Adapta™ ADDR06 (UDI 00643169708662)Recalls & reported problems
Medtronic, Inc. · IPG ADDR06 ADAPTA DR 5/6MM US NS · Class 3 · Implant
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 7
- Deaths reported
- 2
- Injuries reported
- 5
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00643169708662), received between May 15, 2018 and Aug 16, 2023 — classified as 2 deaths, 5 injuries.
The most frequently reported specific device problem is “Under-Sensing” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
15 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Jan 17, 2019Open, Classified
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Z-0806-2019 · Root cause: Component design/selection · FDA record ↗
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available5
- Adverse Event Without Identified Device or Use Problem1
- Under-Sensing1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Death1
- Insufficient Information1
- Muscle Stimulation1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryAug 16, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPLACED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, GIVEN THE POTENTIAL FOR LOSS OF DEVICE FUNCTIONALITY, THE IPG WAS EXPLANTED AND REPLACED TO PRECLUDE HARM TO THE PATIENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17555978 ↗InjuryApr 3, 2023Under-Sensing; Under-Sensing
IT WAS FURTHER REPORTED THAT THE IPG WAS EXPLANTED AND THE RV LEAD WAS CAPPED AND REPLACED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16662329 ↗DeathJan 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Insufficient Information
MAUDE report 13175065 ↗InjuryAug 13, 2021Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS FURTHER REPORTED THAT THE IPG WAS AT ELECTIVE REPLACEMENT INDICATOR (ERI).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12317301 ↗DeathSep 19, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.
Patient problems: Death
MAUDE report 9086943 ↗InjuryMar 13, 2019Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPLACED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE; HOWEVER, GIVEN THE POTENTIAL FOR LOSS OF DEVICE FUNCTIONALITY, THE IPG WAS EXPLANTED AND REPLACED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8418603 ↗InjuryMay 15, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available
IT WAS REPORTED THAT THE PATIENT EXPERIENCED POCKET STIMULATION AFTER A DEVICE CHANGE OUT. IT WAS ALSO NOTED THAT THE DEVICE WAS IMPLANTED SUB-MUSCULAR WHICH WOULD LIKELY HAVE BEEN THE CAUSE OF THE NEW POCKET STIMULATION. REPROGRAMMING WAS PERFORMED WHICH SIGNIFICANTLY DECREASED THE STIMULATION. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Muscle Stimulation
MAUDE report 7514582 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.
- Z-2128-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
- Z-2127-2025 · Jun 17, 2025 · Open, ClassifiedMedtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
- Z-1924-2025 · Apr 29, 2025 · Open, ClassifiedAzure S DR MRI SureScan, Product number W3DR01
- Z-2509-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
- Z-2510-2020 · Jun 4, 2020 · Open, ClassifiedAzure S SR MRI, Model Number W3SR01
- Z-2507-2020 · Jun 4, 2020 · Open, ClassifiedAzure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
- Z-2508-2020 · Jun 4, 2020 · Open, ClassifiedAzure S DR MRI, Model Number W3DR01
- Z-0807-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
- Z-0810-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
- Z-0809-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
- Z-0811-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
- Z-0808-2019 · Jan 17, 2019 · Open, ClassifiedMedtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
- Z-2120-2011 · Apr 4, 2011 · TerminatedMedtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
- Z-1440-2010 · Feb 11, 2010 · TerminatedMedtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
- Z-1515-2009 · May 18, 2009 · TerminatedMedtronic KAPPA 600/700/900 Series Implantable Pulse Generator (IPGs), Dual Chamber Rate Responsive Pacemaker, models KDR601, KDR603, KDR606, KDR651, KDR653, KDR701, KDR703, KDR706
Other Pulse generator, permanent, implantable devices
- Assurity™ PM2240St. Jude Medical, Inc. · Pulse generator, permanent, implantable1 recall84 deaths4,076 reports
- Fortify Assura™ CD2357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable63 deaths3,571 reports
- Advisa DR MRI™ SureScan® A2DR01Medtronic, Inc. · Pulse generator, permanent, implantable51 deaths2,298 reports
- Fortify Assura™ CD1357-40QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable35 deaths1,933 reports
- Ellipse™ CD2411-36QSt. Jude Medical, Inc. · Pulse generator, permanent, implantable42 deaths1,498 reports
- Fortify Assura™ CD2357-40CSt. Jude Medical, Inc. · Pulse generator, permanent, implantable25 deaths1,378 reports
- Accent® DR RFSt. Jude Medical, Inc. · Pulse generator, permanent, implantable73 deaths1,370 reports
- Edora 8 DR-T 407145BIOTRONIK SE & Co. KG · Pulse generator, permanent, implantable10 deaths982 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.