Skip to content

Adapta™ L ADDRL1 (UDI 00643169708686)Recalls & reported problems

Medtronic, Inc. · IPG ADDRL1 ADAPTA DR IS-1 LG US NS · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Adapta implantable pulse generator has 177 adverse event reports, with 38 deaths and 101 injuries. The reports mostly describe pacing problems, early battery drain, and devices being removed, and some death reports give no details. One recall is open, started in January 2019, over a circuit error that can affect how the device works in dual chamber mode with atrial sensing. Reports fell to none in the latest 12 months from 6 in the prior 12 months.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
177
Deaths reported
38
Injuries reported
101
Recalls naming it
1

Summary

The FDA's MAUDE database contains 177 adverse event reports that list this exact device (DI 00643169708686), received between Apr 10, 2018 and Sep 12, 2025 — classified as 38 deaths, 101 injuries, 38 malfunctions.

The most frequently reported specific device problem is “Pacing Problem” (35 reports).

1 FDA recall record name this device's identifier, 1 still open.

15 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 2 reportsDec 2023: 2 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 2 reportsSep 2024: 1 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 17, 2019Open, Classified

    A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

    Z-0806-2019 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available96
  • Pacing Problem35
  • Adverse Event Without Identified Device or Use Problem34
  • Premature Discharge of Battery3
  • Failure to Interrogate2
  • Mechanical Problem2
  • Battery Problem1
  • Device Contamination with Body Fluid1

Patient problems

  • No Known Impact Or Consequence To Patient90
  • Death22
  • No Clinical Signs, Symptoms or Conditions19
  • Insufficient Information13
  • Unspecified Infection12
  • No Consequences Or Impact To Patient7
  • Dizziness5
  • Syncope4

Recent adverse event reports

The 25 most recent of 177 reports, in the reporter's own words as published by the FDA.

  • InjurySep 12, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23037253 ↗
  • MalfunctionJul 9, 2025Premature Discharge of Battery; Premature Discharge of Battery

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED HIGH OUTPUTS WHICH PREMATURELY DEPLETED THE BATTERY. IT WAS FURTHER REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD ALSO EXHIBITED HIGH THRESHOLDS. BOTH PRODUCTS REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22458177 ↗
  • InjuryMay 29, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22103784 ↗
  • InjuryMay 8, 2025Incorrect Measurement; Incorrect Measurement

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED INCORRECTLY MEASURED BATTERY VOLTAGE. THE IPG WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21983385 ↗
  • InjuryApr 18, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED EROSION. THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Erosion

    MAUDE report 21864182 ↗
  • InjuryNov 14, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS FURTHER REPORTED THAT THE IPG WAS EXPLANTED DUE TO ELECTIVE REPLACEMENT INDICATOR (ERI).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20682980 ↗
  • InjuryOct 11, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A STROKE. THE IMPLANTABLE PULSE GENERATOR (IPG) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Stroke/CVA

    MAUDE report 20426273 ↗
  • DeathSep 13, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 20209742 ↗
  • DeathAug 23, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, A CAUSE OF DEATH COULD NOT BE IDENTIFIED. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. THIS PATIENT WAS CONFIRMED TO BE PROGRAMMED TO A PACING MODE SUSCEPTIBLE TO THE DESCRIBED ADVISORY ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 20046109 ↗
  • InjuryAug 6, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED, THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION. HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING, THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR, WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19917523 ↗
  • InjuryApr 5, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) SOFTWARE WAS UPDATED. THE DEVICE WAS CONFIRMED TO BE IN A MODE SUSCEPTIBLE TO THE ERROR. NO PROGRAMMING CHANGES WERE MADE. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19049066 ↗
  • InjuryDec 21, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPROGRAMMED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, THE DEVICE WAS FUNCTIONING IN A MODE SUSCEPTIBLE TO CIRCUIT ERROR AND WAS THEREFORE REPROGRAMMED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18370798 ↗
  • MalfunctionDec 19, 2023Pacing Problem; Pacing Problem

