Skip to content

FlexCath™ Select 990065Recalls & reported problems

Medtronic, Inc. · SHEATH 990065 FLEXCATH SELECT 10 FR 65CM · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00643169688858), received between Aug 22, 2019 and Aug 26, 2019 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Material Deformation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Material Deformation2
  • Material Twisted/Bent1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 26, 2019Material Deformation; Material Deformation

    IT WAS REPORTED THAT DURING A RADIOFREQUENCY (RF) PROCEDURE, THE SHEATH WAS HAVING DIFFICULTY MANEUVERING. IT WAS THEN OBSERVED THE SHEATH WAS KINKED AND WAS REPLACED. THE SECOND SHEATH ALSO HAD MANEUVERABILITY ISSUES AND WAS ABLE TO BE MANIPULATED WITH RESOLVE IN ORDER TO PERFORM RF. THE CASE WAS COMPLETED WITH RADIOFREQUENCY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. THE SECOND SHEATH SUBSEQUENTLY TESTED OUT OF SPECIFICATION PER THE MANUFACTURER'S INVESTIGATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8933550 ↗
  • MalfunctionAug 22, 2019Material Deformation; Material Twisted/Bent; Material Deformation; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A RADIOFREQUENCY (RF) PROCEDURE, THE SHEATH WAS HAVING DIFFICULTY MANEUVERING. IT WAS THEN OBSERVED THE SHEATH WAS KINKED AND WAS REPLACED. THE SECOND SHEATH ALSO HAD MANEUVERABILITY ISSUES AND WAS ABLE TO BE MANIPULATED WITH RESOLVE IN ORDER TO PERFORM RF. THE CASE WAS COMPLETED WITH RADIOFREQUENCY. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8922486 ↗

Other INTRODUCER, CATHETER devices

All INTRODUCER, CATHETER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page