TiMesh 015-042Recalls & reported problems
Medtronic Ps Medical, Inc. · PLATE 015-042 STRAIGHT 1H X 1H 8MM · Class 2 · Implant
All models of TiMesh (15)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Medtronic TiMesh plate is a small titanium plate used to repair the skull. The 26 reports mostly describe the plate breaking, cracking, or arriving bent — during surgery or the cleaning process, often replaced with a backup product. There are no deaths and no injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 26
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 26 adverse event reports that list this exact device (DI 00643169620278), received between Sep 13, 2018 and Feb 12, 2026 — classified as 26 malfunctions.
The most frequently reported specific device problem is “Break” (26 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break26
- Fracture4
Patient problems
- No Clinical Signs, Symptoms or Conditions22
- No Known Impact Or Consequence To Patient18
Recent adverse event reports
The 25 most recent of 26 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 12, 2026Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A TITANIUM PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT THE CONNECTING PLATE BROKE DURING A CRANIOTOMY. THE PROCEDURE WAS COMPLETED WITH BACK UP PRODUCTS. THE PATIENT'S STATUS WAS ALIVE-NO INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24322724 ↗MalfunctionSep 16, 2025Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE USED FOR CRANIAL REPAIR. IT WAS REPORTED THAT THERE WAS BREAKAGE/FRACTURE OF THE PLATE OCCURRED DURING THE DISINFECTION PROCESS. THE PROCEDURE WAS COMPLETED WITH BACK UP PRODUCTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23063843 ↗MalfunctionDec 27, 2024Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT DURING INSPECTION UPON THE ARRIVAL OF THE PURCHASED GOOD, IT WAS FOUND THAT THE PRODUCT WAS KINKED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21025691 ↗MalfunctionSep 3, 2024Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A PLATE FOR CRANIAL REPAIR. IT WAS REPORTED THAT DURING THE DISINFECTION PROCESS, THE HOLE EXPANDED. NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20124948 ↗MalfunctionJan 28, 2022Break; Break
IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13386396 ↗MalfunctionJan 28, 2022Break; Break
IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13386302 ↗MalfunctionJan 28, 2022Break; Break
IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13386260 ↗MalfunctionJan 27, 2022Break; Break
IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13373860 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10781617 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10781556 ↗MalfunctionNov 3, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED REPORTED THAT A DISINFECTING AGENT WAS NOT USED AS PART OF THE DISINFECTION PROCESS. THE DISINFECTING PROCESS THAT WAS FOLLOWED WAS CLEAN WATER SHOWER PLUS HIGH TEMPERATURE AND HIGH PRESSURE DISINFECTION. THE PROCESS OF PUTTING LOOSE SCREWS INTO THE BOX WAS OPERATED BY THE WORKERS RESPONSIBLE FOR THE DISINFECTION IN THE HOSPITAL. IT WAS NOTED THAT THE TEMPERATURE WAS 134 DEGREES CELSIUS, AND THE PRESSURE WAS 3BAR.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10781533 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10781515 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10781089 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10781003 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780819 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780796 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780498 ↗MalfunctionNov 3, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED REPORTED THAT A DISINFECTING AGENT WAS NOT USED AS PART OF THE DISINFECTION PROCESS. THE DISINFECTING PROCESS THAT WAS FOLLOWED WAS CLEAN WATER SHOWER PLUS HIGH TEMPERATURE AND HIGH PRESSURE DISINFECTION. THE PROCESS OF PUTTING LOOSE SCREWS INTO THE BOX WAS OPERATED BY THE WORKERS RESPONSIBLE FOR THE DISINFECTION IN THE HOSPITAL. IT WAS NOTED THAT THE TEMPERATURE WAS 134 DEGREES CELSIUS, AND THE PRESSURE WAS 3BAR.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780490 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780356 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780279 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10780208 ↗MalfunctionNov 3, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10779065 ↗MalfunctionSep 9, 2020Break; Break
ADDITIONAL INFORMATION RECEIVED REPORTED THAT THERE WERE NO ENVIRONMENTAL/EXTERNAL/PATIENT FACTORS THAT MAY HAVE LED OR CONTRIBUTED TO THE ISSUE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10508807 ↗MalfunctionMar 4, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE PLATE WAS BROKEN AND COULDN'T BE USED NORMALLY. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9788045 ↗MalfunctionMar 4, 2020Break; Break
IT WAS REPORTED THAT DURING THE SURGERY, THE PLATE WAS BROKEN AND COULDN'T BE USED NORMALLY. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCT(S). THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9788030 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-MLT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-MRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page