MyCareLink™ 24950J (UDI 00643169657885)Recalls & reported problems
Medtronic, Inc. · MON 24950J MYCARELINK/SVC USA · Class 3
All models of MyCareLink (33)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 12
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 00643169657885), received between Dec 9, 2016 and Feb 20, 2024 — classified as 12 malfunctions.
The most frequently reported specific device problem is “Overheating of Device” (10 reports).
1 FDA recall record name this device's identifier.
9 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- Jul 14, 2017Terminated
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
Z-0399-2018 · Root cause: Error in labeling · Recall details · FDA record ↗
Most reported problems
Device problems
- Overheating of Device10
- Inappropriate/Inadequate Shock/Stimulation1
- Mechanical Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions8
- No Consequences Or Impact To Patient2
- No Known Impact Or Consequence To Patient2
- Burn(s)1
- Shock1
Recent adverse event reports
The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 20, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT THE REMOTE MONITOR READER EXPLODED. NO TROUBLESHOOTING TAKEN. SUGGESTED REPLACEMENT BUT THE PATIENT WOULD LIKE TO WAIT AFTER SPEAKING WITH THE PHYSICIAN REGARDING SENDING IN A TRANSMISSION SOONER THAN REPLACEMENT WOULD ARRIVE. THE MONITOR REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18737990 ↗MalfunctionMar 24, 2023Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE READER OF THE REMOTE MONITOR WAS TOO HOT TO TOUCH. THE PATIENT WAS ADVISED TO LEAVE THE MONITOR DISCONNECTED. NO FURTHER TROUBLESHOOTING AT THE TIME OF CALL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16612718 ↗MalfunctionSep 1, 2022Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING TAKEN. THE CALLER WAS TOLD TO LEAVE THE MONITOR UNPLUGGED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15337473 ↗MalfunctionDec 9, 2021Overheating of Device; Overheating of Device
THE MONITOR WAS RETURNED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12970082 ↗MalfunctionJun 9, 2021Overheating of Device; Overheating of Device
THE MONITOR WAS REPLACED AND RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11970729 ↗MalfunctionMar 8, 2021Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR CAUGHT FIRE AND THE BASE WAS DESTROYED BUT THE READER WAS INTACT. NO TROUBLESHOOTING WAS TAKEN. PATIENT DISCARDED THE DAMAGED BASE BUT STILL HAD THE RECEIVER. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11440933 ↗MalfunctionFeb 19, 2021Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR DISPLAYED A 3230 DIAGNOSTIC CODE INDICATING THE READER IS OFF THE MONITOR BASE AND HAS POWERED DOWN, IS OFF THE BASE STATION AND HAS NO BATTERY LIFE REMAINING OR THE READER IS OUT OF RANGE OF THE MONITOR. ALSO, IT BECAME SO HOT THAT IT BURNED THE PATIENT. NO TROUBLESHOOTING TAKEN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Burn(s); No Clinical Signs, Symptoms or Conditions
MAUDE report 11347443 ↗MalfunctionOct 27, 2020Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE REMOTE MONITOR WAS OVERHEATING. NO TROUBLESHOOTING INDICATED. THE REMOTE MONITOR IS EXPECTED TO BE REPLACED . NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10740929 ↗MalfunctionSep 19, 2020Overheating of Device; Overheating of Device
THE MONITOR WAS REPLACED AND RETURNED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10557642 ↗MalfunctionOct 30, 2019Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE READER WAS OVERHEATING. THE PATIENT UNPLUGGED AND PLUGGED BACK IN THE REMOTE MONITOR. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9255948 ↗MalfunctionOct 6, 2017Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE READER PORTION OF THE REMOTE MONITOR WAS HEATING UP SO MUCH THAT THE PATIENT COULD NOT HOLD IT TO THE SKIN. A NEW MONITOR WAS ORDERED FOR THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6919915 ↗MalfunctionDec 9, 2016Inappropriate/Inadequate Shock/Stimulation; Inappropriate/Inadequate Shock/Stimulation
IT WAS REPORTED THAT THE PATIENT WAS SHOCKED BY THE REMOTE MONITOR DURING THE INTERROGATION PHASE OF A REMOTE TRANSMISSION. THE PATIENT WAS ASSURED THAT THE MONITOR DOES NOT PERFORM THERAPY TO IMPLANTED DEVICE BUT COLLECTS AND SENDS TRANSMISSIONS. THE PATIENT WAS REFERRED TO THE CLINIC TO HAVE IMPLANTED DEVICE AND MONITOR CHECKED. A FOLLOW UP CALL WAS MADE TO THE CLINIC. THE NURSE STATED THAT THE PATIENT HAD BEEN SEEN IN CLINIC TWICE SINCE THE CALL FOR A DEVICE CHECK AND FOLLOW UP AND EVERYTHING WAS WITHIN NORMAL LIMITS. BOTH THE MONITOR AND THE IMPLANTED DEVICE REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock
MAUDE report 6160739 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.
- Z-2168-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
- Z-2169-2025 · Jun 18, 2025 · Open, ClassifiedMedtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-2505-2020 · Jun 4, 2020 · Open, ClassifiedCareLink SmartSync Device Manager, Model Number 24970A
- Z-2506-2020 · Jun 4, 2020 · Open, ClassifiedPatient Connector, Model Number 24967
- Z-0845-2019 · Jan 17, 2019 · TerminatedMedtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
- Z-0400-2018 · Jul 14, 2017 · TerminatedMyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
- Z-2209-2010 · Jun 11, 2010 · TerminatedMedtronic Carelink Monitor Model 2490C. The Medtronic CareLink Monitor Model 2490C is designed to automatically gather information from the implanted heart device. This automatic w
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page