Skip to content

Adapta™ ADDR03 (UDI 00643169708648)Recalls & reported problems

Medtronic, Inc. · IPG ADDR03 ADAPTA DR 3.2MM US NS · Class 3 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
6
Deaths reported
1
Injuries reported
4
Recalls naming it
1

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00643169708648), received between Aug 30, 2018 and Nov 9, 2023 — classified as 1 death, 4 injuries, 1 malfunction.

The most frequently reported specific device problem is “Failure to Capture” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

15 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 17, 2019Open, Classified

    A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

    Z-0806-2019 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available3
  • Adverse Event Without Identified Device or Use Problem2
  • Failure to Capture1
  • Pacing Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Hematoma1
  • Insufficient Information1
  • No Known Impact Or Consequence To Patient1
  • Unspecified Infection1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryNov 9, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPLACED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, GIVEN THE POTENTIAL FOR LOSS OF DEVICE FUNCTIONALITY, THE IPG WAS EXPLANTED AND REPLACED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18099448 ↗
  • InjuryOct 21, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT APPROXIMATELY FIVE DAYS POST OPERATIVELY THE PATIENT EXPERIENCED A HEMATOMA OVER INCISION SITE. SURGERY WAS PERFORMED TO REMOVE THE CLOT AND THE POCKET WAS IRRIGATED. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Hematoma

    MAUDE report 12670588 ↗
  • DeathSep 30, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information

    MAUDE report 12553414 ↗
  • MalfunctionJun 30, 2021Failure to Capture; Pacing Problem; Failure to Capture; Pacing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS FAILING TO PACE AT APPROPRIATE TIMES AND FAILING TO CAPTURE PER TELEMETRY. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12090104 ↗
  • InjuryMar 12, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPLACED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, GIVEN THE POTENTIAL FOR LOSS OF DEVICE FUNCTIONALITY THE IPG WAS EXPLANTED AND REPLACED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8412120 ↗
  • InjuryAug 30, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT PART OF THE PACEMAKER SYSTEM WAS REMOVED DUE TO A SYSTEMIC INFECTION. TWO OF THE LEADS COULD ONLY BE PARTIALLY EXTRACTED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 7832499 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.

  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-1924-2025 · Apr 29, 2025 · Open, Classified
    Azure S DR MRI SureScan, Product number W3DR01
  • Z-2508-2020 · Jun 4, 2020 · Open, Classified
    Azure S DR MRI, Model Number W3DR01
  • Z-2507-2020 · Jun 4, 2020 · Open, Classified
    Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
  • Z-2509-2020 · Jun 4, 2020 · Open, Classified
    Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
  • Z-2510-2020 · Jun 4, 2020 · Open, Classified
    Azure S SR MRI, Model Number W3SR01
  • Z-0810-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
  • Z-0809-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
  • Z-0811-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
  • Z-0807-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
  • Z-0808-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
  • Z-2120-2011 · Apr 4, 2011 · Terminated
    Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
  • Z-1440-2010 · Feb 11, 2010 · Terminated
    Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
  • Z-1515-2009 · May 18, 2009 · Terminated
    Medtronic KAPPA 600/700/900 Series Implantable Pulse Generator (IPGs), Dual Chamber Rate Responsive Pacemaker, models KDR601, KDR603, KDR606, KDR651, KDR653, KDR701, KDR703, KDR706

Other Pulse generator, permanent, implantable devices

All Pulse generator, permanent, implantable devices →