Skip to content

MyCareLink™ 24950JLQ (UDI 00643169658127)Recalls & reported problems

Medtronic, Inc. · MON 24950JLQ MYCARELINK/SVC USA LINQ · Class 3

All models of MyCareLink (33)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00643169658127), received between Jun 8, 2017 and May 29, 2019 — classified as 10 malfunctions.

The most frequently reported specific device problem is “Overheating of Device” (9 reports).

1 FDA recall record name this device's identifier.

9 other recalls cover devices cleared under the same FDA submission (P890003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jul 14, 2017Class IITerminated

    Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.

    Z-0399-2018 · Root cause: Error in labeling · Recall details · FDA record ↗

Most reported problems

Device problems

  • Overheating of Device9
  • Device Emits Odor4
  • Smoking1

Patient problems

  • No Known Impact Or Consequence To Patient10

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 29, 2019Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR STARTED SMOKING. THE PATIENT SAW SMOKE COMING FROM THE SIDES OF THE REMOTE MONITOR AND THEN FROM THE SCREEN. THE PATIENT UNPLUGGED THE MONITOR AND STATED THAT THE POWER CORD HAD BURN MARKS ON AREA CONNECTING TO THE MONITOR. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8649942 ↗
  • MalfunctionMar 5, 2019Overheating of Device; Overheating of Device

    THE MONITOR WAS REPLACED AND RETURNED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8392287 ↗
  • MalfunctionJan 17, 2019Overheating of Device; Overheating of Device

    THE MONITOR WAS REPLACED AND RETURNED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8258610 ↗
  • MalfunctionAug 23, 2018Overheating of Device; Overheating of Device

    IT WAS ALSO REPORTED THAT THE REMOTE MONITOR WAS TOO HOT TO TOUCH. NO TROUBLESHOOTING INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7808018 ↗
  • MalfunctionJul 24, 2018Device Emits Odor; Overheating of Device; Smoking; Device Emits Odor

    IT WAS REPORTED THAT THE REMOTE MONITOR SMELLED LIKE IT WAS SMOKING AND READER HEAD WAS HOT TO THE TOUCH, NO CHAR OR DISCOLORATION. NO TROUBLESHOOTING INDICATED. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7713568 ↗
  • MalfunctionApr 20, 2018Device Emits Odor; Overheating of Device; Device Emits Odor; Overheating of Device

    IT WAS REPORTED THAT THE REMOTE MONITOR PLUG CAUGHT ON FIRE WHILE PLUGGED INTO A RECEPTACLE AT THE PATIENT'S HOME. THE MONITOR IS EXPECTED TO BE REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7446626 ↗
  • MalfunctionApr 16, 2018Device Emits Odor; Overheating of Device; Device Emits Odor; Overheating of Device

    IT WAS REPORTED THAT THE PATIENT CLAIMED THERE HAD BEEN A HOUSE FIRE IN (B)(6) AND WAS TOLD BY THE FIRE MARSHAL THAT THE FIRE WAS INITIATED BY THE REMOTE MONITOR. THE STATEMENT WAS MADE BY THE PATIENT AND NOT BY THE STATE FIRE MARSHAL. THE PATIENT ALREADY DISCARDED THE MONITOR. THE MONITOR REPLACEMENT STATUS WAS UNKNOWN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7432868 ↗
  • MalfunctionNov 16, 2017Device Emits Odor; Device Emits Odor

    IT WAS REPORTED THAT THE "HOME MONITOR BURNED ON A FIRE". THE MONITOR STATUS IS UNKNOWN. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7039134 ↗
  • MalfunctionOct 6, 2017Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE PATIENT LEFT THE MONITOR UNPLUGGED FOR A MONTH AND IT IS TOO HOT TO TOUCH. THE MONITOR IS BEING RETURNED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6924659 ↗
  • MalfunctionJun 8, 2017Overheating of Device; Overheating of Device

    IT WAS REPORTED THAT THE READER ON THE REMOTE MONITOR WAS OVERHEATING TO ONE HUNDRED AND THRITY FOUR DEGREES. IT WAS NOTED THAT THE TEMPERATURE WAS MEASURED BY THE PATIENT USING A HEAT GUN. THE PATIENT WAS ADVISED TO UNPLUG THE MONITOR AND NOT USE IT. A NEW MONITOR WAS ORDERED FOR THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 6624967 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P890003. One submission often covers many models, so check the product and lot details in each record.

  • Z-2169-2025 · Jun 18, 2025 · Open, Classified
    Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria A
  • Z-2168-2025 · Jun 18, 2025 · Open, Classified
    Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt Crome application, Software Model Number D00U005; 2. Claria
  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-2506-2020 · Jun 4, 2020 · Open, Classified
    Patient Connector, Model Number 24967
  • Z-2505-2020 · Jun 4, 2020 · Open, Classified
    CareLink SmartSync Device Manager, Model Number 24970A
  • Z-0845-2019 · Jan 17, 2019 · Terminated
    Medtronic Power Supply 26907, Medtronic CareLink Encore(TM) 29901 Programmer
  • Z-0400-2018 · Jul 14, 2017 · Terminated
    MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates
  • Z-2209-2010 · Jun 11, 2010 · Terminated