Skip to content

TiMesh 015-246Recalls & reported problems

Medtronic Ps Medical, Inc. · PLATE 015-246 STRAIGHT 2H X 2H 2.5MM · Class 2 · Implant

All models of TiMesh (15)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
9
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00643169620452), received between Nov 3, 2020 and Jan 3, 2024 — classified as 9 malfunctions.

The most frequently reported specific device problem is “Break” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 4 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520264
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break9
  • Fracture4

Patient problems

  • No Clinical Signs, Symptoms or Conditions9
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18438688 ↗
  • MalfunctionJan 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18438646 ↗
  • MalfunctionJan 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18438595 ↗
  • MalfunctionJan 3, 2024Break; Fracture; Break; Fracture

    MEDTRONIC RECEIVED INFORMATION REGARDING CRANIAL REPAIR PRODUCTS. IT WAS REPORTED THAT WHEN USING THE HAND TOOLS TO TIGHTEN THE SCREW DURING SURGERY, DUE TO THE LACK OF SELF-TAPPING PROPERTIES OF THE SCREW, THE DOCTOR NEEDED TO USE MORE FORCE TO TIGHTEN THE SCREW CAUSING THE SNOWFLAKE PIECE, CONNECTING PIECE, AND SCREW TO DEFORM AND MAKING IT IMPOSSIBLE TO CONTINUE USING IT. IT WAS NOTED THAT THE HOSPITAL SUPPLY ROOM COLLECTED DAMAGED PRODUCTS FROM VARIOUS MANUFACTURERS ON A DAILY BASIS AND PROVIDED THEM TO THE DISTRIBUTOR AT ONE TIME DURING THE INVENTORY PERIOD AT THE END OF THE YEAR. IT WAS ASKED FOR THE DOCTORS TO USE MEDTRONIC PRODUCTS OF THE SAME MODEL INSTEAD.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18438469 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13387511 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13387394 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13386432 ↗
  • MalfunctionJan 28, 2022Break; Break

    IT WAS REPORTED THAT DURING THE OPERATION, THE SCREW SLIPPED AND BROKE AS WELL AS THE CONNECTING PIECE AND CAP HOLE BROKE ON THE PLATE. THE PRODUCTS WERE REPLACED WITH THE SAME PRODUCT. THIS EVENT RESULTED IN A PROCEDURE DELAY OF 1 MINUTE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13386203 ↗
  • MalfunctionNov 3, 2020Break; Break

    IT WAS REPORTED THAT DURING THE SURGERY, THE SCREW SLIPPED, THE SCREW BROKE, AND THE CONNECTING PLATE AND TITANIUM CAP HOLE BROKE. THIS RESULTED IN A PROCEDURE DELAY OF 2 MINUTES. THE PROCEDURE WAS COMPLETED WITH BACKUP PRODUCTS. THE PATIENT'S STATUS AT THE TIME OF THE REPORT WAS ALIVE-NO INJURY. ADDITIONAL INFORMATION RECEIVED REPORTED THAT THIS PROBLEM HAD BEEN OCCURRING SINCE (B)(6) 2018; HOWEVER, THIS WAS THE FIRST TIME THE HOSPITAL HAD NOTIFIED THE DISTRIBUTOR OF THE QUALITY PROBLEMS. IT WAS NOTED THAT THE TITANIUM PRODUCTS NEED TO BE DISINFECTED IN THE HOSPITAL SUPPLY ROOM AFTER UNPACKING, AND THEN SENT TO THE OPERATING ROOM. SIMILAR PRODUCTS ON THE MARKET MAY CAUSE THE PRODUCT TO BREAK DUE TO MANUAL OPERATION AND OTHER REASONS DURING THE DISINFECTION PROCESS. THIS COMPENSATION MAINLY BELONGED TO THE OBVIOUS BROKEN PRODUCTS PICKED OUT BY THE STAFF IN THE SUPPLY ROOM DURING THE DISINFECTION OR AFTER THE DOCTOR OPENED THE DISINFECTION PACKAGE IN THE OPERATING ROOM.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient

    MAUDE report 10780011 ↗

Other Plate, fixation, bone devices

All Plate, fixation, bone devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page