Adapta™ L ADDRL1 (UDI 00643169708709)Recalls & reported problems

Medtronic, Inc. · IPG ADDRL1 ADAPTA DR IS-1 LG US NS · Class 3 · Implant

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Medtronic Adapta pacemaker has 200 adverse event reports, including 44 deaths and 110 injuries. Many reports describe pacing and sensing problems, and generators being removed for unknown reasons, along with precautionary reprogramming after a circuit error advisory. It has 1 recall, still open, started in January 2019, over a circuit error that can affect how some pacemakers work when set to a certain mode.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
200
Deaths reported
44
Injuries reported
110
Recalls naming it
1

Summary

The FDA's MAUDE database contains 200 adverse event reports that list this exact device (DI 00643169708709), received between Dec 19, 2017 and Feb 13, 2026 — classified as 44 deaths, 110 injuries, 46 malfunctions.

The most frequently reported specific device problem is “Pacing Problem” (54 reports).

1 FDA recall record name this device's identifier, 1 still open.

15 other recalls cover devices cleared under the same FDA submission (P980035); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 2 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Jan 17, 2019Open, Classified

    A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality

    Z-0806-2019 · Root cause: Component design/selection · FDA record ↗

Most reported problems

Device problems

  • Appropriate Device Problem Term/Code Not Available114
  • Pacing Problem54
  • Adverse Event Without Identified Device or Use Problem24
  • Insufficient Device Problem Information3
  • Device Sensing Problem2
  • Failure to Capture2
  • Failure to Interrogate2
  • Premature Discharge of Battery2

Patient problems

  • No Known Impact Or Consequence To Patient105
  • No Clinical Signs, Symptoms or Conditions27
  • Death26
  • Insufficient Information16
  • Unspecified Infection5
  • Dizziness4
  • Syncope4
  • Atrial Fibrillation3

Recent adverse event reports

The 25 most recent of 200 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 13, 2026Appropriate Device Problem Term/Code Not Available

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPROGRAMMED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, THE DEVICE WAS FUNCTIONING IN A MODE SUSCEPTIBLE TO CIRCUIT ERROR AND WAS THEREFORE REPROGRAMMED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24330626 ↗
  • InjurySep 16, 2025Device Sensing Problem; Appropriate Device Problem Term/Code Not Available; Device Sensing Problem; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED AN OCCLUSION. IT WAS FURTHER REPORTED THAT THE RIGHT ATRIAL (RA) LEAD EXHIBITED OVERSENSING OF P-WAVES. IT WAS NOTED THAT THE RIGHT VENTRICULAR (RV) HIS BUNDLE LEAD EXHIBITED UNDERSENSING OF R-WAVES. IT WAS FURTHER NOTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS INACTIVATED AND REPLACED DUE TO UNKNOWN REASONS. THE RA LEAD WAS CAPPED. THE RV HIS LEAD CAPPED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. 2025-10-20: IT WAS FURTHER REPORTED THAT THE IPG DID NOT EXHIBIT A PERFORMANCE ISSUE. 2026-01-30: IT WAS FURTHER REPORTED THAT THE PATIENT EXPERIENCED ATRIAL FIBRILLATION WITH FREQUENT VENTRICULAR PACED COMPLEXES AND RIGHT BUNDLE BRANCH BLOCK. IT WAS NOTED THAT THE RV HIS BUNDLE LEAD EXHIBITED CONTINUED UNDERSENSING WHICH CAUSED INAPPROPRIATE PACING. IT WAS FURTHER NOTED THAT THE IPG, RV HIS BUNDLE AND RA LEAD EXHIBITED FAILURE TO SENSE. THE HIS BUNDLE LEAD WAS REPROGRAMMED.

    Patient problems: Atrial Fibrillation; No Clinical Signs, Symptoms or Conditions; Non specific EKG/ECG Changes

    MAUDE report 23062214 ↗
  • InjuryJun 4, 2025Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22138897 ↗
  • InjuryDec 30, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21033370 ↗
  • DeathNov 22, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED, THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION. HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL, FOR DEVICE CIRCUIT ERROR TO OCCUR, WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20757065 ↗
  • InjuryOct 16, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20460524 ↗
  • InjuryJun 14, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS FURTHER REPORTED THAT THE IPG EXPLANT WAS SUSPECTED TO BE DUE TO NORMAL BATTERY DEPLETION AND NOT DUE TO THE DEVICE CIRCUIT ERROR.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19537157 ↗
  • MalfunctionMay 3, 2024Failure to Interrogate; Failure to Interrogate

