED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21025691 ↗
  • MalfunctionSep 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT DURING THE DISINFECTION PROCESS, THE HOLE EXPANDED. NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20124948 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13386396 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13386302 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13386260 ↗
  • MalfunctionJan 27, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13373860 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10781617 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10781556 ↗
  • MalfunctionNov 3, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT A DISINFECTING AGENT WAS NOT USED AS PART OF THE DISINFECTION PROCESS. THE DISINFECTING PROCESS THAT WAS FOLLOWED WAS CLEAN WATER SHOWER PLUS HIGH TEMPERATURE AND HIGH PRESSURE DISINFECTION. THE PROCESS OF PUTTING LOOSE SCREWS INTO THE BOX WAS OPERATED BY THE WORKERS RESPONSIBLE FOR THE DISINFECTION IN THE HOSPITAL. IT WAS NOTED THAT THE TEMPERATURE WAS 134 DEGREES CELSIUS, AND THE PRESSURE WAS 3BAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10781533 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10781515 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10781089 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10781003 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780819 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780796 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780498 ↗
  • MalfunctionNov 3, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT A DISINFECTING AGENT WAS NOT USED AS PART OF THE DISINFECTION PROCESS. THE DISINFECTING PROCESS THAT WAS FOLLOWED WAS CLEAN WATER SHOWER PLUS HIGH TEMPERATURE AND HIGH PRESSURE DISINFECTION. THE PROCESS OF PUTTING LOOSE SCREWS INTO THE BOX WAS OPERATED BY THE WORKERS RESPONSIBLE FOR THE DISINFECTION IN THE HOSPITAL. IT WAS NOTED THAT THE TEMPERATURE WAS 134 DEGREES CELSIUS, AND THE PRESSURE WAS 3BAR.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780490 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780356 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780279 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780208 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10779065 ↗
  • MalfunctionSep 9, 2020Break; Break

    ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WERE NO ENVIRONMENTAL/EXTERNAL/PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10508807 ↗
  • MalfunctionMar 4, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE PLATE WAS BROKEN AND COULDN'T BE USED NORMALLY. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9788045 ↗
  • MalfunctionMar 4, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE PLATE WAS BROKEN AND COULDN'T BE USED NORMALLY. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9788030 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →