Skip to content

NA CB8S58R2Recalls & reported problems

Medtronic, Inc. · CUSTOM PACK CB8S58R2 3/8 SUPP · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00643169680586), received between Jun 20, 2023 and Jun 20, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).

10 other recalls cover devices cleared under the same FDA submission (K891687); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Fluid/Blood Leak1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 20, 2023Fluid/Blood Leak; Fluid/Blood Leak

    MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A CUSTOM TUBING PACK, A CLEAR PRIMING SOLUTION LEAK AT THE 3/8X3/8 DOUBLE LUER CONNECTOR WAS OBSERVED AT THE TIME OF SET-UP AND PRIMING. THE EXACT LOCATION OF THE LEAK IN THE CUSTOM TUBING PACK SPECIFICATION WAS LINE 4, PAGE 4 OF 8. WHEN THE LEAK WAS IDENTIFIED, THE EXTRA CORPOREAL MEMBRANE OXYGENATION (ECMO) SPECIALIST ISOLATED THE CONNECTOR AND THEN REMOVED IT FROM THE CIRCUIT. THEY REPLACED THE LEAKING CONNECTOR WITH AN OFF THE SHELF 3/8X3/8 DOUBLE LUER CONNECTOR AND PROCEEDED WITH THE SETUP AND PRIMING OF THE CIRCUIT. THERE WAS NO PATIENT INVOLVEMENT SO NO ADVERSE EFFECT OCCURRED. ADDITIONAL INFORMATION: THERE WAS NO VISIBLE AIR IN THE SYSTEM/TUBING. THE SAME LEAK DID NOT APPEAR IN MULTIPLE PACKS. IT WAS NOT INDICATED ON WHICH SIDE OF THE CONNECTOR THE LEAK WAS FOUND.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17163236 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K891687. One submission often covers many models, so check the product and lot details in each record.

  • Z-2495-2020 · Aug 8, 2018 · Terminated
    Extra corporeal circuit with bio-active surface.
  • Z-1023-2016 · Feb 4, 2016 · Terminated
    Medtronic Custom Perfusion Tubing Pack, Rx, Sterile EO. Terumo Cardiovascular Systems (Terumo CVS) has distributed an Urgent Safety Alert regarding the CD I H/S Cuvette. The cuvet
  • Z-2736-2015 · Aug 18, 2015 · Terminated
    MEDTRONIC PERFUSION TUBING PACKS (with potentially affected Hemoconcentrators). Model # 7U65R2, 8C72R4, 6N21R2, BB7M38R11, TL2Q68R17, HY8R66R3, HY2U43R9, 8C72R4, BB5J59R19, 6B65R3
  • Z-2105-2013 · Jul 17, 2013 · Terminated
    Medtronic Custom Perfusion System with Carmeda BioActive Surface. Model Numbers: CB1D82R12 and CB5N73R7. Sterilized by Ethylene Oxide, Do no Reuse. Product Usage: Usage: Th
  • Z-0618-2013 · Nov 19, 2012 · Terminated
    Medtronic Intersept Custom Tubing Pack with or without coating ( Carmeda BioActive Surfact, Carmeda BioActive Surfact and Trillium Biosurface,or Balance Biosurface) with the follow
  • Z-0813-2012 · Dec 14, 2011 · Terminated
    Custom Packs/Mexico, Medtronic Cardio-Kit, Sterilized Using Ethylene Oxide, Nonpyrogenic, Rx Only. Model # DLP6N48R3 & DLP6N48R4. Medtronic Intersept Custom Tubing Pack with T
  • Z-1920-2008 · May 9, 2008 · Terminated
    Medtronic Extracorporeal Circuit with bio-active surface; Cardiopulmonary Bypass Oxygenator, Catalog Numbers: CB0B90R3, CB0L44R15, CB0X97R10, CB1V99R2, CB1Y06R14, CB1Z42R15, CB
  • Z-1227-2007 · Aug 1, 2007 · Terminated
    Medtronic Intersept 3608 Sterile tubing (8') 1/2 x 3/32 in. Quantity 10, Lot 7626621, Use by 2009-07-31. Manufacturer: Medtronic, Inc. Minneapolis, MN 55432-5604 USA. Diffe
  • Z-0583-2007 · Jan 2, 2007 · Terminated
    Medtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and
  • Z-1489-04 · Sep 15, 2004 · Terminated
    Medtronic DLP VAD Venous Cannula for Ventricular Assist, Models 95036 and CB95036.

Other OXYGENATOR, CARDIOPULMONARY BYPASS devices

All OXYGENATOR, CARDIOPULMONARY BYPASS devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.