    IT WAS REPORTED THAT THE REMOTE MONITOR WAS NOT ESTABLISHING TELEMETRY WITH THE IMPLANTABLE PULSE GENERATOR (IPG). TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. THE PATIENT WAS REFERRED TO THE CLINIC. THE MONITOR REMAINS IN USE. THE IPG REMAIN IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19236508 ↗
  • MalfunctionMay 1, 2024Failure to Interrogate; Failure to Interrogate

    IT WAS REPORTED BY THE PATIENT'S REPRESENTATIVE THAT THE REMOTE MONITOR HAD A TELEMETRY ISSUE. THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED A POTENTIAL TELEMETRY ISSUE. TROUBLESHOOTING STEPS WERE TAKEN TO NO AVAIL. CALLER WAS REFERRED TO CLINIC. THE REMOTE MONITOR REMAINS IN USE. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19219774 ↗
  • InjuryFeb 16, 2024Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18721232 ↗
  • MalfunctionOct 27, 2023Communication or Transmission Problem; Communication or Transmission Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS UNABLE TO SEND TRANSMISSION. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT. IT WAS ALSO REPORTED THAT THE APPLICATION WAS NOT INTERROGATING THE PATIENTS DEVICE. NO PATIENT COMPLICATIONS. THIS EVENT IS BEING REPORTED IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18020318 ↗
  • InjuryMay 15, 2023Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED POSSIBLE ASYSTOLE/ BRADYCARDIA. IT WAS ALSO REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PACING BELOW THE PROGRAMMED LOWER RATE. THE IPG REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Asystole; Bradycardia

    MAUDE report 16930830 ↗
  • DeathApr 28, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, A CAUSE OF DEATH COULD NOT BE IDENTIFIED. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. THIS PATIENT WAS CONFIRMED TO BE PROGRAMMED TO A PACING MODE SUSCEPTIBLE TO THE DESCRIBED ADVISORY ISSUE.

    Patient problems: Insufficient Information

    MAUDE report 16830522 ↗
  • DeathApr 18, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, A CAUSE OF DEATH COULD NOT BE IDENTIFIED. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information

    MAUDE report 16761955 ↗
  • InjuryApr 3, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS FURTHER REPORTED THAT THE IPG WAS NOT EXPLANTED TO THE FIELD CORRECTIVE ACTION (FCA) ON THE IPG, BUT DUE TO PATIENT INFECTION UNRELATED TO THE DEVICE. IT WAS NOTED THAT THE PATIENT EXPERIENCED RIGHT KNEE INFECTION FROM RECENT KNEE REPLACEMENT SURGERY WHICH LED TO ENDOCARDITIS AND THE IPG NEEDED TO BE EXPLANTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16663895 ↗
  • InjuryMar 7, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16494228 ↗
  • MalfunctionMar 1, 2023Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) EXHIBITED PACING SPIKES. THE IPG REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16458139 ↗
  • DeathFeb 2, 2023Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, A CAUSE OF DEATH COULD NOT BE IDENTIFIED. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE. THIS PATIENT WAS CONFIRMED TO BE PROGRAMMED TO A PACING MODE SUSCEPTIBLE TO THE DESCRIBED ADVISORY ISSUE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 16282525 ↗
  • DeathNov 24, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information

    MAUDE report 15860549 ↗
  • DeathSep 23, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, A CAUSE OF DEATH COULD NOT BE IDENTIFIED. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information

    MAUDE report 15469318 ↗
  • InjurySep 8, 2022Pacing Problem; Pacing Problem

    IT WAS REPORTED THAT THE PATIENT EXPERIENCED TWO HUNDRED AND TEN PAUSES OVER A TWO WEEK PERIOD. IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) IS INDICATED FOR EXHIBITING PAUSES. IT WAS ALSO NOTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED LOW IMPEDANCE LEADING TO A POLARITY SWITCH, AND THAT UNDERPACING AND OVERSENSING WERE OBSERVED ON A ZIO PATCH. IT WAS ALSO NOTED THAT THE RV LEAD EXHIBITED A LOSS OF OVERSENSING THAT LED TO A LACK OF PACING AND PACING SPIKES WITH A LOSS OF OUTPUT. THE ELECTROPHYSIOLOGIST BELIEVES THAT AN INNER INSULATION ISSUE WAS THE CAUSE. THE IPG AND RV LEAD WERE REMOVED AND REPLACED. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: Non specific EKG/ECG Changes

