TiMesh 016-152Recalls & reported problems
Medtronic Ps Medical, Inc. · PANEL 016-152 CRANIAL MESH 152 X 152MM · Class 2 · Implant
All models of TiMesh (15)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169620643), received between Feb 10, 2023 and Apr 28, 2023 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- Post Operative Wound Infection1
- Unspecified Tissue Injury1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION REGARDING A TIMESH DEVICE. IT WAS REPORTED THAT THE PATIENT EXPERIENCED TITANIUM MESH REJECTION AFTER SURGERY, SKIN LESIONS, AND TITANIUM MESH EXPOSURE. THE PATIENT WAS ADMITTED TO THE HOSPITAL DUE TO "HEAD TRAUMA AND A SKULL DEFECT FOR MORE THAN 4 MONTHS AFTER CRANIOTOMY." THE PATIENT UNDERWENT RIGHT FRONTOTEMPORAL AND PARIETAL SKULL DEFECT REPAIR ON (B)(6) 2023. AFTER THE OPERATION, THE PATIENT EXPERIENCED TITANIUM MESH REJECTION, AND UNDERWENT RIGHT FRONTOTEMPORAL AND PARIETAL SKULL DEFECT REPAIR. AFTER THE OPERATION, THE PATIENT EXPERIENCED TITANIUM MESH REJECTION. 3-2, SKIN LESIONS WITH A DIAMETER OF ABOUT 1.5 CM WERE SEEN ON THE RIGHT FOREHEAD OF THE PATIENT, AND A LITTLE TITANIUM MESH WAS PARTIALLY EXPOSED. 3-8, THE PATIENT WAS SENT TO THE OPERATING ROOM, AND THE TITANIUM MESH WAS REMOVED UNDER GENERAL ANESTHESIA WITH ENDOTRACHEAL INTUBATION.
Patient problems: Post Operative Wound Infection; Unspecified Tissue Injury
MAUDE report 16829574 ↗MalfunctionFeb 10, 2023Break; Fracture; Break; Fracture
MEDTRONIC RECEIVED INFORMATION REGARDING A CRANIAL MESH DEVICE BEING USED IN A CRANIOPLASTY. IT WAS REPORTED THAT THE TAIL OF THE SCREW WAS PROTRUDING AND UNEVEN, SO THE WEAK SKIN WAS EASY TO BE COMPRESSED AND NECROTIC; THE SCREW WAS NOT STRONG ENOUGH, AND ITS TEXTURE WAS BRITTLE, SO IT WAS EASY TO BREAK AFTER BEING STRESSED, AND THE TITANIUM PLATE WAS NOT FIXED FIRMLY. THE PLATE HOLE AT THE EDGE OF THE TITANIUM PLATE WAS OVAL, AND THE UPPER SCREW WAS UNEVEN; THE SCREW STRENGTH WAS POOR, AND THE CONTACT PART BETWEEN THE SCREW'S TAIL AND THE SCREWDRIVER WAS EASY TO CRACK WHEN THE FORCE WAS APPLIED. THE DOCTOR TRIED TO MAKE IT AS SMOOTH AS POSSIBLE DURING OPERATION, AND OBSERVED AFTER THE SURGERY. THERE WAS NO REPORTED IMPACT TO PATIENT OUTCOME.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16353572 ↗
Other Plate, fixation, bone devices
- Lapiplasty System 2 SK14Treace Medical Concepts, Inc. · Plate, Fixation, Bone260 reports
- PRECICE® Plating System LP4.5-PRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls6 reports
- PRECICE® Plating System LP4.5-MLT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls5 reports
- PRECICE® Plating System LP4.5-PLF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-MRT4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls3 reports
- PRECICE® Plating System LP4.5-DRF4BNuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls1 report
- PRECICE® Plating System LP4.5-PLF40Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
- PRECICE® Plating System LP4.5-PRT42Nuvasive Specialized Orthopedics, Inc. · Plate, Fixation, Bone3 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page