    IT WAS FURTHER REPORTED THAT THERE WAS NO PERFORMANCE ISSUE WITH THE IPG. THE RATE RESPONSE WAS ADJUSTED FOR THE PATIENT.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18353635 ↗
  • InjuryNov 27, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) SYSTEM WAS EXPLANTED AND REPLACED DUE TO INFECTION. DEBRIDEMENT WAS PERFORMED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Unspecified Infection

    MAUDE report 18211369 ↗
  • DeathNov 14, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED BY THE PATIENT SPOUSE THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 18129076 ↗
  • InjuryOct 12, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17923456 ↗
  • InjurySep 22, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17796448 ↗
  • InjurySep 19, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPLACED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, GIVEN THE POTENTIAL FOR LOSS OF DEVICE FUNCTIONALITY, THE IPG WAS INACTIVATED AND REPLACED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17772765 ↗
  • InjuryAug 16, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE ORIGIN OF THE INFECTION WAS UNRELATED TO THE DEVICE.

    Patient problems: Unspecified Infection

    MAUDE report 17556331 ↗
  • MalfunctionJun 7, 2023Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED A HEART RATE OF "THIRTY-EIGHT BEATS PER MINUTE", WHICH IS LOWER THAN THE PROGRAMMED LOW ER RATE OF THE IMPLANTABLE PULSE GENERATOR (IPG). THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia

    MAUDE report 17080259 ↗
  • InjuryMay 15, 2023Premature Discharge of Battery; Appropriate Device Problem Term/Code Not Available; Premature Discharge of Battery; Appropriate Device Problem Term/Code Not Available

    IT WAS FURTHER REPORTED THAT THE IPG WAS NOT EXPLANTED IN RESPONSE TO THE DUAL CIRCUIT ERROR FCA BUT DUE TO DECREASED BATTERY LIFE CAUSED BY RIGHT VENTRICULAR (RV) LEAD THRESHOLD ISSUES. THE EPICARDIAL RV LEAD WAS CAPPED AND REPLACED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16929560 ↗
  • InjuryApr 4, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16674399 ↗
  • InjuryDec 21, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16020896 ↗
  • DeathOct 5, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information

    MAUDE report 15544569 ↗
  • InjuryMar 8, 2022Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED SYNCOPE AND BRADYCARDIA. THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED HIGH THRESHOLD, HIGH IMPEDANCE AND NOISE ON VENTRICULAR HIGH RATE EPISODES. BOTH THE RV LEAD AND IMPLANTABLE PULSE GENERATOR (IPG) REMAIN IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Bradycardia; Syncope/Fainting

    MAUDE report 13693060 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.

  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-1924-2025 · Apr 29, 2025 · Open, Classified
    Azure S DR MRI SureScan, Product number W3DR01
  • Z-2510-2020 · Jun 4, 2020 · Open, Classified
    Azure S SR MRI, Model Number W3SR01
  • Z-2508-2020 · Jun 4, 2020 · Open, Classified
    Azure S DR MRI, Model Number W3DR01
  • Z-2507-2020 · Jun 4, 2020 · Open, Classified
    Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
  • Z-2509-2020 · Jun 4, 2020 · Open, Classified
    Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
  • Z-0807-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
  • Z-0811-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
  • Z-0808-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
  • Z-0810-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
  • Z-0809-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
  • Z-2120-2011 · Apr 4, 2011 · Terminated
    Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
  • Z-1440-2010 · Feb 11, 2010 · Terminated
    Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
  • Z-1515-2009 · May 18, 2009 · Terminated
    Medtronic KAPPA 600/700/900 Series Implantable Pulse Generator (IPGs), Dual Chamber Rate Responsive Pacemaker, models KDR601, KDR603, KDR606, KDR651, KDR653, KDR701, KDR703, KDR706

Other Pulse generator, permanent, implantable devices

All Pulse generator, permanent, implantable devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.