    MAUDE report 15377456 ↗
  • DeathAug 29, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    IT WAS REPORTED THAT THE PATIENT DIED. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: Insufficient Information

    MAUDE report 15309582 ↗
  • InjuryJul 15, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15027535 ↗
  • InjuryJul 4, 2022Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT THE IMPLANTABLE PULSE GENERATOR (IPG) WAS EXPLANTED FOR AN UNKNOWN REASON. FOLLOW-UP WAS ATTEMPTED TO OBTAIN THE RELEVANT INFORMATION; HOWEVER, NO ADDITIONAL DETAILS ARE AVAILABLE AT THIS TIME. THIS EVENT IS BEING CONSERVATIVELY REPORTED IN AN ABUNDANCE OF CAUTION IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EPISODE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14904664 ↗
  • InjuryJun 14, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IN RESPONSE TO THE ADVISORY DESCRIBING THE POTENTIAL FOR A DEVICE CIRCUIT ERROR TO OCCUR WHILE THE DEVICE IS PROCESSING AN ATRIAL-SENSED EVENT, THE IMPLANTABLE PULSE GENERATOR (IPG) WAS PROPHYLACTICALLY REPROGRAMMED. NO DEVICE MALFUNCTION WAS OBSERVED WHILE THE DEVICE WAS IN USE, HOWEVER, THE DEVICE WAS FUNCTIONING IN A MODE SUSCEPTIBLE TO CIRCUIT ERROR AND WAS THEREFORE REPROGRAMMED TO PRECLUDE HARM TO THE PATIENT. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14683370 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P980035. One submission often covers many models, so check the product and lot details in each record.

  • Z-2127-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Azure Astra application, Software Model Number D
  • Z-2128-2025 · Jun 17, 2025 · Open, Classified
    Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Azure Astra application, Software Model N
  • Z-1924-2025 · Apr 29, 2025 · Open, Classified
    Azure S DR MRI SureScan, Product number W3DR01
  • Z-2510-2020 · Jun 4, 2020 · Open, Classified
    Azure S SR MRI, Model Number W3SR01
  • Z-2507-2020 · Jun 4, 2020 · Open, Classified
    Azure XT DR MRI, Model Numbers: a) W1DR01 and b) W2DR01
  • Z-2509-2020 · Jun 4, 2020 · Open, Classified
    Azure XT SR MRI, Model Numbers: a) W1SR01 and b) W2SR01
  • Z-2508-2020 · Jun 4, 2020 · Open, Classified
    Azure S DR MRI, Model Number W3DR01
  • Z-0810-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Sphera DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number SPDR01 (b) Model Number SPDRL1
  • Z-0808-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: RELIA, Dual chamber pacemaker (VDD): (a) Model Number RED01 (b) Model Number REDR01 (c) Model Number REVDD01
  • Z-0809-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: SENSIA, Dual chamber pacemaker (DDD): (a) Model Number SED01 (b) Model Number SEDR01 (c) Model Number SEDRL1
  • Z-0807-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: ATTESTA L DR MRI SureScan, Dual chamber rate responsive pacemaker (DDDR): (a) Model Number ATDR01 (b) Model Number ATDRL1 (c) Model Number
  • Z-0811-2019 · Jan 17, 2019 · Open, Classified
    Medtronic implantable pulse generator: Versa, Dual chamber rate responsive pacemaker (DDDR), Model Number VEDR01
  • Z-2120-2011 · Apr 4, 2011 · Terminated
    Medtronic, Kappa 600 KDR600,KDR601,KDR603,KDR606,KDR651,KDR653 The devices are implantable pulse generators (IPG) used to detect bradycardia events and provide pacing support. Wit
  • Z-1440-2010 · Feb 11, 2010 · Terminated
    Medtronic EnRhythm, P1501DR, Dual Chamber Rate Responsive Pacemaker with RapidRead Telemetry. The EnRhythm pacing system is an implantable medical device that monitors, detects,
  • Z-1515-2009 · May 18, 2009 · Terminated
    Medtronic KAPPA 600/700/900 Series Implantable Pulse Generator (IPGs), Dual Chamber Rate Responsive Pacemaker, models KDR601, KDR603, KDR606, KDR651, KDR653, KDR701, KDR703, KDR706

Other Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices

All Drug eluting permanent right ventricular (RV) or right atrial (RA) pacemaker electrodes devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.